Pulsatile control of rotary blood pumps: Does the modulation waveform matter?

被引:58
|
作者
Pirbodaghi, Tohid [1 ]
Axiak, Shannon [2 ]
Weber, Alberto [3 ]
Gempp, Thomas [4 ]
Vandenberghe, Stijn [1 ]
机构
[1] Univ Bern, ARTORG Cardiovasc Engn, Bern, Switzerland
[2] Univ Bern, Dept Vet Anesthesia, Bern, Switzerland
[3] Univ Hosp Bern, Inselspital, Dept Cardiac Surg, CH-3010 Bern, Switzerland
[4] Levitronix GmbH, Zurich, Switzerland
来源
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY | 2012年 / 144卷 / 04期
基金
瑞士国家科学基金会;
关键词
VENTRICULAR ASSIST DEVICE; CARDIOPULMONARY BYPASS; SUPPORT; FLOW; NONPULSATILE; PERFUSION; RECOVERY; FAILURE;
D O I
10.1016/j.jtcvs.2012.02.015
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Mechanical support of a failing heart is typically performed with rotary blood pumps running at constant speed, which results in a limited control on cardiac workload and nonpulsatile hemodynamics. A potential solution to overcome these limitations is to modulate the pump speed to create pulses. This study aims at developing a pulsatile control algorithm for rotary pumps, while investigating its effect on left ventricle unloading and the hemodynamics. Methods: The CentriMag (Levitronix GmbH, Zurich, Switzerland) rotary blood pump was implanted in 5 sheep and cannulated from the ventricular apex to the descending aorta. A modified controller was connected to the pump yielding direct speed control via analog voltage. Pump speed modulation patterns, including sine, saw tooth, triangle, and square waveforms with 2 different phase shifts, were synchronized with heartbeat. Various hemodynamic parameters, such as left ventricular pressure and volume, coronary flow, and arterial pressure, were analyzed to examine the influence of pump support. Results: The pump speed modulation significantly affected left ventricular pressure and volume and arterial pressure, whereas coronary flow was not influenced by pump support mode. Stroke work in the pulsatile modes varied from 69% to 91% of baseline value and from 74% to 96% of constant speed value. Consequently, cardiac workload can be adjusted to provide relaxation, which may lead to myocardial recovery. Conclusions: A synchronized pulsing rotary blood pump offers a simple and powerful control modality for heart unloading. This technique provides pulsatile hemodynamics, which is more physiologic than continuous blood flow and may be useful for perfusion of the other organs. (J Thorac Cardiovasc Surg 2012; 144: 970-7)
引用
收藏
页码:970 / 977
页数:8
相关论文
共 50 条
  • [1] Generation of Pulsatile Flow in Rotary Pumps
    Kuleshov, A. P.
    Buchnev, A. S.
    Esipova, O. Yu
    Itkin, G. P.
    Drobyshev, A. A.
    BIOMEDICAL ENGINEERING-MEDITSINSKAYA TEKNIKA, 2022, 56 (03): : 168 - 171
  • [2] Feasibility of Pump Speed Modulation for Restoring Vascular Pulsatility with Rotary Blood Pumps
    Ising, Mickey S.
    Sobieski, Michael A.
    Slaughter, Mark S.
    Koenig, Steven C.
    Giridharan, Guruprasad A.
    ASAIO JOURNAL, 2015, 61 (05) : 526 - 532
  • [3] Control strategy for rotary blood pumps
    Ohuchi, K
    Kikugawa, D
    Takahashi, K
    Uemura, M
    Nakamura, M
    Murakami, T
    Sakamoto, T
    Takatani, S
    ARTIFICIAL ORGANS, 2001, 25 (05) : 366 - 370
  • [4] Minimizing left ventricular stroke work with iterative learning flow profile control of rotary blood pumps
    Rueschen, Daniel
    Prochazka, Frederik
    Amacher, Raffael
    Bergmann, Lukas
    Leonhardt, Steffen
    Walter, Marian
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2017, 31 : 444 - 451
  • [5] Asymmetric speed modulation of a rotary blood pump affects ventricular unloading
    Pirbodaghi, Tohid
    Weber, Alberto
    Axiak, Shannon
    Carrel, Thierry
    Vandenberghe, Stijn
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2013, 43 (02) : 383 - 388
  • [6] Power Consumption of Rotary Blood Pumps: Pulsatile Versus Constant-Speed Mode
    Pirbodaghi, Tohid
    Cotter, Chris
    Bourque, Kevin
    ARTIFICIAL ORGANS, 2014, 38 (12) : 1024 - 1028
  • [7] Pulse-Pressure-Enhancing Controller for Better Physiologic Perfusion of Rotary Blood Pumps Based on Speed Modulation
    Huang, Feng
    Ruan, Xiaodong
    Fu, Xin
    ASAIO JOURNAL, 2014, 60 (03) : 269 - 279
  • [8] Rotary piston blood pumps: past developments and future potential of a unique pump type
    Wappenschmidt, Johannes
    Autschbach, Ruediger
    Steinseifer, Ulrich
    Schmitz-Rode, Thomas
    Margreiter, Raimund
    Klima, Guenter
    Goetzenich, Andreas
    EXPERT REVIEW OF MEDICAL DEVICES, 2016, 13 (08) : 759 - 771
  • [9] Hydraulic Characterization of Implantable Rotary Blood Pumps
    Boes, Stefan
    Thamsen, Bente
    Haas, Mattia
    Daners, Marianne Schmid
    Meboldt, Mirko
    Granegger, Marcus
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (06) : 1618 - 1627
  • [10] A suction index based control system for rotary blood pumps
    Liang, Lixue
    Meki, Moustafa
    Wang, Weibin
    Sethu, Palaniappan
    El-Baz, Ayman
    Giridharan, Guruprasad A.
    Wang, Yu
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2020, 62