Spectral Shaping Via Tin Prefiltration in Ultra-High-Resolution Photon-Counting and Energy-Integrating Detector CT of the Temporal Bone

被引:42
作者
Grunz, Jan-Peter [1 ]
Heidenreich, Julius Frederik [1 ]
Lennartz, Simon [2 ,3 ]
Weighardt, Jens Philipp [4 ]
Bley, Thorsten Alexander [1 ]
Erguen, Sueleyman [5 ]
Petritsch, Bernhard [1 ]
Huflage, Henner [1 ]
机构
[1] Univ Hosp Wurzburg, Dept Diagnost & Intervent Radiol, Oberdurrbacher Str 6, D-97080 Wurzburg, Germany
[2] Univ Cologne, Fac Med, Inst Diagnost & Intervent Radiol, Cologne, Germany
[3] Univ Cologne, Univ Hosp Cologne, Cologne, Germany
[4] Univ Hosp Wurzburg, Dept Otorhinolaryngol Plast Aesthet & Reconstruct, Wurzburg, Germany
[5] Univ Wurzburg, Inst Anat & Cell Biol, Wurzburg, Germany
关键词
temporal bone; tin prefiltration; photon-counting; dose reduction; COMPUTED-TOMOGRAPHY; IMAGE QUALITY; OPTIMIZATION; CHILDREN; FILTER;
D O I
10.1097/RLI.0000000000000901
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesHardening the x-ray beam, tin prefiltration is established for imaging of high-contrast subjects in energy-integrating detector computed tomography (EID-CT). With this work, we aimed to investigate the dose-saving potential of spectral shaping via tin prefiltration in photon-counting detector CT (PCD-CT) of the temporal bone.MethodsDeploying dose-matched scan protocols with and without tin prefiltration on a PCD-CT and EID-CT system (low-/intermediate-/full-dose: 4.8/7.6-7.7/27.0-27.1 mGy), 12 ultra-high-resolution examinations were performed on each of 5 cadaveric heads. While 120 kVp was applied for standard imaging, the protocols with spectral shaping used the highest potential available with tin prefiltration (EID-CT: Sn 150 kVp, PCD-CT: Sn 140 kVp). Contrast-to-noise ratios and dose-saving potential by spectral shaping were computed for each scanner. Three radiologists independently assessed the image quality of each examination with the intraclass correlation coefficient being computed to measure interrater agreement.ResultsRegardless of tin prefiltration, PCD-CT with low (171.2 +/- 10.3 HU) and intermediate radiation dose (134.7 +/- 4.5 HU) provided less image noise than full-dose EID-CT (177.0 +/- 14.2 HU; P < 0.001). Targeting matched image noise to 120 kVp EID-CT, mean dose reduction of 79.3% +/- 3.9% could be realized in 120 kVp PCD-CT. Subjective image quality of PCD-CT was better than of EID-CT on each dose level (P < 0.050). While no distinction was found between dose-matched PCD-CT with and without tin prefiltration (P >= 0.928), Sn 150 kVp EID-CT provided better image quality than 120 kVp EID-CT at high and intermediate dose levels (P > 0.050). The majority of low-dose EID-CT examinations were considered not diagnostic, whereas PCD-CT scans of the same dose level received satisfactory or better ratings. Interrater reliability was excellent (intraclass correlation coefficient 0.903).ConclusionsPCD-CT provides superior image quality and significant dose savings compared with EID-CT for ultra-high-resolution examinations of the temporal bone. Aiming for matched image noise, high-voltage scan protocols with tin prefiltration facilitate additional dose saving in EID-CT, whereas superior inherent denoising decreases the dose reduction potential of spectral shaping in PCD-CT.
引用
收藏
页码:819 / 825
页数:7
相关论文
共 31 条
[1]   A New Frontier in Temporal Bone Imaging: Photon-Counting Detector CT Demonstrates Superior Visualization of Critical Anatomic Structures at Reduced Radiation Dose [J].
Benson, J. C. ;
Rajendran, K. ;
Lane, J. I. ;
Diehn, F. E. ;
Weber, N. M. ;
Thorne, J. E. ;
Larson, N. B. ;
Fletcher, J. G. ;
McCollough, C. H. ;
Leng, S. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2022, :579-584
[2]   Computed tomography arthrography of the shoulder with tin filter-based spectral shaping at 100 kV and 140 kV [J].
Choi, Yun Seok ;
Choo, Hye Jung ;
Lee, Sun Joo ;
Kim, Dong Wook ;
Han, Ji-Yeon ;
Kim, Da Som .
ACTA RADIOLOGICA, 2021, 62 (10) :1349-1357
[3]   Photon-counting CT review [J].
Flohr, Thomas ;
Petersilka, Martin ;
Henning, Andre ;
Ulzheimer, Stefan ;
Ferda, Jiri ;
Schmidt, Bernhard .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 :126-136
[4]   Ultralow-Dose Chest Computed Tomography for Pulmonary Nodule Detection First Performance Evaluation of Single Energy Scanning With Spectral Shaping [J].
Gordic, Sonja ;
Morsbach, Fabian ;
Schmidt, Bernhard ;
Allmendinger, Thomas ;
Flohr, Thomas ;
Husarik, Daniela ;
Baumueller, Stephan ;
Raupach, Rainer ;
Stolzmann, Paul ;
Leschka, Sebastian ;
Frauenfelder, Thomas ;
Alkadhi, Hatem .
INVESTIGATIVE RADIOLOGY, 2014, 49 (07) :465-473
[5]   Effect of tin filter-based spectral shaping CT on image quality and radiation dose for routine use on ultralow-dose CT protocols: A phantom study [J].
Greffier, J. ;
Pereira, F. ;
Hamard, A. ;
Addala, T. ;
Beregi, J. P. ;
Frandon, J. .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2020, 101 (06) :373-381
[6]   Ultra-Low-Dose Photon-Counting CT Imaging of the Paranasal Sinus With Tin Prefiltration How Low Can We Go? [J].
Grunz, Jan-Peter ;
Petritsch, Bernhard ;
Luetkens, Karsten Sebastian ;
Kunz, Andreas Steven ;
Lennartz, Simon ;
Ergun, Suleyman ;
Bley, Thorsten Alexander ;
Huflage, Henner .
INVESTIGATIVE RADIOLOGY, 2022, 57 (11) :728-733
[7]   Image Quality Assessment for Clinical Cadmium Telluride-Based Photon-Counting Computed Tomography Detector in Cadaveric Wrist Imaging [J].
Grunz, Jan-Peter ;
Huflage, Henner ;
Heidenreich, Julius Frederik ;
Erguen, Suleyman ;
Petersilka, Martin ;
Allmendinger, Thomas ;
Bley, Thorsten Alexander ;
Petritsch, Bernhard .
INVESTIGATIVE RADIOLOGY, 2021, 56 (12) :785-790
[8]   Human Imaging With Photon Counting-Based Computed Tomography at Clinical Dose Levels: Contrast-to-Noise Ratio and Cadaver Studies [J].
Gutjahr, Ralf ;
Halaweish, Ahmed F. ;
Yu, Zhicong ;
Leng, Shuai ;
Yu, Lifeng ;
Li, Zhoubo ;
Jorgensen, Steven M. ;
Ritman, Erik L. ;
Kappler, Steffen ;
McCollough, Cynthia H. .
INVESTIGATIVE RADIOLOGY, 2016, 51 (07) :421-429
[9]   Metal Artifact Reduction With Tin Prefiltration in Computed Tomography A Cadaver Study for Comparison With Other Novel Techniques [J].
Hackenbroch, Carsten ;
Schuele, Simone ;
Halt, Daniel ;
Zengerle, Laura ;
Beer, Meinrad .
INVESTIGATIVE RADIOLOGY, 2022, 57 (03) :194-203
[10]   Dose Reduction in Dental CT: A Phantom Study With Special Focus on Tin Filter Technique [J].
Hackenbroch, Carsten ;
Schnaidt, Sebastian ;
Halt, Daniel ;
Wilde, Frank ;
Beer, Meinrad ;
Wunderlich, Arthur .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 215 (04) :945-953