A novel technique for the effective concentration of 99mTc from a large alumina column loaded with low specific-activity (n,γ)-produced 99Mo

被引:14
作者
Chattopadhyay, Sankha [1 ]
Das, Malay Kanti [1 ]
机构
[1] Ctr Variable Energy Cyclotron, BRIT, Radiopharmaceut Lab, Kolkata 700064, India
关键词
Mo-99-Tc-99m generator; post-concentration; Dowex-1 ion-exchange column; 'Dry' [Tc-99m]-pertechnetate;
D O I
10.1016/j.apradiso.2008.02.085
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Molybdenum-99 of low-to-medium specific activity, 7.4-14.8GBq/g, obtained by the Mo-98(n,gamma)Mo-99 reaction in research reactors requires a large alumina column to adsorb the molybdenum since the capacity of alumina to adsorb Mo is limited (20mgMo/g of alumina). The large size of alumina column, in turn, requires large eluant volumes to elute the Tc-99m, and the radioactive concentration (RAC) of the pertechnetate becomes unacceptably low for radiopharmaceutical applications. This study describes a method for obtaining high RAC pertechnetate solutions from a column generator loaded with 99Mo of low specific activity. Molybdenum-99 of specific activity, 7.4 G Bq/g, was absorbed on an alumina column [60 g, 70 mm, (H) x 35 mm (diam.)]. The (TcO4-)-Tc-99m, eluted with 60 ml normal saline (0.15 N NaCl), was applied to a 12 mm (H) x 1 mm (diam.) column containing 15 mg of the strong anion exchanger Dowex 1 x 8. The retained pertechnetate was eluted with 5 ml of 0.5 mM tetrabutylammonium bromide (TBAB) solution in methylene chloride. The organic solvent was subsequently removed under vacuum and the dry residue was reconstituted with the desired volume of normal saline. All the operations were carried out in a closed cycle with sterile connecting tubes and multi-way stop-cocks. The concentrated pertechnetate obtained contained about 90% of the total saline eluate and had a pH of 6-7, a radiochemical purity (RCP) consistently > 99%. the 99Mo breakthrough was around 10-4%. Colorimetric spot tests (limit tests) for Al and Mo showed < 10 ppm in the final (TcO4-)-Tc-99m solution. The RCP of the labelled compounds prepared using the pertechnetate obtained and five different radiopharmaceutical kits were > 95%. The procedure may be equally applicable for the effective concentration of 188ReO(4)(-) from W-188-Re-188 generators. (c) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1295 / 1299
页数:5
相关论文
共 17 条
[1]   EXTENDING THE LIFE OF A 99TCM GENERATOR - A SIMPLE AND CONVENIENT METHOD FOR CONCENTRATING GENERATOR ELUATE FOR CLINICAL USE [J].
BLOWER, PJ .
NUCLEAR MEDICINE COMMUNICATIONS, 1993, 14 (11) :995-997
[2]  
BOYD RE, 1982, RADIOCHIM ACTA, V30, P123
[3]  
BOYD RE, 1987, RADIOCHIM ACTA, V41, P59
[4]   A novel 99mTc delivery system using (n,γ)99Mo adsorbed on a large alumina column in tandem with Dowex-1 and AgCl columns [J].
Chattopadhyay, S ;
Das, MK ;
Sarkar, SK ;
Saraswathy, P ;
Ramamoorthy, N .
APPLIED RADIATION AND ISOTOPES, 2002, 57 (01) :7-16
[5]  
Guhlke S., 1997, Journal of Nuclear Medicine, V38, p125P
[6]  
Guhlke S, 2000, J NUCL MED, V41, P1271
[7]  
GUHLKE S, 1998, J LABELLED COMPD RAD, V40, P274
[8]   PROCESSING OF REACTOR-PRODUCED W-188 FOR FABRICATION OF CLINICAL-SCALE ALUMINA-BASED W-188/RE-188 GENERATORS [J].
KNAPP, FF ;
CALLAHAN, AP ;
BEETS, AL ;
MIRZADEH, S ;
HSIEH, BT .
APPLIED RADIATION AND ISOTOPES, 1994, 45 (12) :1123-1128
[9]  
KNAPP FF, 1998, IAEA TECDOC, V1029, P485
[10]  
Knapp Jr F. F., 1998, IAEATECDOC1029, P419