The Influence of Orthokeratology on Intraocular Pressure Measurements

被引:9
|
作者
Chang, Chia-Jen [1 ,3 ]
Yang, Hsi-Hsien [1 ]
Chang, Chin-An [1 ]
Wu, Richard [2 ]
Tsai, Hsien-Yang [3 ]
机构
[1] Chaoyang Univ Technol, Grad Inst Biochem Sci & Technol, Taichung 40644, Taiwan
[2] Cent Taiwan Univ Sci & Technol, Dept Optometry, Taichung, Taiwan
[3] Execut Yuan, Vet Affairs Commiss, Taichung Vet Gen Hosp, Dept Ophthalmol, Taichung, Taiwan
关键词
Central corneal thickness; intraocular pressure; myopia; orthokeratology; tonometry; DYNAMIC CONTOUR TONOMETER; CENTRAL CORNEAL THICKNESS; GOLDMANN APPLANATION; LENS WEAR; PNEUMOTONOMETRY; CURVATURE; GLAUCOMA;
D O I
10.3109/08820538.2013.768679
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: Underestimation of IOP in a myopic patient may lead to misjudgment of the risk of glaucoma. This study investigated the influence of orthokeratology-induced change in CCT on IOP measured by the non-contact pneumotonometer (NCT), Goldmann applanation tonometer (GAT), and Pascal dynamic contour tonometer (PDCT). Methods: This study was conducted to examine the eyes of 34 patients who received orthokeratology for myopia. CCT and IOP were measured, and IOP was obtained with the NCT, GAT, and PDCT. The associations between changes in measured IOP and change in CCT at different orthokeratology follow-up time points were evaluated by linear regression analysis. Results: Change in IOP measured by the three tonometries correlated significantly with change in CCT after one-week application of orthokeratology. The correlation was strongest for NCT followed by GAT and PDCT. The changes in measured IOP corresponding to a 10 mm decrease in CCT were 0.7-0.9, 0.4-0.6, and 0.2-0.3 mm Hg for NCT, GAT, and PDCT, respectively. Conclusions: The IOP measured by the three methods - NCT, GAT, and PDCT - decreased as a result of the change in CCT induced by orthokeratology. The influence on NCT and GAT was greater than that on PDCT.
引用
收藏
页码:210 / 215
页数:6
相关论文
共 50 条
  • [21] The Genetics of Intraocular Pressure
    Ojha, Pallavi
    Wiggs, Janey L.
    Pasquale, Louis R.
    SEMINARS IN OPHTHALMOLOGY, 2013, 28 (5-6) : 301 - 305
  • [22] Association Between Central Corneal Thickness and Intraocular Pressure in Patients with Refractive Anomalies and Emmetropes
    Ismaili, Mimoza
    INTERNATIONAL JOURNAL OF BIOMEDICINE, 2023, 13 (03) : 91 - 95
  • [23] Ex-vivo experimental validation of biomechanically-corrected intraocular pressure measurements on human eyes using the CorVis ST
    Eliasy, Ashkan
    Chen, Kai-Jung
    Vinciguerra, Riccardo
    Maklad, Osama
    Vinciguerra, Paolo
    Ambrosio Jr, Renato
    Roberts, Cynthia J.
    Elsheikh, Ahmed
    EXPERIMENTAL EYE RESEARCH, 2018, 175 : 98 - 102
  • [24] Comparisons of intraocular pressure measurements: Goldmann applanation tonometry, noncontact tonometry, Tono-Pen tonometry, and dynamic contour tonometry
    Hsu, S-Y
    Sheu, M-M
    Hsu, A-H
    Wu, K-Y
    Yeh, J-I
    Tien, J-N
    Tsai, R-K
    EYE, 2009, 23 (07) : 1582 - 1588
  • [25] Intraocular pressure measurement in children
    Bresson-Dumont, H.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2009, 32 (03): : 176 - 181
  • [26] Techniques of Intraocular Pressure Measurement
    Neuburger, M.
    Rosentreter, A.
    Dietlein, T. S.
    Jordan, J. F.
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2011, 228 (02) : 118 - 124
  • [27] Comparison of Intraocular Pressure Measurements With Goldmann Applanation Tonometry, Tonopen XL, and Pascal Dynamic Contour Tonometry in Patients With Descemet Membrane Endothelial Keratoplasty
    Yildiz, Izlem
    Altan, Cigdem
    Cakmak, Semih
    Genc, Selim
    Yildirim, Yusuf
    Agca, Alper
    JOURNAL OF GLAUCOMA, 2022, 31 (11) : 909 - 914
  • [28] Twenty-four hour intraocular pressure measurements and home tonometry
    Meier-Gibbons, Frances
    Berlin, Michael S.
    Toteberg-Harms, Marc
    CURRENT OPINION IN OPHTHALMOLOGY, 2018, 29 (02) : 111 - 115
  • [29] Dynamic contour tonometry for post-LASIK intraocular pressure measurements
    Duba, I
    Wirthlin, AC
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2004, 221 (05) : 347 - 350
  • [30] Impact of corneal parameters on intraocular pressure measurements in different tonometry methods
    Aleksandra Zakrzewska
    Marta P.Wi?cek
    Anna Machalińska
    International Journal of Ophthalmology, 2019, (12) : 1853 - 1858