Effect of microthread presence and restoration design (screw versus cemented) in dental implant reliability and failure modes

被引:19
作者
Almeida, Erika O. [3 ]
Freitas Junior, Amilcar C. [4 ]
Bonfante, Estevam A. [5 ]
Rocha, Eduardo Passos [3 ]
Silva, Nelson R. F. A. [6 ]
Coelho, Paulo G. [1 ,2 ]
机构
[1] NYU, Coll Dent, Dept Biomat & Biomimet, New York, NY USA
[2] NYU, Coll Dent, Dept Periodontol & Implant Dent, New York, NY USA
[3] Sao Paulo State Univ, UNESP, Dept Dent Mat & Prosthodont, Aracatuba Sch Dent, Sao Paulo, Brazil
[4] Univ Potiguar, Sch Dent UnP, Postgrad Program Dent, Natal, RN, Brazil
[5] UNIGRANRIO Univ, Postgrad Program Dent, Duque De Caxias, RJ, Brazil
[6] NYU, Coll Dent, Dept Prosthodont, New York, NY USA
关键词
dental implant; fractography; reliability; step-stress accelerated life-testing; Weibull; MARGINAL BONE LEVEL; RADIOGRAPHIC EVALUATION; STRESS LEVELS; ELEMENT; CONNECTION; SURVIVAL; THREAD; NECK;
D O I
10.1111/j.1600-0501.2012.02437.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives This study evaluated the reliability and failure modes of implants with a microthreaded or smooth design at the crestal region, restored with screwed or cemented crowns. The postulated null hypothesis was that the presence of microthreads in the implant cervical region would not result in different reliability and strength to failure than smooth design, regardless of fixation method, when subjected to step-stress accelerated life-testing (SSALT) in water. Materials and methods Eighty four dental implants (3.3 x 10 mm) were divided into four groups (n = 21) according to implant macrogeometric design at the crestal region and crown fixation method: Microthreads Screwed (MS); Smooth Screwed (SS); Microthreads Cemented (MC), and Smooth Cemented (SC). The abutments were torqued to the implants and standardized maxillary central incisor metallic crowns were cemented (MC, SC) or screwed (MS, SS) and subjected to SSALT in water. The probability of failure versus cycles (90% two-sided confidence intervals) was calculated and plotted using a power law relationship for damage accumulation. Reliability for a mission of 50,000 cycles at 150 N (90% 2-sided confidence intervals) was calculated. Differences between final failure loads during fatigue for each group were assessed by Kruskal-Wallis along with Benferroni's post hoc tests. Polarized-light and scanning electron microscopes were used for failure analyses. Results The Beta (beta) value (confidence interval range) derived from use level probability Weibull calculation of 1.30 (0.76-2.22), 1.17 (0.70-1.96), 1.12 (0.71-1.76), and 0.52 (0.30-0.89) for groups MC, SC, MS, and SS respectively, indicated that fatigue was an accelerating factor for all groups, except for SS. The calculated reliability was higher for SC (99%) compared to MC (87%). No difference was observed between screwed restorations (MS 29%, SS 43%). Failure involved abutment screw fracture for all groups. The cemented groups (MC, SC) presented more abutment and implant fractures. Significantly higher load to fracture values were observed for SC and MC relative to MS and SS (P < 0.001). Conclusion Since reliability and strength to failure was higher for SC than for MC, our postulated null hypothesis was rejected.
引用
收藏
页码:191 / 196
页数:6
相关论文
共 32 条
[1]   Tissue characteristics at microthreaded implants: An experimental study in dogs [J].
Abrahamsson, Ingemar ;
Berglundh, Tord .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2006, 8 (03) :107-113
[2]   The effect of thread pattern upon implant osseointegration [J].
Abuhussein, Heba ;
Pagni, Giorgio ;
Rebaudi, Alberto ;
Wang, Hom-Lay .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (02) :129-136
[3]   A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW [J].
ADELL, R ;
LEKHOLM, U ;
ROCKLER, B ;
BRANEMARK, PI .
INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06) :387-416
[4]  
[Anonymous], ASM INT
[5]  
[Anonymous], WEIB DISTR BET
[6]  
[Anonymous], 2006, The New Weibull Handbook
[7]   Fatigue testing of two porcelain-zirconia all-ceramic crown systems [J].
Coelho, P. G. ;
Silva, N. R. ;
Bonfante, E. A. ;
Guess, P. C. ;
Rekow, E. D. ;
Thompson, V. P. .
DENTAL MATERIALS, 2009, 25 (09) :1122-1127
[8]   Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crowns [J].
Freitas, Amilcar C., Jr. ;
Bonfante, Estevam A. ;
Rocha, Eduardo P. ;
Silva, Nelson R. F. A. ;
Marotta, Leonard ;
Coelho, Paulo G. .
EUROPEAN JOURNAL OF ORAL SCIENCES, 2011, 119 (04) :323-330
[9]  
Gunne J, 1994, Int J Prosthodont, V7, P143
[10]   The implant thread as a retention element in cortical bone: the effect of thread size and thread profile: a finite element study [J].
Hansson, S ;
Werke, M .
JOURNAL OF BIOMECHANICS, 2003, 36 (09) :1247-1258