Development and Calibration of a New Automated Method to Measure Air Conduction Auditory Thresholds Using an Active Earplug

被引:3
|
作者
Vinay [1 ,3 ]
Henriksen, Viggo [2 ]
Kvaloy, Olav [2 ]
Svensson, U. Peter [1 ]
机构
[1] Norwegian Univ Sci & Technol, Acoust Res Ctr, Dept Elect & Telecommun, N-7034 Trondheim, Norway
[2] SINTEF ICT, Acoust Grp, Trondheim, Norway
[3] Sor Trondelag Univ Coll, Program Audiol Educ, Trondheim, Norway
关键词
RELIABILITY; ACCURACY; VALIDITY;
D O I
10.3813/AAA.918691
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A new automated method termed as 'New Early Warning Test (NEWT)' has been designed to provide automatic auditory threshold monitoring in individuals exposed to high noise levels. The NEWT method is incorporated inside an active communication earplug called Quietpro (R), which has high attenuation characteristics for background noise. The NEWT measures auditory thresholds using a two-peep pure tone stimulus. The aim of the present study was to establish the calibration values required for the NEWT method by comparing the deviations in the thresholds obtained in the NEWT against standard pure tone audiometry (PTA). Sixteen subjects (twenty-eight ears) with normal hearing participated in this study. Auditory thresholds were measured using the NEWT and PTA for frequencies of 1, 3, 4 and 6 kHz. It was observed that there was no significant difference in inter-subject variability. Because this inter-subject variability also included intra-subject variability, results suggested that the NEWT method had at least not any higher intra-subject variability than the PTA. Results of the present study will be used to calibrate the NEWT method, which helps in regular monitoring of hearing status in individuals exposed to noise. These results might have important implications to further evaluate the NEWT method on various groups of population having different degrees of hearing loss.
引用
收藏
页码:113 / 117
页数:5
相关论文
共 2 条
  • [1] Development and Validation of a New Method to Measure Walking Speed in Free-Living Environments Using the Actibelt® Platform
    Schimpl, Michaela
    Lederer, Christian
    Daumer, Martin
    PLOS ONE, 2011, 6 (08):
  • [2] A New Automated Way to Measure Polyethylene Wear in THA Using a High Resolution CT Scanner: Method and Analysis
    Maguire, Gerald Q., Jr.
    Noz, Marilyn E.
    Olivecrona, Henrik
    Zeleznik, Michael P.
    Weidenhielm, Lars
    SCIENTIFIC WORLD JOURNAL, 2014,