Study of the radioluminesence spectra of doped silica optical fibre dosimeters for stem effect removal

被引:26
作者
Veronese, I. [1 ,2 ]
Cantone, M. C. [1 ,2 ]
Catalano, M. [3 ]
Chiodini, N. [4 ,5 ]
Fasoli, M. [4 ,5 ]
Mancosu, P. [3 ]
Mones, E. [6 ]
Moretti, F. [4 ,5 ]
Scorsetti, M. [3 ]
Vedda, A. [4 ,5 ]
机构
[1] Univ Milan, Dipartimento Fis, I-20133 Milan, Italy
[2] INFN, I-20133 Milan, Italy
[3] Ist Clin Humanitas, Milan, Italy
[4] Univ Milano Bicocca, Dipartimento Sci Mat, I-20125 Milan, Italy
[5] INFN, I-20125 Milan, Italy
[6] Azienda Osped Maggiore Carita, Dept Med Phys, I-28100 Novara, Italy
关键词
SCINTILLATION DOSIMETRY; CERENKOV RADIATION; ELECTRON; PHOTON; RADIOTHERAPY; CE3+; BAND;
D O I
10.1088/0022-3727/46/1/015101
中图分类号
O59 [应用物理学];
学科分类号
摘要
In this study, the spectral emission of Eu3+-doped silica optical fibres is investigated under irradiation with photons and electrons of different energies, field sizes and orientations, in order to discover the origin of stem effect and evaluate its influence on the radioluminescence (RL) spectral shape under experimental conditions. A comparison with previously investigated Ce-doped silica fibres is also made. Below the Cerenkov radiation energy threshold, a slight contribution to the total RL signal is observed. This is due to fluorescence effects occurring in the passive fibre. Above the energy threshold, the stem effect is mainly caused by Cerenkov radiation. A method for removing the stem effect, based on RL spectral analysis, is proposed and validated by measuring the output factors of extended photon fields and by studying the angular dependence of the dosimeter.
引用
收藏
页数:7
相关论文
共 31 条
  • [1] An algorithm for real-time dosimetry in intensity-modulated radiation therapy using the radioluminescence signal from Al2O3:C
    Andersen, C. E.
    Marckmann, C. J.
    Aznar, M. C.
    Botter-Jensen, L.
    Kjaer-Kristoffersen, F.
    Medin, J.
    [J]. RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) : 7 - 13
  • [2] Fiber-coupled radioluminescence dosimetry with saturated Al2O3:C crystals: Characterization in 6 and 18 MV photon beams
    Andersen, C. E.
    Damkjaer, S. M. S.
    Kertzscher, G.
    Greilich, S.
    Aznar, M. C.
    [J]. RADIATION MEASUREMENTS, 2011, 46 (10) : 1090 - 1098
  • [3] Real-time optical-fibre luminescence dosimetry for radiotherapy:: physical characteristics and applications in photon beams
    Aznar, MC
    Andersen, CE
    Botter-Jensen, L
    Bäck, SÅJ
    Mattsson, S
    Kjær-Kristoffersen, F
    Medin, J
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (09) : 1655 - 1669
  • [4] Plastic scintillation dosimetry and its application to radiotherapy
    Beddar, A. S.
    [J]. RADIATION MEASUREMENTS, 2006, 41 : S124 - S133
  • [5] CERENKOV LIGHT GENERATED IN OPTICAL FIBERS AND OTHER LIGHT PIPES IRRADIATED BY ELECTRON-BEAMS
    BEDDAR, AS
    MACKIE, TR
    ATTIX, FH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04) : 925 - 935
  • [6] THE 2.7 EV PHOTOLUMINESCENCE BAND IN HIGH-PURITY SYNTHETIC SILICA
    BERTINO, M
    CORAZZA, A
    MARTINI, M
    MERVIC, A
    SPINOLO, G
    [J]. JOURNAL OF PHYSICS-CONDENSED MATTER, 1994, 6 (31) : 6345 - 6352
  • [7] Chiodini N., 2009, P SPIE, V7316
  • [8] Ciocca Mario, 2012, Int J Radiat Oncol Biol Phys, V82, pe305, DOI 10.1016/j.ijrobp.2011.05.010
  • [9] A temporal method of avoiding the Cerenkov radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron and photon beams
    Clift, MA
    Johnston, PN
    Webb, DV
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) : 1421 - 1433
  • [10] Water equivalence of plastic organic scintillators in megavoltage radiotherapy bremsstrahlung beams
    Clift, MA
    Sutton, RA
    Webb, DV
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (07) : 1885 - 1895