Automatic Segmentation of the Central Epithelium Imaged With Three Optical Coherence Tomography Devices

被引:18
作者
Ge, Lili [1 ]
Shen, Meixiao [1 ,2 ]
Tao, Aizhu [1 ]
Wang, Jianhua [2 ]
Dou, Guopeng [1 ]
Lu, Fan [1 ]
机构
[1] Wenzhou Med Coll, Sch Ophthalmol & Optometry, Wenzhou 325027, Zhejiang, Peoples R China
[2] Univ Miami, Bascom Palmer Eye Inst, Dept Ophthalmol, Miami, FL USA
来源
EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE | 2012年 / 38卷 / 03期
关键词
Automatic segmentation; Cornea; Optical coherence tomography; TOPOGRAPHICAL THICKNESS; TOTAL CORNEA; REPEATABILITY; DOMAIN; KERATOCONUS; REPRODUCIBILITY; DIAGNOSIS; WEAR;
D O I
10.1097/ICL.0b013e3182499b64
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The aim of this study was to investigate the feasibility of automatic segmentation of the central corneal thickness (CCT) and epithelial thickness (ET) of the human cornea obtained with different spectral domain optical coherence tomography (OCT) instruments. Methods: Ten left eyes from 10 healthy subjects with a mean age of 22.5 +/- 1.5 years participated in this study. A custom-built ultra-high resolution OCT (UHR-OCT) with a 3-mu m axial resolution, ultralong scan depth OCT (UL-OCT) with a 7.5-mu m resolution, and commercial RTVue OCT with a 5-mu m resolution were used to image the cornea. An automated segmentation algorithm was developed to process the OCT images and yield the CCT and ET. The measurement was verified by a manual measurement. Results: The automatic algorithm successfully processed the central thickness of the cornea and corneal epithelium for all images. The average CCT obtained by the automatic segmentation algorithm was 528.1 +/- 22.4 mu m, 526.1 +/- 23.4 mu m, and 525.2 +/- 23.7 mu m for UHR-OCT, UL-OCT, and RTVue, respectively. The average ET was 53.2 +/- 2.0 mu m, 54.1 +/- 3.0 mu m, and 52.1 +/- 2.5 mu m for UHR-OCT, UL-OCT, and RTVue, respectively. These measurements were in agreement with those of the manual method for the CCT (all r>0.997, P<0.05) and for the ET (all r>0.71, P<0.05). Conclusions: The algorithm seemed to be feasible for automatically segmenting the CCT and ET in OCT images using these tested OCT devices. The segmented results were equivalent to that obtained with the manual method.
引用
收藏
页码:150 / 157
页数:8
相关论文
共 31 条
[1]   Diagnosis and Management of Conjunctival and Corneal Intraepithelial Neoplasia Using Ultra High-Resolution Optical Coherence Tomography [J].
Abou Shousha, Mohamed ;
Karp, Carol L. ;
Perez, Victor L. ;
Hoffmann, Rodrigo ;
Ventura, Roberta ;
Chang, Victoria ;
Dubovy, Sander R. ;
Wang, Jianhua .
OPHTHALMOLOGY, 2011, 118 (08) :1531-1537
[2]   Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry [J].
Chen, Shihao ;
Huang, Jinhai ;
Wen, Daizong ;
Chen, Wuhe ;
Huang, David ;
Wang, Qinmei .
ACTA OPHTHALMOLOGICA, 2012, 90 (05) :449-455
[3]   Use of Anterior Segment Optical Coherence Tomography to Study Corneal Changes After Collagen Cross-linking [J].
Doors, Muriel ;
Tahzib, Nayyirih G. ;
Eggink, Fred A. ;
Berendschot, Tos T. J. M. ;
Webers, Carroll A. B. ;
Nuijts, Rudy M. M. A. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (06) :844-851
[4]   Assessment of central corneal thickness using optical coherence tomography [J].
Fishman, GR ;
Pons, ME ;
Seedor, JA ;
Liebmann, JM ;
Ritch, R .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04) :707-711
[5]   Corneal and epithelial thickness in keratoconus: A comparison of ultrasonic pachymetry Orbscan II, and optical coherence tomography [J].
Haque, S ;
Simpson, T ;
Jones, L .
JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (05) :486-493
[6]  
Haque Sameena, 2004, Eye Contact Lens, V30, P189, DOI 10.1097/01.ICL.0000140223.60892.16
[7]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[8]   Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography [J].
Huang, Jehn-Yu ;
Pekmezci, Melike ;
Yaplee, Stephanie ;
Lin, Shan .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (11) :1647-1656
[9]   Swelling of the Human Cornea Revealed by High-Speed, Ultrahigh-Resolution Optical Coherence Tomography [J].
Hutchings, Natalie ;
Simpson, Trefford L. ;
Hyun, Chulho ;
Moayed, Alireza A. ;
Hariri, Sepideh ;
Sorbara, Luigina ;
Bizheva, Kostadinka .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (09) :4579-4584
[10]   MICROMETER-SCALE RESOLUTION IMAGING OF THE ANTERIOR EYE IN-VIVO WITH OPTICAL COHERENCE TOMOGRAPHY [J].
IZATT, JA ;
HEE, MR ;
SWANSON, EA ;
LIN, CP ;
HUANG, D ;
SCHUMAN, JS ;
PULIAFITO, CA ;
FUJIMOTO, JG .
ARCHIVES OF OPHTHALMOLOGY, 1994, 112 (12) :1584-1589