Dual-slice spiral versus single-slice spiral scanning: Comparison of the physical performance of two computed tomography scanners

被引:66
作者
Liang, Y
Kruger, RA
机构
[1] Department of Radiology, Indiana University, School of Medicine, Indianapolis
关键词
spiral CT; dual-slice technology; pitch; resolution; modulation transfer function; interpolation;
D O I
10.1118/1.597705
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper we deal with two types of spiral scanners; one is a single-slice spiral scanner, while the other employs dual-slice technology into spiral scanning. Physical performance parameters, including image noise, contrast resolution, spatial resolution (transversal and longitudinal), and radiation exposure are measured. Computer simulations based on two interpolation methods (180 degrees and 360 degrees linear interpolation) are also used in evaluating the slice-sensitivity profile (SSP) and noise. The results show that the noise behaves in the same way for both types of scanners. The noise change, relative to that of the standard scan with the same scanning parameters, depends solely on the interpolation algorithm. Table speed and scanner geometry (either single slice or dual slice) have no effect on the noise value. For the given table speed, as well as individual detector collimation (slice width) the dual-slice scan results in better longitudinal resolution (SSP) compared to a single-slice scan if the scan is obtained with nonoverlapping slices (pitch greater than 2). This is because the dual-slice scan obtains twice the number of nonoverlapped projections for the same length, which reduces the degradation of the slice profile by using more densely arranged projections (in the longitudinal direction) for the interpolation. In the dual-slice scanner the workable scan rate is extended up to pitch 4 compared to a pitch of 2 for the single-slice scanner. Therefore, the dual-slice spiral scanner is preferred in applications requiring an increased scan rate with comparative image quality. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:205 / 220
页数:16
相关论文
共 11 条
  • [11] Diagnostic Performance of 64-versus 256-Slice Computed Tomography Coronary Angiography Compared with Conventional Coronary Angiography in Patients with Suspected Coronary Artery Disease
    Chua, Su-Kiat
    Hung, Huei-Fong
    Cheng, Jun-Jack
    Tseng, Min-Tsung
    Law, Wai-Yip
    Kuo, Chu-Jen
    Chiu, Chiung-Zuan
    Chang, Che-Ming
    Lee, Shih-Huang
    Lo, Huey-Ming
    Lin, Sheng-Chang
    Liou, Jer-Young
    Shyu, Kou-Gi
    ACTA CARDIOLOGICA SINICA, 2013, 29 (02) : 151 - 159