A Monte Carlo investigation of the dose distribution for 60Co high dose rate brachytherapy source in water and in different media

被引:12
作者
Badry, Hamza [1 ]
Oufni, Lhoucine [1 ]
Ouabi, Hmad [2 ]
Hirayama, Hideo [3 ]
机构
[1] Sultan Moulay Slimane Univ, Fac Sci & Tech, Dept Phys, LPM ERM, BP 523, Beni Mellal 23000, Morocco
[2] Oncol Ctr Al Azhar, Rabat, Morocco
[3] High Energy Accelerator Res Org, KEK, Appl Res Lab, 1-1 Oho, Tsukuba, Ibaraki 3050801, Japan
关键词
Monte Carlo; Brachytherapy; Dose; Co-60; TECHNICAL NOTE;
D O I
10.1016/j.apradiso.2018.02.016
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In this study, the dosimetric characterization for The BEBIG Co-60 High Dose Rate (HDR) brachytherapy source model Co0.A86 was investigated and the validity of the EGS5 Monte Carlo code to reproduce the dosimetric parameters in water phantom was checked. In addition, the dose distribution for different tissue phantoms was calculated. The BEBIG Co-60 HDR brachytherapy source was modeled using EGS5 Monte Carlo simulation code. A description of the source design, geometry and materials used in this work were provided. According to the update TG43-U1 formalism of AAPM, the air kerma strength, the dose rate constant, 2D rectangular dose distribution in water were calculated, moreover, the results of the radial dose function were obtained in water and different tissue phantoms; bone, lung, adipose tissue, breast and muscle. The obtained results were tabulated and presented in graphical formats for the comparison with available data. The calculated value of the air kerma strength of this study, 3.0419 U Bq(-1), agree well with that of the other Monte Carlo calculation. The 2D look-up along-away rectangular dose were obtained in water, the results were similar to the published data for all distances larger than 1 an, for the distances near to the source region on the transversal source axis small differences are apparent. The radial dose function were presented in graphical format for the comparison between the dose distribution in water and different tissue phantoms. The EGS5 results obtained in this study shows good consistency with the published data for the dosimetric parameters of the of the BEBIG Co-60 HDR brachytherapy source. It seems that the radial dose function calculated in water differed in tissue phantoms due to the atomic composition and densities for media that are not taken account by the TG43-U1 formalism.
引用
收藏
页码:104 / 110
页数:7
相关论文
共 17 条
[1]  
[Anonymous], MED PHYS
[2]  
[Anonymous], SLACR730 EGS5
[3]  
[Anonymous], DLC 136 PHOTOX PHOT
[4]   On the use of HDR 60Co source with the MammoSite® Radiation Therapy System [J].
Baltas, D. ;
Lymperopoulou, G. ;
Zamboglou, N. .
MEDICAL PHYSICS, 2008, 35 (12) :5263-5268
[5]   Investigation of dose distribution of TheraSeed 200 103Pd brachytherapy seed on coaxial cylindrical phantom shells using Monte Carlo simulation [J].
Camgoz, Berkay ;
Kumru, Mehmet Nurullah .
ANNALS OF NUCLEAR MEDICINE, 2011, 25 (05) :371-379
[6]   Effect of tissue Inhomogeneities on dose distributions from Cf-252 brachytherapy source [J].
Ghassoun, J. .
APPLIED RADIATION AND ISOTOPES, 2013, 71 (01) :1-6
[7]   Technical note:: Dosimetric study of a. new Co-60 source used in brachytherapy [J].
Granero, D. ;
Perez-Calatayud, J. ;
Ballester, F. .
MEDICAL PHYSICS, 2007, 34 (09) :3485-3488
[8]   Dosimetric characterization of the 60Co BEBIG Co0.A86 high dose rate brachytherapy source using PENELOPE [J].
Guerrero, Rafael ;
Almansa, Julio F. ;
Torres, Javier ;
Lallena, Antonio M. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (08) :960-967
[9]  
Hubbell J.H., 1997, Tables of X-ray mass attenuation coefficients and mass energy-absorption coefficients (version 1.03), DOI [10.18434/T4D01F, DOI 10.18434/T4D01F]
[10]  
ICRU, 1984, Technical Report 37