Simulation Models of Misclassification Error for Single Thresholds of High-Sensitivity Cardiac Troponin I Due to Assay Bias and Imprecision

被引:62
作者
Lyon, Andrew W. [1 ]
Kavsak, Peter A. [2 ]
Lyon, Oliver A. S.
Worster, Andrew [3 ]
Lyon, Martha E. [1 ]
机构
[1] Saskatoon Hlth Reg, Dept Pathol & Lab Med, Saskatoon, SK, Canada
[2] McMaster Univ, Dept Pathol & Mol Med, Hamilton, ON, Canada
[3] McMaster Univ, Div Emergency Med, Hamilton, ON, Canada
基金
加拿大健康研究院;
关键词
99TH PERCENTILE; RULE-OUT; VALIDATION; VALUES; TIME; CARE;
D O I
10.1373/clinchem.2016.265058
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: Clinical outcome studies for cardiac troponins (cTn) are expensive and difficult to design owing to variation in patients, in the assays, and in the incidence of different types of myocardial infarction (MI). To overcome these difficulties, simulation models were used to estimate the rate of misclassification error for MI and risk prediction resulting from assay bias and imprecision. METHODS: Finite mixture analysis of Abbott high sensitivity cTnI (hs-cTnI) results at time 0 h in patients presenting early with acute coronary syndrome (ACS) symptoms to the emergency department (ED) [n = 145, Reducing the Time Interval for Identifying New Guideline (RING) study] allowed derivation of a simulation data set (n = 10 000). hs-cTnI concentrations were modified by addition of bias or imprecision error. The percentage of all 10 000 modified hs-cTnI results that were misclassified for MI at thresholds of 2, 5, 26.2, and 52 ng/L was determined by Monte Carlo simulation. Analyses were replicated with an all-comer emergency department (ED) population (n = 1137) ROMI (Optimum Troponin Cutoffs for ACS in the ED) study. RESULTS: In the RING study, simulation at 26.2-ng/L (99th percentile) and 52-ng/L thresholds were affected by both bias +/- 2 ng/L and imprecision (10%-20%) and had misclassification rates of 0.4% to 0.6%. Simulations at the +/- 2-ng/L and 5-ng/L thresholds were only affected by bias. Misclassification rates at bias of +/- 1 ng/L were 10% for the +/- 2-ng/L threshold, and 5% for the 5-ng/L threshold. CONCLUSIONS: Simulation models predicted that hscTnI results are seldom misclassified (<1% of patients) when interpretative thresholds are near or exceed the overall 99th percentile. However, simulation models also predicted that low hs-cTnI results, as recommended in guidelines, are prone to misclassification of 5%-10% of patients. (C) 2016 American Association for Clinical Chemistry
引用
收藏
页码:585 / 592
页数:8
相关论文
共 32 条
[1]   Determination of 19 Cardiac Troponin I and T Assay 99th Percentile Values from a Common Presumably Healthy Population [J].
Apple, Fred S. ;
Ler, Ranka ;
Murakami, MaryAnn M. .
CLINICAL CHEMISTRY, 2012, 58 (11) :1574-1581
[2]   Validation of the 99th percentile cutoff independent of assay imprecision (CV) for cardiac troponin monitoring for ruling out myocardial infarction [J].
Apple, FS ;
Parvin, CA ;
Buechler, KF ;
Christenson, RH ;
Wu, AHB ;
Jaffe, AS .
CLINICAL CHEMISTRY, 2005, 51 (11) :2198-2200
[3]  
Box G. E., 1978, Statistics for experimenters, V664
[4]   Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models [J].
Boyd, James C. ;
Bruns, David E. .
CLINICAL CHEMISTRY, 2014, 60 (04) :644-650
[5]  
Boyd JC, 2001, CLIN CHEM, V47, P209
[6]  
Chang M., 2010, MONTE CARLO SIMULATI
[7]  
Deb P, 1997, J APPL ECONOMET, V12, P313, DOI 10.1002/(SICI)1099-1255(199705)12:3<313::AID-JAE440>3.0.CO
[8]  
2-G
[9]   Age, sex, and racial influences on the Beckman Coulter AccuTnI+3 99th percentile [J].
Greene, Dina N. ;
Leong, Thomas K. ;
Collinson, Paul O. ;
Kamer, Sandy M. ;
Huang, Karl ;
Lorey, Thomas S. ;
Go, Alan S. .
CLINICA CHIMICA ACTA, 2015, 444 :149-153
[10]   Long-time quality assessment of the Elecsys Troponin T hs assay [J].
Hammarsten, Ola ;
Jacobsson, Carl-Eric ;
Widegren, Monica ;
Danylchenko, Tatyana ;
Jaffe, Allan S. .
CLINICAL BIOCHEMISTRY, 2013, 46 (12) :1055-1057