Investigation of gated cone-beam CT to reduce respiratory motion blurring

被引:17
|
作者
Kincaid, Russell E., Jr. [1 ]
Yorke, Ellen D. [1 ]
Goodman, Karyn A. [2 ]
Rimner, Andreas [2 ]
Wu, Abraham J. [2 ]
Mageras, Gig S. [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10065 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10065 USA
关键词
image guided radiation therapy; cone-beam CT; respiratory motion; motion management; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; LUNG-TUMORS; QUALITY-ASSURANCE; CANCER; LIVER;
D O I
10.1118/1.4795336
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Methods of reducing respiratory motion blurring in cone-beam CT (CBCT) have been limited to lung where soft tissue contrast is large. Respiration-correlated cone-beam CT uses slow continuous gantry rotation but image quality is limited by uneven projection spacing. This study investigates the efficacy of a novel gated CBCT technique. Methods: In gated CBCT, the linac is programmed such that gantry rotation and kV image acquisition occur within a gate around end expiration and are triggered by an external respiratory monitor. Standard CBCT and gated CBCT scans are performed in 22 patients (11 thoracic, 11 abdominal) and a respiration-correlated CT (RCCT) scan, acquired on a standard CT scanner, from the same day serves as a criterion standard. Image quality is compared by calculating contrast-to-noise ratios (CNR) for tumors in lung, gastroesophageal junction (GEJ) tissue, and pancreas tissue, relative to surrounding background tissue. Congruence between the object in the CBCT images and that in the RCCT is measured by calculating the optimized normalized cross-correlation (NCC) following CBCT-to-RCCT rigid registrations. Results: Gated CBCT results in reduced motion artifacts relative to standard CBCT, with better visualization of tumors in lung, and of abdominal organs including GEJ, pancreas, and organs at risk. CNR of lung tumors is larger in gated CBCT in 6 of 11 cases relative to standard CBCT. A paired two-tailed t-test of lung patient mean CNR shows no statistical significance (p = 0.133). In 4 of 5 cases where CNR is not increased, lung tumor motion observed in RCCT is small (range 1.3-5.2 mm). CNR is increased and becomes statistically significant for 6 out of 7 lung patients with > 5 mm tumor motion (p = 0.044). CNR is larger in gated CBCT in 5 of 7 GEE cases and 3 of 4 pancreas cases (p = 0.082 and 0.192). Gated CBCT yields improvement with lower NCC relative to standard CBCT in 10 of 11, 7 of 7, and 3 of 4 patients for lung, GEJ, and pancreas images, respectively (p = 0.0014, 0.0030, 0.165). Conclusions: Gated CBCT reduces image blurring caused by respiratory motion. The gated gantry rotation yields uniformly and closely spaced projections resulting in improved reconstructed image quality. The technique is shown to be applicable to abdominal sites, where image contrast of soft tissues is low. (C) 2013 American Association of Physicists in Medicine. [http ://dx.doi.org/10.1118/1.4795336]
引用
收藏
页数:11
相关论文
共 50 条
  • [41] A motion algorithm to extract physical and motion parameters of mobile targets from cone-beam computed tomographic images
    Alsbou, Nesreen
    Ahmad, Salahuddin
    Ali, Imad
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (04) : 599 - 613
  • [42] AUTOMATIC MOTION CORRECTION IN CONE-BEAM COMPUTED TOMOGRAPHY
    Ens, S.
    Ulrici, J.
    Hell, E.
    Buzug, T. M.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3248 - 3251
  • [43] Investigation on substantial asymmetry of HU value in dental cone-beam CT with offset detector
    Yun, Sungho
    Choi, Dain
    Ye, Sung-Joon
    Cho, Gyuseong
    Cho, Seungryon
    MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1, 2024, 12925
  • [44] Motion Vector Field Upsampling for Improved 4D Cone-Beam CT Motion Compensation of the Thorax
    Sauppe, Sebastian
    Rank, Christopher M.
    Brehm, Marcus
    Paysan, Pascal
    Seghers, Dieter
    Kachelriess, Marc
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [45] kV Cone-Beam CT-Based IGRT A Clinical Review
    Boda-Heggemann, Judit
    Lohr, Frank
    Wenz, Frederik
    Flentje, Michael
    Guckenberger, Matthias
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (05) : 284 - 291
  • [46] Monte Carlo evaluation of scatter mitigation strategies in cone-beam CT
    Lazos, Dimitrios
    Williamson, Jeffrey F.
    MEDICAL PHYSICS, 2010, 37 (10) : 5456 - 5470
  • [47] Two methods for reducing moving metal artifacts in cone-beam CT
    Hahn, Andreas
    Knaup, Michael
    Brehm, Marcus
    Sauppe, Sebastian
    Kachelriess, Marc
    MEDICAL PHYSICS, 2018, 45 (08) : 3671 - 3680
  • [48] Prostate Image-Guided Radiotherapy by Megavolt Cone-Beam CT
    Zucca, Sergio
    Carau, Barbara
    Solla, Ignazio
    Garibaldi, Elisabetta
    Farace, Paolo
    Lay, Giancarlo
    Meleddu, Gianfranco
    Gabriele, Pietro
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (08) : 473 - 478
  • [49] Interleaved acquisition for cross scatter avoidance in dual cone-beam CT
    Giles, William
    Bowsher, James
    Li, Hao
    Yin, Fang-Fang
    MEDICAL PHYSICS, 2012, 39 (12) : 7719 - 7728
  • [50] Image-Based Motion Compensation for High-Resolution Extremities Cone-Beam CT
    Sisniega, A.
    Stayman, J. W.
    Cao, Q.
    Yorkston, J.
    Siewerdsen, J. H.
    Zbijewski, W.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783