Validation of the Beurer BM 44 upper arm blood pressure monitor for home measurement, according to the European Society of Hypertension International Protocol 2002

被引:2
|
作者
Lueders, Stephan [1 ,2 ]
Krueger, Ralf [2 ,3 ]
Zemmrich, Claudia [3 ]
Forstner, Klaus [4 ]
Sturm, Claus-Dieter [2 ]
Bramlage, Peter [2 ,3 ]
机构
[1] St Josefs Hosp, Med Clin, D-49661 Cloppenburg, Germany
[2] TEMPESTAS Inst Med Res, Cloppenburg, Germany
[3] Inst Pharmacol & Prevent Med GmbH, Potsdam, Germany
[4] Forschungsinst Klin Med Tech, Res Inst Clin Engn, Asperg, Germany
关键词
automated upper arm blood pressure; Beurer BM 44; European Society of Hypertension International Protocol 2002; validation; OFFICE;
D O I
10.1097/MBP.0b013e32835b9e8e
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The present study aimed to validate the automated upper arm blood pressure (BP) measuring device BM 44 for home BP monitoring according to the 2002 Protocol of the European Society of Hypertension. The most important new feature of the new device was an integrated 'WHO indicator', which categorizes the patient's individual result within the WHO recommendations for target BP by a coloured scale. Methods Systolic and diastolic BPs were measured sequentially in 35 adult participants (16 men, 19 women) using a standard mercury y-tubed reference sphygmomanometer (two observers) and the BM 44 device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and a further 18 participants in phase 2 of the validation study. Results The BM 44 device passed phase 1 of the validation study successfully with a number of absolute differences between device and observers of 5, 10 and 15mmHg for at least 28 out of 25, 35 out of 35 and 40 out of 40 measurements, respectively. The device also achieved the targets for phases 2.1 and 2.2, with 23 and 26 participants having had at least two of three device-observers differences within 5mmHg for systolic and diastolic BP, respectively. Conclusion The Beurer BM 44 upper arm BP monitor has passed the International Protocol requirements, and hence can be recommended for home use in adults. Blood Press Monit 17:248-252 (C) 2012 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins. Blood Pressure Monitoring 2012, 17:248-252
引用
收藏
页码:248 / 252
页数:5
相关论文
共 50 条
  • [11] Validation of the HONSUN LD-578 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Zhang, Yi
    Wang, Jie
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2009, 14 (03) : 128 - 131
  • [12] Validation of the ANDON KD-5915 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Huang, Qi-Fang
    Wang, Jie
    Sheng, Chang-Sheng
    Zhang, Na-Na
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2010, 15 (04) : 232 - 234
  • [13] A NOVEL DUAL-MODE AMBULATORY AND HOME BLOOD PRESSURE MONITOR: VALIDATION ACCORDING TO THE EUROPEAN SOCIETY OF HYPERTENSION INTERNATIONAL PROTOCOL
    Stergiou, G. S.
    Tzamouranis, D.
    Nasothimiou, E.
    Karpettas, N.
    Protogerou, A.
    JOURNAL OF HYPERTENSION, 2009, 27 : S195 - S195
  • [14] Validation of the Medipro MediCare 100f upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Yi, Jun
    Wan, Yi
    Pan, Feng
    Yu, Xiaorong
    Zhao, Huadong
    Shang, Fujun
    Xu, Yongyong
    BLOOD PRESSURE MONITORING, 2011, 16 (04) : 197 - 199
  • [15] Validation of the iHealth BP5 wireless upper arm blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Shang, Fujun
    Zhu, Yizheng
    Zhu, Zhenlai
    Liu, Lei
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2013, 18 (05) : 278 - 281
  • [16] Validation of the Kinetik Blood Pressure Monitor-Series 1 for use in adults at home and in clinical settings, according to the 2002 European Society of Hypertension International Protocol on the validation of blood pressure devices
    Schwartz, C. L.
    Edwards, K.
    Gamble, W.
    Kirkham, A.
    Lacy, P.
    Lewis, P.
    McDonagh, S. T. J.
    Peers, C.
    Sheppard, J. P.
    Swales, P.
    Howarth, J.
    Williams, B.
    JOURNAL OF HUMAN HYPERTENSION, 2021, 35 (11) : 1046 - 1050
  • [17] Validation of the Kingyield BP101H blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2010, 15 (06) : 329 - 331
  • [18] Validation of the AGILIS ambulatory blood pressure monitor according to the European Society of Hypertension International protocol for validation of blood pressure measuring devices in adults
    Mallion, JJM
    Pierre, H
    Neuder, Y
    Ormezzano, O
    Baguet, JP
    BLOOD PRESSURE MONITORING, 2005, 10 (02) : 97 - 101
  • [19] QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease: A Validation Study According to the European Society of Hypertension International Protocol Revision 2010
    Mazoteras-Pardo, Victoria
    Becerro-De-Bengoa-Vallejo, Ricardo
    Losa-Iglesias, Marta Elena
    Lopez-Lopez, Daniel
    Rodriguez-Sanz, David
    Casado-Hernandez, Israel
    Calvo-Lobo, Cesar
    Palomo-Lopez, Patricia
    JOURNAL OF MEDICAL INTERNET RESEARCH, 2019, 21 (12)
  • [20] Validation of the A&D UA-1020 upper-arm blood pressure monitor for home blood pressure monitoring according to the British Hypertension Society Protocol
    Zeng, Wei-Fang
    Kang, Yuan-Yuan
    Liu, Ming
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2013, 18 (03) : 177 - 181