Dosimetry of a prototype retractable eMLC for fixed-beam electron therapy

被引:70
作者
Hogstrom, KR
Boyd, RA
Antolak, JA
Svatos, MM
Faddegon, BA
Rosenman, JG
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Siemens Med Solut USA Inc, Oncol Care Syst, Concord, CA 94520 USA
[3] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA 94143 USA
[4] Univ N Carolina, Dept Radiat Oncol, Chapel Hill, NC 27599 USA
关键词
electron collimation; multileaf collimator; electron beam dosimetry;
D O I
10.1118/1.1644516
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An electron multileaf collimator (eMLC) has been designed that is unique in that it retracts to 37 cm from the isocenter [63-cm source-to-collimator distance (SCD)] and can be deployed to distances of 20 and 10 cm from the isocenter (80 and 90 cm SCD, respectively). It is expected to be capable of arc therapy at 63 cm SCD; isocentric, fixed-beam therapy at 80 cm SCD; and source-to-surface distance (SSD), fixed-beam therapy at 90 cm SCD. In all positions, its leaves could be used for unmodulated or intensity-modulated therapy. Our goal in the present work is to describe the general characteristics of the eMLC and to demonstrate that its leakage characteristics and dosimetry are adequate for SSD, fixed-beam therapy as an alternative to Cerrobend cutouts with applicators once the prototype's leaves are motorized. Our eMLC data showed interleaf electron leakage at 15 MeV to be less than 0.1% based on a 0.0025 cm manufacturing tolerance, and lateral electron leakage at 5 and 15 MeV to be less than 2%. X-ray leakage through the leaves was 1.6% at 15 MeV. Our data showed that beam penumbra was independent of direction and leaf position. The dosimetric properties of square fields formed by the eMLC were very consistent with those formed by Cerrobend inserts in the 20 X 20 cm 2 applicator. Output factors exhibited similar fieldsize dependence. Airgap factors exhibited almost identical field-size dependence at two SSDs (105 and 110 cm), consistent with the common assumption that airgap factors are applicator independent. Percent depth-dose curves were similar, but showed variations up to 3% in the buildup region. The pencil-beam algorithm (PBA) fit measured data from the eMLC and applicator-cutout systems equally well, and the resulting two-dimensional (2-D) dose distributions, as predicted by the PBA, agreed well at common airgap, distance. Simulating patient setups for breast and head and neck treatments showed that almost all fields could be treated using similar SSDs as when using applicators, although head and neck treatments require placing the patient's head on a head-holder treatment table extension. The results of this work confirmed our design goals and support the potential use of the eMLC design in the clinical setting. The eMLC should allow the same treatments as are typically delivered with the electron applicator-cutout system currently used for fixedbeam therapy. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:443 / 462
页数:20
相关论文
共 36 条
  • [1] DESIGN PRINCIPLES AND CLINICAL POSSIBILITIES WITH A NEW GENERATION OF RADIATION-THERAPY EQUIPMENT - A REVIEW
    BRAHME, A
    [J]. ACTA ONCOLOGICA, 1987, 26 (06) : 403 - 412
  • [2] Clinical considerations of Monte Carlo for electron radiotherapy treatment planning
    Faddegon, B
    Balogh, J
    Mackenzie, R
    Scora, D
    [J]. RADIATION PHYSICS AND CHEMISTRY, 1998, 53 (03) : 217 - 227
  • [3] Faddegon BA, 2002, MED PHYS, V29, P1285
  • [4] Hogstrom K., 2003, INTENSITY MODULATED, P749
  • [5] DESIGN OF METALLIC ELECTRON-BEAM CONES FOR AN INTRAOPERATIVE THERAPY LINEAR-ACCELERATOR
    HOGSTROM, KR
    BOYER, AL
    SHIU, AS
    OCHRAN, TG
    KIRSNER, SM
    KRISPEL, F
    RICH, T
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (05): : 1223 - 1232
  • [6] DOSIMETRIC EVALUATION OF A PENCIL-BEAM ALGORITHM FOR ELECTRONS EMPLOYING A TWO-DIMENSIONAL HETEROGENEITY CORRECTION
    HOGSTROM, KR
    MILLS, MD
    MEYER, JA
    PALTA, JR
    MELLENBERG, DE
    MEOZ, RT
    FIELDS, RS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1984, 10 (04): : 561 - 569
  • [7] HOGSTROM KR, 1985, P 1985 MEV US C, P251
  • [8] HOGSTROM KR, 1987, RAD ONCOLOGY PHYSICS, P265
  • [9] Radiotherapy of breast cancer after breast-conserving surgery:: an improved technique using mixed electron-photon beams with a multileaf collimator
    Jansson, T
    Lindman, H
    Nygård, K
    Dahlgren, CV
    Montelius, A
    Öberg-Kreuger, C
    Asplund, S
    Bergh, J
    [J]. RADIOTHERAPY AND ONCOLOGY, 1998, 46 (01) : 83 - 89
  • [10] MATCHING OF ELECTRON AND PHOTON BEAMS WITH A MULTI-LEAF COLLIMATOR
    KARLSSON, M
    ZACKRISSON, B
    [J]. RADIOTHERAPY AND ONCOLOGY, 1993, 29 (03) : 317 - 326