Influence of Implant Design and Under-Preparation of the Implant Site on Implant Primary Stability. An In Vitro Study

被引:24
|
作者
Herrero-Climent, Mariano [1 ]
Lemos, Bernardo Ferreira [1 ,2 ]
Herrero-Climent, Federico [3 ]
Falcao, Carlos [1 ,2 ]
Oliveira, Helder [2 ]
Herrera, Manuela [4 ]
Javier Gil, Francisco [5 ]
Rios-Carrasco, Blanca [6 ]
Rios-Santos, Jose-Vicente [6 ]
机构
[1] Porto Dent Inst, P-4150518 Porto, Portugal
[2] Fernando Pessoa Univ, Fac Hlth Sci, P-4249004 Porto, Portugal
[3] Univ Complutense Madrid, Dept Periodontol, Madrid 28040, Spain
[4] Univ Seville, Dept Stomatol, Seville 41009, Spain
[5] Univ Int Catalunya, Fac Dent, Barcelona 08017, Spain
[6] Univ Seville, Dept Periodontol, Seville 41009, Spain
关键词
dental implant; tapered; under-preparation; implant preparation; implant stability; insertion torque; ISQ; RFA; RESONANCE FREQUENCY-ANALYSIS; LOW-DENSITY BONE; DENTAL IMPLANTS; SURGICAL TECHNIQUE; SURFACE-ROUGHNESS; STEPPED OSTEOTOMY; INSERTION TORQUE; HUMAN CADAVERS; OSSEOINTEGRATION; RELIABILITY;
D O I
10.3390/ijerph17124436
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The aim of this study was to evaluate the effects of different implant sites an under-preparation sequence associated with two different implant designs on implant primary stability measured by two parameters: insertion torque (IT) and implant stability quotient (ISQ). It used two different implants: one cylindrical as a control and another one with a tapered design. The implants were inserted in type III fresh humid bovine bone and four drilling sequences were used: one control, the one proposed by the implant company (P1), and three different undersized (P2, P3 and P4). P2 was the same as P1 without the cortical drill, P3 was without the last pilot drill and P4 was without both of them. The sample size wasn= 40 for each of the eight groups. Final IT was measured with a torquemeter and the ISQ was measured with Penguin resonance frequency analysis. Results showed that both ISQ and IT have a tendency to increase as the preparation technique reduces the implant site diameter when compared with the standard preparation, P1. The preparations without cortical drill, P2 and P4, showed the best results when compared with the ones with a cortical drill. Tapered implants always showed higher or the same ISQ and IT values when compared with the cylindrical implants. Giving the limitations of this study, it can be concluded that reducing implant preparation can increase IT and ISQ values. Removing the cortical drill and the use of a tapered design implant are also effective methods of increasing primary implant stability.
引用
收藏
页码:1 / 13
页数:13
相关论文
共 50 条
  • [1] Investigation of the Influence of Roughness and Dental Implant Design on Primary Stability via Analysis of Insertion Torque and Implant Stability Quotient: An In Vitro Study
    Romero, Marta
    Herrero-Climent, Mariano
    Rios-Carrasco, Blanca
    Brizuela, Aritza
    Romero, Manuel Maria
    Gil, Javier
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (13)
  • [2] Enhancing primary implant stability by undersizing implant site preparation: A human cadaver study
    Turkyilmaz, I
    Tozum, T. F.
    JOURNAL OF STOMATOLOGY ORAL AND MAXILLOFACIAL SURGERY, 2020, 121 (01) : 58 - 62
  • [3] Efficacy of biologically guided implant site preparation to obtain adequate primary implant stability
    Anitua, Eduardo
    Alkhraisat, Mohammad Hamdan
    Pinas, Laura
    Orive, Gorka
    ANNALS OF ANATOMY-ANATOMISCHER ANZEIGER, 2015, 199 : 9 - 15
  • [4] Primary Implant Stability Analysis of Different Dental Implant Connections and Designs-An In Vitro Comparative Study
    Raz, Perry
    Meir, Haya
    Levartovsky, Shifra
    Sebaoun, Alon
    Beitlitum, Ilan
    MATERIALS, 2022, 15 (09)
  • [5] Influence of Implant Macro-Design, -Length, and -Diameter on Primary Implant Stability Depending on Different Bone Qualities Using Standard Drilling Protocols-An In Vitro Analysis
    Stoilov, Milan
    Shafaghi, Ramin
    Stark, Helmut
    Marder, Michael
    Kraus, Dominik
    Enkling, Norbert
    JOURNAL OF FUNCTIONAL BIOMATERIALS, 2023, 14 (09)
  • [6] Effect of Implant Macro-Design and Magnetodynamic Surgical Preparation on Primary Implant Stability: An In Vitro Investigation
    Antonelli, Alessandro
    Barone, Selene
    Attanasio, Ferdinando
    Salviati, Marianna
    Cerra, Maria Giulia
    Calabria, Elena
    Bennardo, Francesco
    Giudice, Amerigo
    DENTISTRY JOURNAL, 2023, 11 (10)
  • [7] Validation of the Implant Stability Test for Implant Provisional Crowns: An In Vitro Study
    Okuhama, Yurie
    Nagata, Koudai
    Nakashizu, Toshifumi
    Wakamori, Kana
    Okubo, Manabu
    Tsuruoka, Hayato
    Kawana, Hiromasa
    APPLIED SCIENCES-BASEL, 2023, 13 (06):
  • [8] Analysis of the relationship between a modified method for implant site preparation and primary implant stability: A pilot study
    Unsal, Gokce Soganci
    Turkyilmaz, Ilser
    JOURNAL OF DENTAL SCIENCES, 2020, 15 (02) : 234 - 235
  • [9] Effect of bone quality, implant design, and surgical technique on primary implant stability
    Karl, Matthias
    Grobecker-Karl, Tanja
    QUINTESSENCE INTERNATIONAL, 2018, 49 (03): : 189 - 198
  • [10] Reliability and Correlation of Different Devices for the Evaluation of Primary Implant Stability: An In Vitro Study
    Raz, Perry
    Meir, Haya
    Levartovsky, Shifra
    Peleg, Maia
    Sebaoun, Alon
    Beitlitum, Ilan
    MATERIALS, 2021, 14 (19)