Development of a Novel Breast MRI Phantom for Quality Control

被引:8
作者
Tuong, Betty [1 ]
Gardiner, Ian [1 ,2 ]
机构
[1] Univ British Columbia, Fac Med, Dept Radiol, Vancouver, BC V5Z 4E3, Canada
[2] Canada Diagnost Ctr, Breast Imaging Program, Vancouver, BC V6J 1B3, Canada
关键词
breast MRI; breast phantom; quality control; CANCER; TISSUE;
D O I
10.2214/AJR.12.9571
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. Many indications for breast MRI exist. MRI screening can identify pre-invasive breast cancer in women at high risk and in that regard is superior to mammography and ultrasound. Quality control standards exist for mammography and breast ultrasound screening with phantoms designed specifically for this purpose. Given the growing importance of breast MRI, we propose the development of a breast MRI phantom for quality control purposes. MATERIALS AND METHODS. A breast phantom with dual cavities containing water and fat was developed. A resolution plate inside the phantom contains various shapes ranging in size from 1 to 20 mm. Twenty studies of the phantom were performed with a 1.5-T system. STIR, T1-weighted fat-suppressed, and T2-weighted sequences were completed. Relaxation times of water and fat, number of step shapes resolved on STIR and T2-weighted images, number of circles resolved on T2-weighted images, and the diameter of a 20-mm circle on T1-weighted fat-suppressed images were recorded. RESULTS. On STIR images the TR of fat was 238.70 +/- 96.31 ms and of water was 1231.92 +/- 399.14 ms. On T2-weighted images the TR of fat was 778.73 +/- 62.60 ms and of water was 1737.60 +/- 121.63 ms. On STIR images, steps 3 mm and larger were visualized in 95% of instances. On T2-weighted images steps 3 mm and larger were seen in all instances. Measurements of a 20-mm circle were 19 +/- 0.3 mm. CONCLUSION. The proposed breast MRI phantom can be used to obtain reproducible measurements and allows implementation of quality control measures for a modality that is being increasingly used.
引用
收藏
页码:W511 / W515
页数:5
相关论文
共 17 条
[1]  
American College of Radiology, 2011, MAGN RES IM MRI ACCR
[2]  
American College of Radiology, 2011, BREAST MAGN RES IM M
[3]  
American College of Radiology, 2011, MAMM ACCR PROGR REQ
[4]  
American College of Radiology, 2011, BREAST ULTR ACCR PRO
[5]  
DeMartini Wendy, 2008, Top Magn Reson Imaging, V19, P143, DOI 10.1097/RMR.0b013e31818a40a5
[6]   An anthropomorphic phantom for quantitative evaluation of breast MRI [J].
Freed, Melanie ;
de Zwart, Jacco A. ;
Loud, Jennifer T. ;
El Khouli, Riham H. ;
Myers, Kyle J. ;
Greene, Mark H. ;
Duyn, Jeff H. ;
Badano, Aldo .
MEDICAL PHYSICS, 2011, 38 (02) :743-753
[7]  
Hatakenaka Masamitsu, 2008, Magn Reson Med Sci, V7, P23, DOI 10.2463/mrms.7.23
[8]   Diffusion-Weighted Imaging of Breast Cancer: Correlation of the Apparent Diffusion Coefficient Value with Prognostic Factors [J].
Kim, Sung Hun ;
Cha, Eun Suk ;
Kim, Hyeon Sook ;
Kang, Bong Joo ;
Choi, Jae Jeong ;
Jung, Ji Han ;
Park, Yong Gyu ;
Suh, Young Jin .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 30 (03) :615-620
[9]   The current status of breast MR imaging - Part I. Choice of technique, image interpretation, diagnostic accuracy, and transfer to clinical practice [J].
Kuhl, Christiane .
RADIOLOGY, 2007, 244 (02) :356-378
[10]   Prospective Multicenter Cohort Study to Refine Management Recommendations for Women at Elevated Familial Risk of Breast Cancer: The EVA Trial [J].
Kuhl, Christiane ;
Weigel, Stefanie ;
Schrading, Simone ;
Arand, Birke ;
Bieling, Heribert ;
Koenig, Roy ;
Tombach, Bernd ;
Leutner, Claudia ;
Rieber-Brambs, Andrea ;
Nordhoff, Dennis ;
Heindel, Walter ;
Reiser, Maximilian ;
Schild, Hans H. .