Visual performance with accommodating and multifocal intraocular lenses

被引:4
作者
Lan, Jie [1 ,2 ]
Huang, Yu-Sen [2 ]
Dai, Yun-Hai [2 ]
Wu, Xiao-Ming [2 ]
Sun, Jia-Jun [2 ]
Xie, Li-Xin [2 ]
机构
[1] Qingdao Univ, Coll Med, Qingdao 266071, Shandong, Peoples R China
[2] Shandong Acad Med Sci, Qingdao Eye Hosp, Shandong Eye Inst, Qingdao 266071, Shandong, Peoples R China
基金
中国国家自然科学基金;
关键词
intraocular lens; accommodating; multifocal; visual performance; CONTRAST SENSITIVITY; OPTICAL-PERFORMANCE; EYE MODEL; ADD-ON; IMPLANTATION; RESTOR; REZOOM; ACUITY; ABERRATION; CATARACT;
D O I
10.18240/ijo.2017.02.09
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
AIM: To compare the visual functional outcomes with accommodating and multifocal intraocular lenses (IOLs). 0 METHODS: Our retrospective comparative study included 51 patients (60 eyes) received implantation of an accommodating IOL (Tetraflex; 16 patients, 20 eyes), a refractive multifocal IOL (ReZoom; 18 patients, 20 eyes), or a diffractive multifocal IOL (ZMA00; 17 patients, 20 eyes). Subjective refraction, visual acuity, contrast sensitivity (CS), intraocular aberration, and subjective photic phenomena were detected at 3mo after surgery. RESULTS: The spherical equivalent in the three groups was -0.38 +/- 0.54 D, 0.14 +/- 0.56 D, and 0.35 +/- 0.41 D, respectively. No statistically significant differences were found in uncorrected and corrected distance visual acuity and uncorrected intermediate visual acuity among the groups (P=0.39). The ReZoom group had significantly better distance-corrected intermediate visual acuity than the ZMA00 group (P=0.003). The ZMA00 group had significantly better near visual acuity than the other groups (P<0.05). Better contrast sensitivity values were observed in the Tetraflex group under most of the spatial frequencies conditions (P=0.025). The total aberration was lowest in the ZMA00 group (P=0.000), and the spherical aberration was highest in the Tetraflex group (P=0.000). The three groups had similar frequency of ghosting and glare, and the Tetraflex group had a low rate of halos (P=0.01). CONCLUSION: Both accommodating and multifocal 1IOLs can successfully restore distance and uncorrected intermediate visual acuities. Tetraflex accommodating IOLs perform better in CS and with less halos of photic phenomena. ReZoom refractive multifocal 10Ls have
引用
收藏
页码:235 / 240
页数:6
相关论文
共 30 条
[1]   Visual and Optical Performance With Two Different Diffractive Multifocal Intraocular Lenses Compared to a Monofocal Lens [J].
Alio, Jorge L. ;
Pinero, David P. ;
Plaza-Puche, Ana B. ;
Amparo, Francisco ;
Jimenez, Ramon ;
Rodriguez-Prats, Jose L. ;
Javaloy, Jaime .
JOURNAL OF REFRACTIVE SURGERY, 2011, 27 (08) :570-581
[2]   Multifocal intraocular lenses in cataract surgery: Literature review of benefits and side effects [J].
de Vries, Niels E. ;
Nuijts, Rudy M. M. A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (02) :268-278
[3]   The ReZoom multifocal intraocular lens: 2-year follow-up [J].
Forte, Raimondo ;
Ursoleo, Paolo .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2009, 19 (03) :380-383
[4]   Visual acuity, contrast sensitivity, subjective quality of vision, and quality of life with 4 different multifocal IOLs [J].
Gil, Miguel A. ;
Varon, Consuelo ;
Rosello, Noemi ;
Cardona, Genis ;
Buil, Jose A. .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2012, 22 (02) :175-187
[5]   Far and near visual acuity with multifocal intraocular lenses in an optomechanical eye model with imaging capability [J].
Gobbi, Pier Giorgio ;
Fasce, Francesco ;
Bozza, Stefano ;
Calori, Giliola ;
Brancato, Rosario .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (06) :1082-1094
[6]   Retinal Straylight and Complaint Scores 18 Months After Implantation of the AcrySof Monofocal and ReSTOR Diffractive Intraocular Lenses [J].
Hofmann, Thomas ;
Zuberbuhler, Bruno ;
Cervino, Alejandro ;
Montes-Mico, Robert ;
Haefliger, Eduard .
JOURNAL OF REFRACTIVE SURGERY, 2009, 25 (06) :485-492
[7]   Long-term visual outcomes after Crystalens (R) HD intraocular lens implantation [J].
Karavitaki, Alexandra E. ;
Pallikaris, Ioannis G. ;
Panagopoulou, Sophia I. ;
Kounis, George A. ;
Kontadakis, George ;
Kymionis, George D. .
CLINICAL OPHTHALMOLOGY, 2014, 8 :937-943
[8]   Objective evaluation of through-focus optical performance of presbyopia-correcting intraocular lenses using an optical bench system [J].
Kim, Myoung Joon ;
Zheleznyak, Len ;
MacRae, Scott ;
Tchah, Hungwon ;
Yoon, Geunyoung .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (07) :1305-1312
[9]  
Lan Jie, 2013, Zhonghua Yan Ke Za Zhi, V49, P139
[10]   Contrast visual acuity after multifocal intraocular lens implantation: aspheric versus spherical design [J].
Li, Jun-Hua ;
Feng, Yi-Fan ;
Zhao, Yun-E ;
Zhao, Yin-Ying ;
Lin, Lei .
INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2014, 7 (01) :100-103