Dual-slice spiral versus single-slice spiral scanning: Comparison of the physical performance of two computed tomography scanners

被引:66
作者
Liang, Y
Kruger, RA
机构
[1] Department of Radiology, Indiana University, School of Medicine, Indianapolis
关键词
spiral CT; dual-slice technology; pitch; resolution; modulation transfer function; interpolation;
D O I
10.1118/1.597705
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper we deal with two types of spiral scanners; one is a single-slice spiral scanner, while the other employs dual-slice technology into spiral scanning. Physical performance parameters, including image noise, contrast resolution, spatial resolution (transversal and longitudinal), and radiation exposure are measured. Computer simulations based on two interpolation methods (180 degrees and 360 degrees linear interpolation) are also used in evaluating the slice-sensitivity profile (SSP) and noise. The results show that the noise behaves in the same way for both types of scanners. The noise change, relative to that of the standard scan with the same scanning parameters, depends solely on the interpolation algorithm. Table speed and scanner geometry (either single slice or dual slice) have no effect on the noise value. For the given table speed, as well as individual detector collimation (slice width) the dual-slice scan results in better longitudinal resolution (SSP) compared to a single-slice scan if the scan is obtained with nonoverlapping slices (pitch greater than 2). This is because the dual-slice scan obtains twice the number of nonoverlapped projections for the same length, which reduces the degradation of the slice profile by using more densely arranged projections (in the longitudinal direction) for the interpolation. In the dual-slice scanner the workable scan rate is extended up to pitch 4 compared to a pitch of 2 for the single-slice scanner. Therefore, the dual-slice spiral scanner is preferred in applications requiring an increased scan rate with comparative image quality. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:205 / 220
页数:16
相关论文
共 11 条
  • [1] Single-slice rebinning reconstruction in spiral cone-beam computed tomography
    Bruder, H
    Kachelriess, M
    Schaller, S
    Stierstorfer, K
    Flohr, T
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (09) : 873 - 887
  • [2] COMPARISON OF SINGLE-SLICE COMPUTED TOMOGRAPHY PROTOCOLS FOR DETECTION OF PULMONARY NODULES IN DOGS
    Joly, Hugo
    d'Anjou, Marc-Andre
    Alexander, Kate
    Beauchamp, Guy
    VETERINARY RADIOLOGY & ULTRASOUND, 2009, 50 (03) : 279 - 284
  • [3] Advanced single-slice rebinning in cone-beam spiral CT
    Kachelriess, M
    Schaller, S
    Kalender, WA
    MEDICAL PHYSICS, 2000, 27 (04) : 754 - 772
  • [4] Fourier-based approach to interpolation in single-slice helical computed tomography
    La Rivière, PJ
    Pan, XC
    MEDICAL PHYSICS, 2001, 28 (03) : 381 - 392
  • [5] Favorable noise uniformity properties of Fourier-based interpolation and reconstruction approaches in single-slice helical computed tomography
    La Rivière, PJ
    Pan, XC
    MEDICAL PHYSICS, 2002, 29 (06) : 943 - 951
  • [6] Evaluation of spiral computed tomography versus ventilation/perfusion scanning in patients clinically suspected of pulmonary embolism
    Katsouda, E
    Mystakidou, K
    Rapti, A
    Argentos, S
    Kelekis, NL
    Mastorakou, R
    Anagnostopoulou, O
    Kelekis, DA
    IN VIVO, 2005, 19 (05): : 873 - 878
  • [7] Temporal resolution measurement of 128-slice dual source and 320-row area detector computed tomography scanners in helical acquisition mode using the impulse method
    Hara, Takanori
    Urikura, Atsushi
    Ichikawa, Katsuhiro
    Hoshino, Takashi
    Nishimaru, Eiji
    Niwa, Shinji
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (04): : 625 - 630
  • [8] Dose and image quality of high-pitch dual source computed tomography for the evaluation of cervical lymph node status - Comparison to regular 128-slice single source computed tomography
    Bodelle, Boris
    Bauer, Ralf W.
    Holthaus, Lara
    Schulz, Boris
    Al-Butmeh, Firas
    Wichmann, Julian L.
    Beeres, Martin
    Vogl, Thomas J.
    Kerl, J. Matthias
    EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (06) : E281 - E285
  • [9] Assessment of image quality of two cone-beam computed tomography of the Varian Linear accelerators: Comparison with spiral CT simulator
    Ragab, H.
    Abdelaziz, D. M.
    Khalil, M. M.
    Elbakry, M. N. Yasein
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (03): : 491 - 498
  • [10] Patient doses from X-ray computed tomography examinations by a single-array detector unit: Axial versus spiral mode
    Ghavami, S. M.
    Mesbahi, A.
    Pesianian, I.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2012, 10 (02): : 89 - 94