Breast volume assessment based on 3D surface geometry:: verification of the method using MR imaging

被引:36
作者
Eder, Maximilian [1 ]
Schneider, Armin [2 ]
Feussner, Hubertus [2 ]
Zimmermann, Alexander [3 ]
Hoehnke, Christoph [1 ]
Papadopulos, Nikolaos A. [1 ]
Kovacs, Laszlo [1 ]
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Klin & Poliklin Plast Chirurg & Handchirurg, Arbeitsgrp Comp Aided Plast Surg, D-81675 Munich, Germany
[2] Tech Univ Munich, Klinikum Rechts Isar, Forsch Grp MITI, D-81675 Munich, Germany
[3] Tech Univ Munich, Klinikum Rechts Isar, Abt Gefasschirurg, D-81675 Munich, Germany
来源
BIOMEDIZINISCHE TECHNIK | 2008年 / 53卷 / 03期
关键词
breast; breast volume; plastic surgery; 3D imaging; 3D scan; 3D segmentation;
D O I
10.1515/BMT.2008.017
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Differences in breast volume and contour are subjectively estimated by surgeons. 3D surface imaging using 3D scanners provides objective breast volume quantification, but precision and accuracy of the method requires verification. Breast volumes of five test individuals were assessed using a 3D surface scanner. Magnetic resonance imaging (MRI) reference volumes were obtained to verify and compare the 3D scan measurements. The anatomical thorax wall curvature was segmented using MRI data and compared to the interpolated curvature of the posterior breast volume delimitation of 3D scan data. MRI showed higher measurement precision, mean deviation (expressed as percentage of volume) of 1.1 0 0.34% compared to 1.63 0.53% for the 3D scanner. Mean MRI [right (left) breasts: 638 (629) 143 (138) cc] and 3D scan [right (left) breasts: 493 (497) 112 (116) cc] breast volumes significantly correlated [right (left) breasts: r=0.982 (0.977), p=0.003 (0.004)]. The posterior thorax wall of the 3D scan model showed high agreement with the MRI thorax wall curvature [mean positive (negative) deviation: 0.33 (-0.17) 0.37 cm]. High correspondence and correlation of 3D scan data with MR]based verifications support 3D surface imaging as sufficiently precise and accurate for breast volume measurements.
引用
收藏
页码:112 / 121
页数:10
相关论文
共 32 条
[1]   Estimation of breast prosthesis volume by the Cavalieri principle using magnetic resonance images [J].
Akbas, H ;
Sahin, B ;
Eroglu, L ;
Odaci, E ;
Bilgic, S ;
Kaplan, S ;
Uzun, A ;
Ergur, H ;
Bek, Y .
AESTHETIC PLASTIC SURGERY, 2004, 28 (05) :275-280
[2]  
BOUMAN F G, 1970, British Journal of Plastic Surgery, V23, P263, DOI 10.1016/S0007-1226(70)80053-4
[3]   A formula for surgical modifications of the breast [J].
Brown, RW ;
Cheng, YC ;
Kurtay, M .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2000, 106 (06) :1342-1345
[4]   Breast volume assessment: comparing five different techniques [J].
Bulstrode, N ;
Bellamy, E ;
Shrotria, S .
BREAST, 2001, 10 (02) :117-123
[5]   3-DIMENSIONAL INPUT OF BODY-SURFACE DATA USING A LASER-LIGHT SCANNER [J].
CUTTING, CB ;
MCCARTHY, JG ;
KARRON, DB .
ANNALS OF PLASTIC SURGERY, 1988, 21 (01) :38-45
[6]   3-DIMENSIONAL ALPHA-SHAPES [J].
EDELSBRUNNER, H ;
MUCKE, EP .
ACM TRANSACTIONS ON GRAPHICS, 1994, 13 (01) :43-72
[7]   Virtual 3-dimensional modeling as a valuable adjunct to aesthetic and reconstructive breast surgery [J].
Eder, Maximilian ;
Papadopulos, Nikolaos A. ;
Kovacs, Laszlo .
AMERICAN JOURNAL OF SURGERY, 2007, 194 (04) :563-565
[8]   Breast volume determination in breast hypertrophy [J].
Eder, Maximilian ;
Papadopulos, Nikolaos A. ;
Kovacs, Laszlo .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2007, 120 (01) :356-357
[9]   Measurement of breast volume with thermoplastic casts [J].
EdsanderNord, A ;
Wickman, M ;
Jurell, G .
SCANDINAVIAN JOURNAL OF PLASTIC AND RECONSTRUCTIVE SURGERY AND HAND SURGERY, 1996, 30 (02) :129-132
[10]  
FALCAO AX, 1997, SPIE MED IMAGING, V3034, P191