Failure analysis of cemented metal-on-metal total hip replacements from a single centre cohort

被引:16
作者
Bryant, Michael [1 ]
Ward, Michael [2 ]
Farrar, Richard [3 ]
Freeman, Robert [3 ]
Brummitt, Ken [3 ]
Nolan, John [4 ]
Neville, Anne [1 ]
机构
[1] Univ Leeds, Inst Engn Thermofluids Surfaces & Interfaces, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Inst Mat Res, Leeds LS2 9JT, W Yorkshire, England
[3] DePuy Int, Leeds, W Yorkshire, England
[4] Norfolk & Norwich Teaching Hosp, Norwich, Norfolk, England
关键词
Fretting-corrosion; Femoral stems; Retrieval study; Bone cement; Case study; TRIBOCHEMICAL REACTION; FEMORAL COMPONENTS; CORROSION; COCRMO; WEAR; BIOTRIBOCORROSION; ARTHROPLASTIES; ALLOY;
D O I
10.1016/j.wear.2012.12.013
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
This paper presents the failure analysis of cemented metal on metal (MoM) total hip replacements (THR) from a single centre cohort. In this paper all retrieved femoral stems were implanted and revised at the Norfolk and Norwich University hospital by three specialist hip surgeons. All patients were subjected to the same combination of metallic implants, operative and follow up procedure. Between 1997 and 2008, 652 Ultima TPS (TM) femoral stems with the Ultima MoM articulation (DePuy International, Leeds) were implanted. At 2-9 years follow up, 90 revisions had been performed. Upon removal extensive fretting-corrosion of the cemented portions of the femoral stems was seen, resulting in high metal ion levels. In order to fully understand the factors resulting in failure of these cemented MoM THR optical light microscopy, Scanning Electron Microscopy with integrated Energy Dispersive X-ray spectroscopy (SEM/EDX) and X-ray Photo-electron spectroscopy (XPS) were utilised in order to assess the surface morphology and chemistry of retrieved MoM THR and surface chemistry of the films found on the surface. Gross slip, plastic deformation and directionality of the surface were extensively seen on the proximal surfaces of the retrievals. A more corrosive phenomenon was in the distal regions of the stem, demonstrating a seemingly inter-granular attack. SEM/EDX analysis demonstrated that black films present commonly seen on the highly worn regions of the cemented portions of the femoral stem were a complex mixture of chromium oxide, protein and amorphous carbon. XPS demonstrated films were Cr, O rich, mixed with organic material. Crown Copyright (C) 2012 Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:226 / 233
页数:8
相关论文
共 22 条
[1]  
[Anonymous], 1995, Handbook of X-ray Photoelectron Spectroscopy. A Reference Book of Standard Spectra for Identification and Interpretation of XPS Data
[2]   High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS [J].
Bolland, B. J. R. F. ;
Culliford, D. J. ;
Langton, D. J. ;
Millington, J. P. S. ;
Arden, N. K. ;
Latham, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :608-615
[3]   Clinical and radiological results of the collarless polished tapered stem at 15 years follow-up [J].
Burston, B. J. ;
Barnett, A. J. ;
Amirfeyz, R. ;
Yates, P. J. ;
Bannister, G. C. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (07) :889-894
[4]   Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs [J].
Donell, S. T. ;
Darrah, C. ;
Nolan, J. F. ;
Wimhurst, J. ;
Toms, A. ;
Barker, T. H. W. ;
Case, C. P. ;
Tucker, J. K. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (11) :1501-1508
[5]  
Fairley Neal., 2009, CasaXPS Manual
[6]  
FOWLER JL, 1988, ORTHOP CLIN N AM, V19, P477
[7]  
GIE GA, 1994, J BONE JOINT SURG BR, V76B, P506, DOI 10.1302/0301-620X.76B3.8175874
[8]   Passive and transpassive behaviour of CoCrMo in simulated biological solutions [J].
Hodgson, AWE ;
Kurz, S ;
Virtanen, S ;
Fervel, V ;
Olsson, COA ;
Mischler, S .
ELECTROCHIMICA ACTA, 2004, 49 (13) :2167-2178
[9]   In vivo surface wear mechanisms of femoral components of cemented total hip arthroplasties -: The influence of wear mechanism on clinical outcome [J].
Howell, JR ;
Blunt, LA ;
Doyle, C ;
Hooper, RM ;
Lee, AJC ;
Ling, RSM .
JOURNAL OF ARTHROPLASTY, 2004, 19 (01) :88-101
[10]   THE INITIATION OF FAILURE IN CEMENTED FEMORAL COMPONENTS OF HIP ARTHROPLASTIES [J].
JASTY, M ;
MALONEY, WJ ;
BRAGDON, CR ;
OCONNOR, DO ;
HAIRE, T ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (04) :551-558