Water equivalence of some plastic-water phantom materials for clinical proton beam dosimetry

被引:12
作者
Al-Sulaiti, L. [1 ,2 ]
Shipley, D. [2 ]
Thomas, R. [2 ]
Owen, P. [1 ,2 ]
Kacperek, A. [3 ]
Regan, P. H. [1 ]
Palmans, H. [2 ]
机构
[1] Univ Surrey, Dept Phys, Guildford GU2 5XH, Surrey, England
[2] Natl Phys Lab, Radiat Dosimetry Team, Teddington TW11 0LW, Middx, England
[3] Clatterbridge Ctr Oncol, Wirral, Merseyside, England
关键词
Proton dosimetry; Fluence correction factor; Plastic water; Monte Carlo; Water equivalent; SOLID MATERIALS; ELECTRON-BEAMS; PHOTON;
D O I
10.1016/j.apradiso.2012.02.002
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Plastic-water phantom materials are not exactly water equivalent since they have a different elemental composition and different interaction cross sections for protons than water. Several studies of the water equivalence of plastic-water phantom materials have been reported for photon and electron beams, but none for clinical proton beams. In proton beams, the difference between non-elastic nuclear interactions in plastic-water phantom materials compared to those in water should be considered. In this work, the water equivalence of Plastic Water (R) (PW)(1), Plastic Water (R), Diagnostic Therapy (PWDT)(1) and solid water (WT1)(2) phantoms was studied for clinical proton energies of 60 MeV and 200 MeV. This was done by evaluating the fluence correction factor at equivalent depths; first with respect to water and then with respect to graphite by experiment and Monte Carlo (MC) simulations using FLUKA. MC simulations showed that the fluence correction with respect to water was less than 0.5% up to the entire penetration depth of the protons at 60 MeV and less than 1% at 200 MeV up to 20 cm depth for PWDT, PW and WT1. With respect to graphite the fluence correction was about 0.5% for 60 MeV and about 4% for 200 MeV. The experimental results for modulated and un-modulated 60 MeV proton beams showed good agreement with the MC simulated fluence correction factors with respect to graphite deviating less than 1% from unity for the three plastic-water phantoms. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1052 / 1057
页数:6
相关论文
共 16 条
[1]   Water equivalence of various materials for clinical proton dosimetry by experiment and Monte Carlo simulation [J].
Al-Sulaiti, Leena ;
Shipley, David ;
Thomas, Russell ;
Kacperek, Andrzej ;
Regan, Patrick ;
Palmans, Hugo .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3) :344-347
[2]  
[Anonymous], 2000, ICRU632000
[3]  
[Anonymous], 2000, IAEATRS398
[4]   Monte Carlo calculations of correction factors for plastic phantoms in clinical photon and electron beam dosimetry [J].
Araki, Fujio ;
Hanyu, Yuji ;
Fukuoka, Miyoko ;
Matsumoto, Kenji ;
Okumura, Masahiko ;
Oguchi, Hiroshi .
MEDICAL PHYSICS, 2009, 36 (07) :2992-3001
[5]  
Battistoni G, 2007, AIP CONF PROC, V896, P31, DOI 10.1063/1.2720455
[6]  
Fasso F., 2005, CERN200510
[7]  
ICRU, 1984, ICRU37
[8]   Nuclear collision processes around the Bragg peak in proton therapy [J].
Matsuzaki Y. ;
Date H. ;
Sutherland K.L. ;
Kiyanagi Y. .
Radiological Physics and Technology, 2010, 3 (1) :84-92
[9]   Characterization of the phantom material Virtual Water™ in high-energy photon and electron beams [J].
McEwen, MR ;
Niven, D .
MEDICAL PHYSICS, 2006, 33 (04) :876-887
[10]   Nuclear interactions in proton therapy: dose and relative biological effect distributions originating from primary and secondary particles [J].
Paganetti, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (05) :747-764