Computer-assisted knee arthroplasty versus a conventional jig-based technique -: A randomised, prospective trial

被引:347
作者
Chauhan, SK [1 ]
Scott, RG [1 ]
Breidahl, W [1 ]
Beaver, RJ [1 ]
机构
[1] Royal Perth Hosp, Perth, WA 6008, Australia
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2004年 / 86B卷 / 03期
关键词
D O I
10.1302/0301-620X.86B3.14643
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We have compared a new technique of computer-assisted knee arthroplasty with the current conventional jig-based technique in 70 patients randomly allocated to receive either of the methods. Post-operative CT was performed according to the Perth CT Knee Arthroplasty protocol and pre- and post-operative Maquet views of the limb were taken. Intra-operative and peri-operative morbidity data were collected and blood loss measured. Post-operative CT showed a significant improvement in the alignment of the components using computer-assisted surgery in regard to femoral varus/valgus (p = 0.032), femoral rotation (p = 0.001), tibial varus/valgus (p = 0.047) tibial posterior slope (p = 0.0001), tibial rotation (p 0.011) and femorotibial mismatch (p = 0.037). Standing alignment was also improved (p 0.004) and blood loss was less (p = 0.0001). Computer-assisted surgery took longer with a mean increase of 13 minutes (p = 0.0001).
引用
收藏
页码:372 / 377
页数:6
相关论文
共 17 条