Urease was doped into polypyrrole (PPy) film when pyrrole (Py) was polymerized by cyclic voltammogram. By Combination of PPy modified urease electrode and CO2 gas-sensitive electrode a bioelectrochemical, sensor was formed, which responses to urea. Factors affecting the activity of urease in PPy film were studied and activation energy of the ferment was determined. It was found that electrode potential is in direct proportion to the logarithm of the concentration of urea within the concentration range of 5.0 x 10(-5)similar to 1.0 x 10(-2) mol/L.
引用
收藏
页码:417 / 421
页数:5
相关论文
共 12 条
[1]   A QUARTZ MICROBALANCE STUDY OF THE ELECTROSYNTHESIS OF POLYPYRROLE [J].
BAKER, CK ;
REYNOLDS, JR .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1988, 251 (02) :307-322
[2]   POLYPYRROLE AND POLY N-METHYLPYRROLE - AN ELECTROCHEMICAL STUDY IN AN AQUEOUS-MEDIUM [J].
GENIES, EM ;
SYED, AA .
SYNTHETIC METALS, 1984, 10 (01) :21-30
[3]  
HE SW, 1994, J HUNAN U, V21, P28
[4]   CONFERMENT OF CHOLESTEROL SENSITIVITY ON POLYPYRROLE FILMS BY IMMOBILIZATION OF CHOLESTEROL OXIDASE AND FERROCENECARBOXYLATE IONS [J].
KAJIYA, Y ;
TSUDA, R ;
YONEYAMA, H .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1991, 301 (1-2) :155-164
[5]  
LOURSE LD, 1989, SYNTHETIC MET, pC761
[6]   CONDUCTING POLYMER GAS SENSORS [J].
MIASIK, JJ ;
HOOPER, A ;
TOFIELD, BC .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS I, 1986, 82 :1117-&
[7]  
MO SL, 1993, ACTA CHIMICA SINCICA, V5, P632
[8]  
NIE JC, 1988, BRIEF INDUCTION BIOC, P97
[9]   EVALUATION OF POLYPYRROLE GLUCOSE-OXIDASE ELECTRODES IN FLOW-INJECTION SYSTEMS FOR SUCROSE DETERMINATION [J].
SCHUHMANN, W ;
KITTSTEINEREBERLE, R .
BIOSENSORS & BIOELECTRONICS, 1991, 6 (03) :263-273
[10]   PEROXIDASE-INCORPORATED POLYPYRROLE MEMBRANE ELECTRODES [J].
TATSUMA, T ;
GONDAIRA, M ;
WATANABE, T .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1183-1187