Radiation dose reduction in cerebral CT perfusion imaging using iterative reconstruction

被引:41
作者
Niesten, Joris M. [1 ]
van der Schaaf, Irene C. [1 ]
Riordan, Alan J. [1 ]
de Jong, Hugo W. A. M. [1 ]
Horsch, Alexander D. [2 ]
Eijspaart, Daniel [1 ]
Smit, Ewoud J. [1 ]
Mali, Willem P. T. M. [1 ]
Velthuis, Birgitta K. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiol, NL-3584 CX Utrecht, Netherlands
[2] Rijnstate Hosp, Dept Radiol, Arnhem, Netherlands
关键词
Cerebral perfusion; Stroke; Infarction; Computed tomography; Radiation dose; TEMPORAL RESOLUTION; ISCHEMIC-STROKE; PARAMETERS; ALGORITHMS; ACCURACY; QUALITY; VOLUME; NOISE;
D O I
10.1007/s00330-013-3042-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To investigate whether iterative reconstruction (IR) in cerebral CT perfusion (CTP) allows for 50 % dose reduction while maintaining image quality (IQ). A total of 48 CTP examinations were reconstructed into a standard dose (150 mAs) with filtered back projection (FBP) and half-dose (75 mAs) with two strengths of IR (middle and high). Objective IQ (quantitative perfusion values, contrast-to-noise ratio (CNR), penumbra, infarct area and penumbra/infarct (P/I) index) and subjective IQ (diagnostic IQ on a four-point Likert scale and overall IQ binomial) were compared among the reconstructions. Half-dose CTP with high IR level had, compared with standard dose with FBP, similar objective (grey matter cerebral blood volume (CBV) 4.4 versus 4.3 mL/100 g, CNR 1.59 versus 1.64 and P/I index 0.74 versus 0.73, respectively) and subjective diagnostic IQ (mean Likert scale 1.42 versus 1.49, respectively). The overall IQ in half-dose with high IR level was scored lower in 26-31 %. Half-dose with FBP and with the middle IR level were inferior to standard dose with FBP. With the use of IR in CTP imaging it is possible to examine patients with a half dose without significantly altering the objective and diagnostic IQ. The standard dose with FBP is still preferable in terms of subjective overall IQ in about one quarter of patients. Key points Computed tomography perfusion (CTP) is increasingly important in ischaemia imaging. Radiation exposure of CTP is a drawback. Iterative reconstruction (IR) allows reduction of radiation dose in unenhanced head CT. CTP IR enables 50 % dose reduction without altering objective and diagnostic quality.
引用
收藏
页码:484 / 493
页数:10
相关论文
共 50 条
  • [31] Effects of adaptive statistical iterative reconstruction on radiation dose reduction and diagnostic accuracy of pediatric abdominal CT
    Bae, Sohi
    Kim, Myung-Joon
    Yoon, Choon-Sik
    Kim, Dong Wook
    Hong, Jung Hwa
    Lee, Mi-Jung
    [J]. PEDIATRIC RADIOLOGY, 2014, 44 (12) : 1541 - 1547
  • [32] Radiation Dose Reduction of Chest CT with Iterative Reconstruction in Image Space - Part II: Assessment of Radiologists' Preferences Using Dual Source CT
    Hwang, Hye Jeon
    Seo, Joon Beom
    Lee, Jin Seong
    Song, Jae-Woo
    Kim, Song Soo
    Lee, Hyun Joo
    Lim, Chae Hun
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2012, 13 (06) : 720 - 727
  • [33] Reduction in Radiation Exposure of CT Perfusion by Optimized Imaging Timing Using Temporal Information of the Preceding CT Angiography of the Carotid Artery in the Stroke Protocol
    Deak, Zsuzsanna
    Schuettoff, Lara
    Lohse, Ann-Kathrin
    Fabritius, Matthias
    Reidler, Paul
    Forbrig, Robert
    Kunz, Wolfgang
    Dimitriadis, Konstantin
    Ricke, Jens
    Sabel, Bastian
    [J]. DIAGNOSTICS, 2022, 12 (11)
  • [34] Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography
    Gordic, Sonja
    Desbiolles, Lotus
    Sedlmair, Martin
    Manka, Robert
    Plass, Andre
    Schmidt, Bernhard
    Husarik, Daniela B.
    Maisano, Francesco
    Wildermuth, Simon
    Alkadhi, Hatem
    Leschka, Sebastian
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (02) : 459 - 468
  • [35] Standard chest CT using combined automated tube potential selection and iterative reconstruction: image quality and radiation dose reduction
    Chae, In Hye
    Kim, Yookyung
    Lee, So Won
    Park, Ji Eun
    Shim, Sung Shine
    Lee, Jin Hwa
    [J]. CLINICAL IMAGING, 2014, 38 (05) : 641 - 647
  • [36] Radiation Dose Reduction at Pediatric CT: Use of Low Tube Voltage and Iterative Reconstruction
    Nagayama, Yasunori
    Oda, Seitaro
    Nakaura, Takeshi
    Tsuji, Akinori
    Urata, Joji
    Furusawa, Mitsuhiro
    Utsunomiya, Daisuke
    Funama, Yoshinori
    Kidoh, Masafumi
    Yamashita, Yasuyuki
    [J]. RADIOGRAPHICS, 2018, 38 (05) : 1421 - +
  • [37] Efficient radiation dose reduction in whole-brain CT perfusion imaging using a 3D GAN: performance and clinical feasibility
    Dashtbani Moghari, Mahdieh
    Zhou, Luping
    Yu, Biting
    Young, Noel
    Moore, Krystal
    Evans, Andrew
    Fulton, Roger R.
    Kyme, Andre Z.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (07)
  • [38] Impact of temporal resolution on perfusion metrics, therapy decision, and radiation dose reduction in brain CT perfusion in patients with suspected stroke
    Rau, Alexander
    Reisert, Marco
    Stein, Thomas
    Mueller-Peltzer, Katharina
    Rau, Stephan
    Bamberg, Fabian
    Taschner, Christian A.
    Urbach, Horst
    Kellner, Elias
    [J]. NEURORADIOLOGY, 2024, 66 (05) : 749 - 759
  • [39] Iterative Reconstruction for Head CT: Effects on Radiation Dose and Image Quality
    Rivers-Bowerman, Michael D.
    Shankar, Jai Jai Shiva
    [J]. CANADIAN JOURNAL OF NEUROLOGICAL SCIENCES, 2014, 41 (05) : 620 - 625
  • [40] Dose reduction with iterative reconstruction: Optimization of CT protocols in clinical practice
    Greffier, J.
    Macri, F.
    Larbi, A.
    Fernandez, A.
    Khasanova, E.
    Pereira, F.
    Mekkaoui, C.
    Beregi, J. P.
    [J]. DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2015, 96 (05) : 477 - 486