Analysis of Explanted Magnetically Controlled Growing Rods From Seven UK Spinal Centers

被引:55
作者
Joyce, Thomas J. [1 ]
Smith, Simon L. [1 ]
Rushton, Paul R. P. [2 ]
Bowey, Andrew J. [2 ]
Gibson, Michael J. [2 ]
机构
[1] Newcastle Univ, Sch Mech & Syst Engn, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
[2] Royal Victoria Infirm, Great North Childrens Hosp, Newcastle Upon Tyne, Tyne & Wear, England
基金
英国工程与自然科学研究理事会;
关键词
early onset scoliosis; EDX; explant; failure analysis; growing rod; implant; MAGEC; magnetically controlled; metallosis; titanium; wear; EARLY-ONSET SCOLIOSIS; TOTAL HIP-ARTHROPLASTY; ON-METAL BEARINGS; KNEE REPLACEMENTS; WEAR; FAILURE; PROSTHESES; MULTICENTER; TRUNNION; CHILDREN;
D O I
10.1097/BRS.0000000000002221
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Analysis of explanted MAGnetic Expansion Control (MAGEC) growing rods. Objective. To analyze explanted MAGEC rods used in management of early onset scoliosis and identify the mode of failure in such cases. Summary of Background Data. Magnetically controlled growing rods are increasingly used as the option of choice for early onset scoliosis. However, being more complex than conventional growing rods they are perhaps more likely to succumb to multifarious failure modes. In addition, metallosis has been reported around failed MAGEC rods. Methods. Explanted MAGEC rods from seven UK spinal centers were obtained for independent analysis. Thirty-four MAGEC rods, from 18 children, explanted for reasons including failure of rod lengthening and maximum rod distraction reached, were cut open to allow internal components to be evaluated and assessed. Results. Externally, all MAGEC rods showed localized marks, which were termed "growth marks'' as they indicated growth of the rod in vivo, on the extending bar component. After cutting open, titanium wear debris was found inside all 34 (100%) MAGEC rods. Ninety-one percent (31/34) of MAGEC rods showed measurable wear of the extending bar, towards the magnet end. Substantial damage to the radial bearing was seen inside 74% (25/34) of MAGEC rods while O-ring seal failure was seen in 53% (18/34) of cases. In 44% (15/34) of MAGEC rods the drive pin was fractured but this was felt to be an effect of rod failure, not a cause. Conclusion. The combination of high volumes of titanium wear debris alongside O-ring seal damage likely accounts for the metallosis reported clinically around some MAGEC rods. Based on this explant data, a failure mechanism in MAGEC rods due to the natural off axis loading in the spine was proposed. This is the largest data set reporting a complete analysis of explanted MAGEC rods to date.
引用
收藏
页码:E16 / E22
页数:7
相关论文
共 36 条
  • [1] Dual growing rod technique for the treatment of progressive early-onset scoliosis - A multicenter study
    Akbarnia, BA
    Marks, DS
    Boachie-Adjei, O
    Thompson, AG
    Asher, MA
    [J]. SPINE, 2005, 30 (17) : S46 - S57
  • [2] [Anonymous], 2016, FDA Drug Safety Communication: FDA review results in new warnings about using general anesthetics and sedation drugs in young children and pregnant women
  • [3] [Anonymous], 2014, ISO128912
  • [4] [Anonymous], 2016, 510 K SUMMARY MAGEC
  • [5] Complications of Growing-Rod Treatment for Early-Onset Scoliosis Analysis of One Hundred and Forty Patients
    Bess, Shay
    Akbarnia, Behrooz A.
    Thompson, George H.
    Sponseller, Paul D.
    Shah, Suken A.
    El Sebaie, Hazem
    Boachie-Adjei, Oheneba
    Karlin, Lawrence I.
    Canale, Sarah
    Poe-Kochert, Connie
    Skaggs, David L.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (15) : 2533 - 2543
  • [6] Determining material loss from the femoral stem trunnion in hip arthroplasty using a coordinate measuring machine
    Bone, Martin C.
    Sidaginamale, Raghavendra P.
    Lord, James K.
    Scholes, Susan C.
    Joyce, Thomas J.
    Nargol, Anthony V. F.
    Langton, David J.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2015, 229 (01) : 69 - 76
  • [7] Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal-on-metal total hip arthroplasty system
    Brock, Timothy M.
    Sidaginamale, Raghavendra
    Rushton, Steven
    Nargol, Antoni V. F.
    Bowsher, John G.
    Savisaar, Christina
    Joyce, Tom J.
    Deehan, David J.
    Lord, James K.
    Langton, David J.
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2015, 33 (12) : 1868 - 1874
  • [8] General anesthesia, surgery and hospitalization in children and their effects upon cognitive, academic, emotional and sociobehavioral development - a review
    Caldas, JCS
    Pais-Ribeiro, JL
    Carneiro, SR
    [J]. PEDIATRIC ANESTHESIA, 2004, 14 (11) : 910 - 915
  • [9] Special article: Update on the magnetically controlled growing rod: tips and pitfalls
    Cheung, Jason Pui Yin
    Cahill, Patrick
    Yaszay, Burt
    Akbarnia, Behrooz A.
    Cheung, Kenneth M. C.
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY, 2015, 23 (03) : 383 - 390
  • [10] Cobalt intoxication diagnosed with the help of Dr House
    Dahms, Kirsten
    Sharkova, Yulia
    Heitland, Peter
    Pankuweit, Sabine
    Schaefer, Juergen R.
    [J]. LANCET, 2014, 383 (9916) : 574 - 574