microwave plasma torch discharge;
hollow cathode lamp;
high current microsecond pulsed power supply;
ionic fluorescence spectrometry;
atomic fluorescence spectrometry;
europium;
D O I:
暂无
中图分类号:
O433 [光谱学];
学科分类号:
0703 ;
070302 ;
摘要:
Eu atomic and ionic fluorescence spectrometry in microwave plasma torch discharge excited by high current microsecond pulsed hollow cathode lamp (HCMP HCL-MPT AFS/IFS) was studied. Operating conditions were optimized. The best detection limits for AFS and IES obtained with a desolvated ultrasonic nebulization system were 42. 0 ng/mL for Eu I 462.7 nm and 21.8 ng/mL for Eu II 381.97 nm, respectively, both were better than those given by the instruction manual of a Baird ICP AFS-2000 spectrometer using pneumatic concentric nebulizer with desolvation for AFS,but were significantly higher than those obtained by using the Baird spectrometer with a mini-monochromator and a ultrasonic nebulzer system.