Dual energy CT for attenuation correction with PET/CT

被引:19
作者
Xia, Ting [1 ]
Alessio, Adam M. [2 ]
Kinahan, Paul E. [1 ,2 ]
机构
[1] Univ Washington, Dept Bioengn, Seattle, WA 98105 USA
[2] Univ Washington, Dept Radiol, Seattle, WA 98105 USA
关键词
dual energy CT; PET/CT; attenuation correction; noise and bias; X-RAY CT; COMPUTED-TOMOGRAPHY; QUANTITATIVE ASSESSMENT; SPECT; ALGORITHM; TRANSFORMATION; COEFFICIENTS; ARTIFACT; IMAGES; NOISE;
D O I
10.1118/1.4828838
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The authors evaluate the energy dependent noise and bias properties of monoenergetic images synthesized from dual-energy CT (DECT) acquisitions. These monoenergetic images can be used to estimate attenuation coefficients at energies suitable for positron emission tomography (PET) and single-photon emission computed tomography (SPECT) imaging. This is becoming more relevant with the increased use of quantitative imaging by PET/CT and SPECT/CT scanners. There are, however, potential variations in the noise and bias of synthesized monoenergetic images as a function of energy. Methods: The authors used analytic approximations and simulations to estimate the noise and bias of synthesized monoenergetic images of water-filled cylinders with different shapes and the NURBSbased cardiac-torso (NCAT) phantom from 40 to 520 keV, the range of SPECT and PET energies. The dual-kVp spectra were based on the GE Lightspeed VCT scanner at 80 and 140 kVp with added filtration of 0.5 mm Cu. The authors evaluated strategies of noise suppression with sinogram smoothing and dose minimization with reduction of tube currents at the two kVp settings. The authors compared the impact of DECT-based attenuation correction with single-kVp CT-based attenuation correction on PET quantitation for the NCAT phantom for soft tissue and high-Z materials of bone and iodine contrast enhancement. Results: Both analytic calculations and simulations displayed the expected minimum noise value for a synthesized monoenergetic image at an energy between the mean energies of the two spectra. In addition the authors found that the normalized coefficient of variation in the synthesized attenuation map increased with energy but reached a plateau near 160 keV, and then remained constant with increasing energy up to 511 keV and beyond. The bias was minimal, as the linear attenuation coefficients of the synthesized monoenergetic images were within 2.4% of the known true values across the entire energy range. Compared with no sinogram smoothing, sinogram smoothing can dramatically reduce noise in the DECT-derived attenuation map. Through appropriate selection of tube currents for high and low kVp scans, DECT can deliver roughly the same amount of radiation dose as that of a single kVp CT scan, but could be used for PET attenuation correction with reduced bias in contrast agent regions by a factor of similar to 2.6 and slightly reduced RMSE for the total image. Conclusions: When DECT is used for attenuation correction at higher energies, there is a noise amplification that is dependent on the energy of the synthesized monoenergetic image of linear attenuation coefficients. Sinogram smoothing reduces the noise amplification in DECT-derived attenuation maps without increasing bias. With an appropriate selection of CT techniques, a DECT scan with the same radiation dose as a single CT scan can result in a PET image with improved quantitative accuracy. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:11
相关论文
共 41 条
[1]   Accuracy of CT-based attenuation correction in PET/CT bone imaging [J].
Abella, Monica ;
Alessio, Adam M. ;
Mankoff, David A. ;
MacDonald, Lawrence R. ;
Jose Vaquero, Juan ;
Desco, Manuel ;
Kinahan, Paul E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (09) :2477-2490
[2]   COMPARISON OF NOISE AND DOSE IN CONVENTIONAL AND ENERGY SELECTIVE COMPUTED-TOMOGRAPHY [J].
ALVAREZ, R ;
SEPPI, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (02) :2853-2855
[3]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[4]  
Antoch G, 2002, J NUCL MED, V43, P1339
[5]   Comparative Assessment of Energy-Mapping Approaches in CT-Based Attenuation Correction for PET [J].
Ay, Mohammad R. ;
Shirmohammad, Maryam ;
Sarkar, Saeed ;
Rahmim, Arman ;
Zaidi, Habib .
MOLECULAR IMAGING AND BIOLOGY, 2011, 13 (01) :187-198
[6]  
Beyer T, 2000, J NUCL MED, V41, P1369
[7]   Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: Myocardial perfusion assessment [J].
Blankespoor, SC ;
Wu, X ;
Kalki, K ;
Brown, JK ;
Tang, HR ;
Cann, CE ;
Hasegawa, BH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (04) :2263-2274
[8]   PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients [J].
Burger, C ;
Goerres, G ;
Schoenes, S ;
Buck, A ;
Lonn, AHR ;
von Schulthess, GK .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (07) :922-927
[9]  
Cherry S. R., 2012, Physics in Nuclear Medicine
[10]   COMPARISON OF 4 DUAL ENERGY IMAGE DECOMPOSITION METHODS [J].
CHUANG, KS ;
HUANG, HK .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (04) :455-466