Radiation dose reduction in cerebral CT perfusion imaging using iterative reconstruction

被引:42
作者
Niesten, Joris M. [1 ]
van der Schaaf, Irene C. [1 ]
Riordan, Alan J. [1 ]
de Jong, Hugo W. A. M. [1 ]
Horsch, Alexander D. [2 ]
Eijspaart, Daniel [1 ]
Smit, Ewoud J. [1 ]
Mali, Willem P. T. M. [1 ]
Velthuis, Birgitta K. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiol, NL-3584 CX Utrecht, Netherlands
[2] Rijnstate Hosp, Dept Radiol, Arnhem, Netherlands
关键词
Cerebral perfusion; Stroke; Infarction; Computed tomography; Radiation dose; TEMPORAL RESOLUTION; ISCHEMIC-STROKE; PARAMETERS; ALGORITHMS; ACCURACY; QUALITY; VOLUME; NOISE;
D O I
10.1007/s00330-013-3042-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To investigate whether iterative reconstruction (IR) in cerebral CT perfusion (CTP) allows for 50 % dose reduction while maintaining image quality (IQ). A total of 48 CTP examinations were reconstructed into a standard dose (150 mAs) with filtered back projection (FBP) and half-dose (75 mAs) with two strengths of IR (middle and high). Objective IQ (quantitative perfusion values, contrast-to-noise ratio (CNR), penumbra, infarct area and penumbra/infarct (P/I) index) and subjective IQ (diagnostic IQ on a four-point Likert scale and overall IQ binomial) were compared among the reconstructions. Half-dose CTP with high IR level had, compared with standard dose with FBP, similar objective (grey matter cerebral blood volume (CBV) 4.4 versus 4.3 mL/100 g, CNR 1.59 versus 1.64 and P/I index 0.74 versus 0.73, respectively) and subjective diagnostic IQ (mean Likert scale 1.42 versus 1.49, respectively). The overall IQ in half-dose with high IR level was scored lower in 26-31 %. Half-dose with FBP and with the middle IR level were inferior to standard dose with FBP. With the use of IR in CTP imaging it is possible to examine patients with a half dose without significantly altering the objective and diagnostic IQ. The standard dose with FBP is still preferable in terms of subjective overall IQ in about one quarter of patients. Key points Computed tomography perfusion (CTP) is increasingly important in ischaemia imaging. Radiation exposure of CTP is a drawback. Iterative reconstruction (IR) allows reduction of radiation dose in unenhanced head CT. CTP IR enables 50 % dose reduction without altering objective and diagnostic quality.
引用
收藏
页码:484 / 493
页数:10
相关论文
共 28 条
[1]   CT Perfusion in Acute Ischemic Stroke: A Comparison of 2-Second and 1-Second Temporal Resolution [J].
Abels, B. ;
Klotz, E. ;
Tomandl, B. F. ;
Villablanca, J. P. ;
Kloska, S. P. ;
Lell, M. M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (09) :1632-1639
[2]   Acute stroke: a comparison of different CT perfusion algorithms and validation of ischaemic lesions by follow-up imaging [J].
Abels, Benjamin ;
Villablanca, J. Pablo ;
Tomandl, Bernd F. ;
Uder, Michael ;
Lell, Michael M. .
EUROPEAN RADIOLOGY, 2012, 22 (12) :2559-2567
[3]   Reducing the radiation dose for low-dose CT of the paranasal sinuses using iterative reconstruction: Feasibility and image quality [J].
Bulla, Stefan ;
Blanke, Philipp ;
Hassepass, Frederike ;
Krauss, Tobias ;
Winterer, Jan Thorsten ;
Breunig, Christine ;
Langer, Mathias ;
Pache, Gregor .
EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (09) :2246-2250
[4]   Combination of a Low-Tube-Voltage Technique With Hybrid Iterative Reconstruction (iDose) Algorithm at Coronary Computed Tomographic Angiography [J].
Funama, Yoshinori ;
Taguchi, Katsuyuki ;
Utsunomiya, Daisuke ;
Oda, Seitaro ;
Yanaga, Yumi ;
Yamashita, Yasuyuki ;
Awai, Kazuo .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2011, 35 (04) :480-485
[5]   Dynamic perfusion CT Optimizing the temporal resolution for the calculation of perfusion CT parameters in stroke patients [J].
Kaemena, Andreas ;
Streitparth, Florian ;
Grieser, Christian ;
Lehrnkuhl, Lukas ;
Jamil, Basil ;
Wojtal, Katarzyna ;
Ricke, Jens ;
Pech, Maciej .
EUROPEAN JOURNAL OF RADIOLOGY, 2007, 64 (01) :111-118
[6]   Lowering the Dose in Head CT Using Adaptive Statistical Iterative Reconstruction [J].
Kilic, K. ;
Erbas, G. ;
Guryildirim, M. ;
Arac, M. ;
Ilgit, E. ;
Coskun, B. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (09) :1578-1582
[7]   Increasing Sampling Interval in Cerebral Perfusion CT: Limitation for the Maximum Slope Model [J].
Kloska, Stephan P. ;
Fischer, Tobias ;
Sauerland, Cristina ;
Buerke, Boris ;
Dziewas, Rainer ;
Fischbach, Roman ;
Heindel, Walter .
ACADEMIC RADIOLOGY, 2010, 17 (01) :61-66
[8]   Iterative Reconstruction in Head CT: Image Quality of Routine and Low-Dose Protocols in Comparison with Standard Filtered Back-Projection [J].
Korn, A. ;
Fenchel, M. ;
Bender, B. ;
Danz, S. ;
Hauser, T. K. ;
Ketelsen, D. ;
Flohr, T. ;
Claussen, C. D. ;
Heuschmid, M. ;
Ernemann, U. ;
Brodoefel, H. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (02) :218-224
[9]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174
[10]   Perfusion CT Improves Diagnostic Accuracy for Hyperacute Ischemic Stroke in the 3-Hour Window: Study of 100 Patients with Diffusion MRI Confirmation [J].
Lin, Ke ;
Do, Kinh G. ;
Ong, Phat ;
Shapiro, Maksim ;
Babb, James S. ;
Siller, Keith A. ;
Pramanik, Bidyut K. .
CEREBROVASCULAR DISEASES, 2009, 28 (01) :72-79