Validation of the Beurer BM 44 upper arm blood pressure monitor for home measurement, according to the European Society of Hypertension International Protocol 2002

被引:2
|
作者
Lueders, Stephan [1 ,2 ]
Krueger, Ralf [2 ,3 ]
Zemmrich, Claudia [3 ]
Forstner, Klaus [4 ]
Sturm, Claus-Dieter [2 ]
Bramlage, Peter [2 ,3 ]
机构
[1] St Josefs Hosp, Med Clin, D-49661 Cloppenburg, Germany
[2] TEMPESTAS Inst Med Res, Cloppenburg, Germany
[3] Inst Pharmacol & Prevent Med GmbH, Potsdam, Germany
[4] Forschungsinst Klin Med Tech, Res Inst Clin Engn, Asperg, Germany
关键词
automated upper arm blood pressure; Beurer BM 44; European Society of Hypertension International Protocol 2002; validation; OFFICE;
D O I
10.1097/MBP.0b013e32835b9e8e
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The present study aimed to validate the automated upper arm blood pressure (BP) measuring device BM 44 for home BP monitoring according to the 2002 Protocol of the European Society of Hypertension. The most important new feature of the new device was an integrated 'WHO indicator', which categorizes the patient's individual result within the WHO recommendations for target BP by a coloured scale. Methods Systolic and diastolic BPs were measured sequentially in 35 adult participants (16 men, 19 women) using a standard mercury y-tubed reference sphygmomanometer (two observers) and the BM 44 device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and a further 18 participants in phase 2 of the validation study. Results The BM 44 device passed phase 1 of the validation study successfully with a number of absolute differences between device and observers of 5, 10 and 15mmHg for at least 28 out of 25, 35 out of 35 and 40 out of 40 measurements, respectively. The device also achieved the targets for phases 2.1 and 2.2, with 23 and 26 participants having had at least two of three device-observers differences within 5mmHg for systolic and diastolic BP, respectively. Conclusion The Beurer BM 44 upper arm BP monitor has passed the International Protocol requirements, and hence can be recommended for home use in adults. Blood Press Monit 17:248-252 (C) 2012 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins. Blood Pressure Monitoring 2012, 17:248-252
引用
收藏
页码:248 / 252
页数:5
相关论文
共 50 条
  • [31] Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension
    Pini, Claudio
    Natalizi, Anna
    Gerosa, Pietro Francesco
    Frigerioa, Mauro
    Omboni, Stefano
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2008, 13 (03) : 177 - 182
  • [32] Validation of the Artsana CSI 610 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension
    Pini, Claudio
    Pastori, Marco
    Baccheschi, Jordan
    Omboni, Stefano
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2007, 12 (03) : 179 - 184
  • [33] Validation of the SCIAN LD-735 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2016, 21 (04) : 255 - 258
  • [34] Validation of the AVITA BPM17 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Liu, Chang-Yuan
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2017, 22 (04) : 230 - 233
  • [35] Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2012, 17 (05) : 204 - 206
  • [36] Self-measurement of blood pressure: validation of the Braun BP VitalScan Plus 1650 monitor according to the European Society of Hypertension International Protocol
    Nolly, Hector
    Osso, Pablo
    Nolly, Alejandro
    Silva, Martin
    Nolly, Mariela
    Romero, Marcelo
    BLOOD PRESSURE MONITORING, 2008, 13 (02) : 101 - 106
  • [37] Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high-altitude environment, according to the European Society of Hypertension International Protocol revision 2010
    K Cho
    M Tian
    Y Lan
    X Zhao
    L L Yan
    Journal of Human Hypertension, 2013, 27 : 487 - 491
  • [38] Validation of the PHILIPS DL8760 upper arm blood pressure monitor, in oscillometry mode, for self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010
    Chahine, Mirna N.
    Bou Harb, Souad
    Saad, Abdel Rahman
    Sarkis, Patrick
    Azaki, Alaa
    Harb, Aya
    Allouch, Ali
    Asmar, Roland
    BLOOD PRESSURE MONITORING, 2021, 26 (03) : 237 - 241
  • [39] Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high-altitude environment, according to the European Society of Hypertension International Protocol revision 2010
    Cho, K.
    Tian, M.
    Lan, Y.
    Zhao, X.
    Yan, L. L.
    JOURNAL OF HUMAN HYPERTENSION, 2013, 27 (08) : 487 - 491
  • [40] Validation of noninvasive oscillometric blood pressure 2020 up pressure upper arm blood pressure monitoring technology according to the European Society of Hypertension International Protocol revision 2010
    Reshetnik, Alexander
    Gohlisch, Christopher
    Abou-Dakn, Michael
    Toelle, Markus
    Zidek, Walter
    van der Giet, Markus
    BLOOD PRESSURE MONITORING, 2019, 24 (02) : 99 - 101