Investigation of gated cone-beam CT to reduce respiratory motion blurring

被引:17
|
作者
Kincaid, Russell E., Jr. [1 ]
Yorke, Ellen D. [1 ]
Goodman, Karyn A. [2 ]
Rimner, Andreas [2 ]
Wu, Abraham J. [2 ]
Mageras, Gig S. [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10065 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10065 USA
关键词
image guided radiation therapy; cone-beam CT; respiratory motion; motion management; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; LUNG-TUMORS; QUALITY-ASSURANCE; CANCER; LIVER;
D O I
10.1118/1.4795336
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Methods of reducing respiratory motion blurring in cone-beam CT (CBCT) have been limited to lung where soft tissue contrast is large. Respiration-correlated cone-beam CT uses slow continuous gantry rotation but image quality is limited by uneven projection spacing. This study investigates the efficacy of a novel gated CBCT technique. Methods: In gated CBCT, the linac is programmed such that gantry rotation and kV image acquisition occur within a gate around end expiration and are triggered by an external respiratory monitor. Standard CBCT and gated CBCT scans are performed in 22 patients (11 thoracic, 11 abdominal) and a respiration-correlated CT (RCCT) scan, acquired on a standard CT scanner, from the same day serves as a criterion standard. Image quality is compared by calculating contrast-to-noise ratios (CNR) for tumors in lung, gastroesophageal junction (GEJ) tissue, and pancreas tissue, relative to surrounding background tissue. Congruence between the object in the CBCT images and that in the RCCT is measured by calculating the optimized normalized cross-correlation (NCC) following CBCT-to-RCCT rigid registrations. Results: Gated CBCT results in reduced motion artifacts relative to standard CBCT, with better visualization of tumors in lung, and of abdominal organs including GEJ, pancreas, and organs at risk. CNR of lung tumors is larger in gated CBCT in 6 of 11 cases relative to standard CBCT. A paired two-tailed t-test of lung patient mean CNR shows no statistical significance (p = 0.133). In 4 of 5 cases where CNR is not increased, lung tumor motion observed in RCCT is small (range 1.3-5.2 mm). CNR is increased and becomes statistically significant for 6 out of 7 lung patients with > 5 mm tumor motion (p = 0.044). CNR is larger in gated CBCT in 5 of 7 GEE cases and 3 of 4 pancreas cases (p = 0.082 and 0.192). Gated CBCT yields improvement with lower NCC relative to standard CBCT in 10 of 11, 7 of 7, and 3 of 4 patients for lung, GEJ, and pancreas images, respectively (p = 0.0014, 0.0030, 0.165). Conclusions: Gated CBCT reduces image blurring caused by respiratory motion. The gated gantry rotation yields uniformly and closely spaced projections resulting in improved reconstructed image quality. The technique is shown to be applicable to abdominal sites, where image contrast of soft tissues is low. (C) 2013 American Association of Physicists in Medicine. [http ://dx.doi.org/10.1118/1.4795336]
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Investigation of effect of reconstruction filters on cone-beam computed tomography image quality
    Srinivasan, Kavitha
    Mohammadi, Mohammad
    Shepherd, Justin
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (03) : 607 - 614
  • [32] Estimation of Local Deformable Motion in Image-Based Motion Compensation for Interventional Cone-Beam CT
    Sisniega, A.
    Capostagno, S.
    Zbijewski, W.
    Stayman, J. W.
    Weiss, C. R.
    Ehtiati, T.
    Siewerdsen, J. H.
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [33] Integration of cone-beam CT in stereotactic body radiation therapy
    Yin, Fang-Fang
    Wang, Zhiheng
    Yoo, Sua
    Wu, Q. Jackie
    Kirkpatrick, John
    Larrier, Nicole
    Meyer, Jeffrey
    Willett, Christopher G.
    Marks, Lawrence B.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2008, 7 (02) : 133 - 139
  • [34] Cone-beam CT reconstruction for planar object
    Liu, Tong
    NDT & E INTERNATIONAL, 2012, 45 (01) : 9 - 15
  • [35] Motion vector field phase-to-amplitude resampling for 4D motion-compensated cone-beam CT
    Sauppe, Sebastian
    Kuhm, Julian
    Brehm, Marcus
    Paysan, Pascal
    Seghers, Dieter
    Kachelriess, Marc
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (03):
  • [36] Development of a 4D phantom for respiratory motion modeling during Cone-Beam CT (CBCT) imaging on the Varian On-Board Imager (OBI)
    Amin, Adam Tan Mohd
    Mokri, Siti Salasiah
    Ahmad, Rozilawati
    Abd Rahni, Ashrani Aizzuddin
    2018 IEEE-EMBS CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES), 2018, : 451 - 456
  • [37] Unsupervised motion artifacts reduction for cone-beam CT via enhanced landmark detection
    Viriyasaranon, Thanaporn
    Ma, Serie
    Thies, Mareike
    Maier, Andreas
    Choi, Jang-Hwan
    EXPERT SYSTEMS WITH APPLICATIONS, 2025, 278
  • [38] Reconstruction of implanted marker trajectories from cone-beam CT projection images using interdimensional correlation modeling
    Chung, Hyekyun
    Poulsen, Per Rugaard
    Keall, Paul J.
    Cho, Seungryong
    Cho, Byungchul
    MEDICAL PHYSICS, 2016, 43 (08) : 4643 - 4654
  • [39] Calculating tumor trajectory and dose-of-the-day using cone-beam CT projections
    Jones, Bernard L.
    Westerly, David
    Miften, Moyed
    MEDICAL PHYSICS, 2015, 42 (02) : 694 - 702
  • [40] An image-based method to synchronize cone-beam CT and optical surface tracking
    Fassi, Aurora
    Schaerer, Joel
    Riboldi, Marco
    Sarrut, David
    Baroni, Guido
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (02): : 117 - 128