Automated Image Registration for Stand-Alone Micro-PET and Micro-CT

被引:2
作者
Yang, Jing [1 ]
Li, Guang [1 ]
Sun, Yi [1 ]
Huang, Hongbo [2 ]
Zhao, Fukuan [2 ]
Luo, Shouhua [1 ]
机构
[1] Southeast Univ, Sch Biol Sci & Med Engn, Nanjing 210096, Jiangsu, Peoples R China
[2] Jiangsu Inst Nucl Med, Wuxi 214063, Jiangsu, Peoples R China
基金
中国国家自然科学基金;
关键词
Micro-PET; Micro-CT; Multi-Modality Image Fusion; Small Animal Imaging; CLOSED-FORM SOLUTION; GOLD NANORODS; IN-VIVO; FUSION; COREGISTRATION; QUATERNIONS;
D O I
10.1166/jnn.2016.11389
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Combination of micro Positron Emission Tomography with micro Computed Tomography (microPET/micro-CT) has been widely used in clinical and scientific research. Multimodality of microPET/micro-CT makes it a representative hybrid imaging device. However, its high prices hampered its widespread use. Previous work was concentrated on achieving complicated process of fusion images manually. Registration problem is a prerequisite for image fusion. Therefore, an effective method for accurate image registration between stand-alone micro-PET and micro-CT scanners is needed. Presented work proposes a multimodal images using separate micro-PET and microCT devices as well as implements a registration method by combining hardware and software. An animal chamber was designed which can be mounted on micro-PET and micro-CT scanners separately. Four catheters filled with (18)F2-deoxy-2-fluoro-D-glucose (F-18-FDG) solution were used to generate the spatial relation between PET and CT images. The transformed matrix was calculated during each registration experiment instead of a predetermined one, which decreased errors introduced by uncertainty in the position of animal chamber. In this way multi-modality biomedical images are acquired automatically without human intervention. These in vivo and test experimental results show the accuracy of this method. As stated above, this method is effective and robust for multi-modality image fusion.
引用
收藏
页码:6903 / 6909
页数:7
相关论文
共 50 条
  • [31] Methods of in-vivo mouse lung micro-CT
    Recheis, W
    Nixon, E
    Thiesse, J
    McLennan, G
    Ross, A
    Hoffman, EA
    MEDICAL IMAGING 2005: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, PTS 1 AND 2, 2005, 5746 : 40 - 49
  • [32] Cardiac Micro-PET-CT
    Croteau E.
    Renaud J.M.
    deKemp R.A.
    Current Cardiovascular Imaging Reports, 2013, 6 (2) : 179 - 190
  • [33] Assessment of scaffold porosity: the new route of micro-CT
    Bertoldi, Serena
    Fare, Silvia
    Tanzi, Maria Cristina
    JOURNAL OF APPLIED BIOMATERIALS & BIOMECHANICS, 2011, 9 (03) : 165 - 175
  • [34] Evaluation of microvascular anatomy by micro-CT
    Lerman, A
    Ritman, EL
    HERZ, 1999, 24 (07) : 531 - 533
  • [35] The effect of scatter correction on dual energy micro-CT
    Clark, D.
    Johnston, S. M.
    Johnson, G. A.
    Badea, C. T.
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [36] A comparison of sampling strategies for dual energy micro-CT
    Guo, Xiaolian
    Johnston, Samuel M.
    Johnson, G. Allan
    Badea, Cristian T.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [37] Micro-CT of fossils preserved in amber
    Dierick, M.
    Cnudde, V.
    Masschaele, B.
    Vlassenbroeck, J.
    Van Hoorebeke, L.
    Jacobs, P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01) : 641 - 643
  • [38] Micro-PET探测器单元的研究
    丁先利
    王金川
    祁辉荣
    肖国青
    肖志刚
    徐瑚珊
    孙志宇
    徐治国
    武丽杰
    胡正国
    郭忠言
    核技术, 2006, (05) : 388 - 391
  • [39] Study on characterization of micro-fracture of shale based on micro-CT
    Gou Q.
    Xu S.
    Hao F.
    Yang F.
    Wang Y.
    Lu Y.
    Zhang A.
    Cheng X.
    Qing J.
    Gao M.
    Dizhi Xuebao/Acta Geologica Sinica, 2019, 93 (09): : 2372 - 2382
  • [40] Research on modeling of skeletal composite voxel model based on micro-CT image
    Li, Jun-Li, 1600, Atomic Energy Press (48): : 773 - 779