An optical fiber Bragg grating force sensor for monitoring sub-bandage pressure during compression therapy

被引:39
作者
Wang, David Hsiao-Chuan [1 ]
Blenman, Neil [1 ]
Maunder, Simon [1 ]
Patton, Vicki [2 ]
Arkwright, John [1 ]
机构
[1] CSIRO Mat Sci & Engn, Lindfield, NSW 2070, Australia
[2] Univ New S Wales, St George Hosp, Dept Anorectal Physiol, Kogarah, NSW, Australia
来源
OPTICS EXPRESS | 2013年 / 21卷 / 17期
关键词
INTERFACE PRESSURE; VENOUS DISEASE; IN-VIVO; DEVICE;
D O I
10.1364/OE.21.019799
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Graduated compression bandaging of the lower limbs is the primary therapy for venous leg ulcers with its efficacy believed to be predominantly dependent on the amount and the distribution of the compressive pressure applied. There has been on-going demand for an ideal sensor to facilitate in-vivo monitoring of the sub-bandage pressure. Several methods and devices have been reported but each has its limitations, such as bulkiness, low tolerance to movement, susceptible to thermal noise and single point sensing. An optical fiber force sensor is demonstrated, consisting of two arrays of fiber Bragg grating (FBG) entwined in a double helix form and packaged with contact-force sensitivity. This sensor array has inherent temperature immunity and is capable of real-time, distributed sensing of sub-bandage pressure. The calibration results of the sensor array, as well as the validation human trial results, are presented. (C) 2013 Optical Society of America
引用
收藏
页码:19799 / 19807
页数:9
相关论文
共 14 条
  • [1] In-vivo demonstration of a high resolution optical fiber manometry catheter for diagnosis of gastrointestinal motility disorders
    Arkwright, J. W.
    Blenman, N. G.
    Underhill, I. D.
    Maunder, S. A.
    Szczesniak, M. M.
    Dinning, P. G.
    Cook, I. J.
    [J]. OPTICS EXPRESS, 2009, 17 (06): : 4500 - 4508
  • [2] DEVICE FOR MEASURING SOFT-TISSUE INTERFACE PRESSURES
    BARBENEL, JC
    SOCKALINGHAM, S
    [J]. JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (06): : 519 - 522
  • [3] Mechanisms of disease:: Chronic venous disease
    Bergan, John J.
    Schmid-Schoenbein, Geert W.
    Smith, Philip D. Coleridge
    Nicolaides, Andrew N.
    Boisseau, Michel R.
    Eklof, Bo
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2006, 355 (05) : 488 - 498
  • [4] Evaluation of a sensor for low interface pressure applications
    Ferguson-Pell, M
    Hagisawa, S
    Bain, D
    [J]. MEDICAL ENGINEERING & PHYSICS, 2000, 22 (09) : 657 - 663
  • [5] Instruction of compression therapy by means of interface pressure measurement
    Hafner, J
    Lüthi, W
    Hänssle, H
    Kammerlander, G
    Burg, G
    [J]. DERMATOLOGIC SURGERY, 2000, 26 (05) : 481 - 486
  • [6] ABC of arterial and venous disease - Ulcerated lower limb
    London, NJM
    Donnelly, R
    [J]. BRITISH MEDICAL JOURNAL, 2000, 320 (7249) : 1589 - 1591
  • [7] Moore Z, 2002, J Wound Care, V11, P265
  • [8] Measurement of lower leg compression in vivo:: Recommendations for the performance of measurements of interface pressure and stiffness
    Partsch, H
    Clark, M
    Bassez, S
    Benigni, JP
    Becker, F
    Blazek, V
    Caprini, J
    Cornu-Thénard, A
    Hafner, J
    Flour, M
    Jünger, M
    Moffatt, C
    Neumann, M
    [J]. DERMATOLOGIC SURGERY, 2006, 32 (02) : 224 - 233
  • [9] HOW LONG DO COMPRESSION BANDAGES MAINTAIN THEIR PRESSURE DURING AMBULATORY TREATMENT OF VARICOSE-VEINS
    RAJ, TB
    GODDARD, M
    MAKIN, GS
    [J]. BRITISH JOURNAL OF SURGERY, 1980, 67 (02) : 122 - 124
  • [10] In-fibre Bragg grating sensors
    Rao, YJ