Ultra-high spatial resolution BOLD fMRI in humans using combined segmented-accelerated VFA-FLEET with a recursive RF pulse design

被引:14
作者
Berman, Avery J. L. [1 ,2 ]
Grissom, William A. [3 ,4 ]
Witzel, Thomas [1 ,2 ]
Nasr, Shahin [1 ,2 ]
Park, Daniel J. [1 ]
Setsompop, Kawin [1 ,2 ,5 ]
Polimeni, Jonathan R. [1 ,2 ,5 ]
机构
[1] Massachusetts Gen Hosp, Athinoula A Martinos Ctr Biomed Imaging, 149 13th St,Room 2301, Charlestown, MA 02129 USA
[2] Harvard Med Sch, Dept Radiol, Boston, MA 02115 USA
[3] Vanderbilt Univ, Inst Imaging Sci, 221 Kirkland Hall, Nashville, TN 37235 USA
[4] Vanderbilt Univ, Dept Biomed Engn, Nashville, TN 37235 USA
[5] MIT, Harvard MIT Div Hlth Sci & Technol, Cambridge, MA 02139 USA
关键词
BOLD; FLEET; fMRI; high spatial resolution; multi-shot EPI; segmented EPI; SMS; variable flip angle; OCULAR DOMINANCE COLUMNS; HUMAN BRAIN; VISUAL-CORTEX; NEGATIVE BOLD; ECHO; MRI; FIELD; CONTRAST; IMAGES; ACTIVATION;
D O I
10.1002/mrm.28415
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To alleviate the spatial encoding limitations of single-shot echo-planar imaging (EPI) by developing multi-shot segmented EPI for ultra-high-resolution functional MRI (fMRI) with reduced ghosting artifacts from subject motion and respiration. Theory and Methods Segmented EPI can reduce readout duration and reduce acceleration factors, however, the time elapsed between segment acquisitions (on the order of seconds) can result in intermittent ghosting, limiting its use for fMRI. Here, "FLEET" segment ordering, where segments are looped over before slices, was combined with a variable flip angle progression (VFA-FLEET) to improve inter-segment fidelity and maximize signal for fMRI. Scaling a sinc pulse's flip angle for each segment (VFA-FLEET-Sinc) produced inconsistent slice profiles and ghosting, therefore, a recursive Shinnar-Le Roux (SLR) radiofrequency (RF) pulse design was developed (VFA-FLEET-SLR) to generate unique pulses for every segment that together produce consistent slice profiles and signals. Results The temporal stability of VFA-FLEET-SLR was compared against conventional-segmented EPI and VFA-FLEET-Sinc at 3T and 7T. VFA-FLEET-SLR showed reductions in both intermittent and stable ghosting compared to conventional-segmented and VFA-FLEET-Sinc, resulting in improved image quality with a minor trade-off in temporal SNR. Combining VFA-FLEET-SLR with acceleration, we achieved a 0.6-mm isotropic acquisition at 7T, without zoomed imaging or partial Fourier, demonstrating reliable detection of blood oxygenation level-dependent (BOLD) responses to a visual stimulus. To counteract the increased repetition time from segmentation, simultaneous multi-slice VFA-FLEET-SLR was demonstrated using RF-encoded controlled aliasing. Conclusions VFA-FLEET with a recursive RF pulse design supports acquisitions with low levels of artifact and spatial blur, enabling fMRI at previously inaccessible spatial resolutions with a "full-brain" field of view.
引用
收藏
页码:134 / 153
页数:20
相关论文
共 94 条
  • [21] INNER VOLUME MR IMAGING - TECHNICAL CONCEPTS AND THEIR APPLICATION
    FEINBERG, DA
    HOENNINGER, JC
    CROOKS, LE
    KAUFMAN, L
    WATTS, JC
    ARAKAWA, M
    [J]. RADIOLOGY, 1985, 156 (03) : 743 - 747
  • [22] PHASE ERRORS IN MULTISHOT ECHO-PLANAR IMAGING
    FEINBERG, DA
    OSHIO, K
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) : 535 - 539
  • [23] Feinberg DA, 2013, P INT SOC MAGN RESON, V21, P3316
  • [24] Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging
    Feinberg, David A.
    Moeller, Steen
    Smith, Stephen M.
    Auerbach, Edward
    Ramanna, Sudhir
    Glasser, Matt F.
    Miller, Karla L.
    Ugurbil, Kamil
    Yacoub, Essa
    [J]. PLOS ONE, 2010, 5 (12):
  • [25] Feiweier T, 2011, US Patent, Patent No. 20110234221
  • [26] Laminar imaging of positive and negative BOLD in human visual cortex at 7 T
    Fracasso, Alessio
    Luijten, Peter R.
    Dumoulin, Serge O.
    Petridou, Natalia
    [J]. NEUROIMAGE, 2018, 164 : 100 - 111
  • [27] Whole-Brain Cerebral Blood Flow Mapping Using 3D Echo Planar Imaging and Pulsed Arterial Tagging
    Gai, Neville D.
    Talagala, S. Lalith
    Butman, John A.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 33 (02) : 287 - 295
  • [28] High-Resolution fMRI Reveals Laminar Differences in Neurovascular Coupling between Positive and Negative BOLD Responses
    Goense, Jozien
    Merkle, Hellmut
    Logothetis, Nikos K.
    [J]. NEURON, 2012, 76 (03) : 629 - 639
  • [29] Brief visual stimulation allows mapping of ocular dominance in visual cortex using fMRI
    Goodyear, BG
    Menon, RS
    [J]. HUMAN BRAIN MAPPING, 2001, 14 (04) : 210 - 217
  • [30] Homogeneous non-selective and slice-selective parallel-transmit excitations at 7 Tesla with universal pulses: A validation study on two commercial RF coils
    Gras, Vincent
    Boland, Markus
    Vignaud, Alexandre
    Ferrand, Guillaume
    Amadon, Alexis
    Mauconduit, Franck
    Le Bihan, Denis
    Stoecker, Tony
    Boulant, Nicolas
    [J]. PLOS ONE, 2017, 12 (08):