System Architecture of the LabPET Small Animal PET Scanner

被引:32
作者
Tetrault, Mare-Andre [1 ]
Viscogliosi, Nicolas [1 ]
Riendeau, Joel [1 ]
Belanger, Francois [1 ]
Michaud, Jean-Baptiste [1 ]
Semmaoui, Hicham [1 ]
Berard, Philippe [2 ]
Lemieux, Francois [1 ]
Arpin, Louis [1 ]
Bergeron, Melanie [2 ]
Cadorette, Jules [2 ]
Pepin, Catherine M. [2 ]
Robert, Ghislain [3 ]
Lepage, Martin D. [3 ]
Lecomte, Roger [2 ]
Fontaine, Rejean [1 ]
机构
[1] Univ Sherbrooke, Dept Elect & Comp Engn, Sherbrooke, PQ J1K 2R1, Canada
[2] Univ Sherbrooke, Dept Med Nucl & Radiobiol, Sherbrooke, PQ J1K 2R1, Canada
[3] Gamma Med Ideas Inc, Sherbrooke, PQ J1K 2R1, Canada
基金
加拿大健康研究院; 加拿大自然科学与工程研究理事会;
关键词
Positron Emission Tomography (PET);
D O I
10.1109/TNS.2008.2002326
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
To address modern molecular imaging requirements, a digital positron emission tomography (PET) scanner for small animals has been developed at Universite de Sherbrooke. Based on individual readout of avalanche photodiodes (APD) coupled to LYSO/LGSO phoswich detectors, the scanner supports up to 4608 channels in a 16.2 cm diameter, 11.25 cm axial field of view with an isotropic similar to 1.2 mm FWHM intrinsic spatial resolution at the center of the field of view. Custom data acquisition boards preprocess and sample APD signals at 45 MHz and compute in real time crystal identification, energy and timing information of detected events at an average sustained rate of up to 1250 raw counts per second per mm(2) (10 000 cps/channel). Real time digital signal analysis also filters out events outside the pre-selected energy window with crystal granularity to eliminate Compton events and electronic noise. Retained events are then merged into a single stream through a real-time sorting tree, at which end prompt and delayed coincidences are extracted. A single Firewire link handles both control and data transfers with a host computer. The LabPET features four data recording modes, giving the user the choice to retain data for research or to minimize file size for high coincidence count rate and imaging purposes. The electronic system also supports time synchronized data insertion for flags such as vital signs used in gated image reconstruction. Aside from data acquisition, hardware can generate live energy and discrimination spectra suitable for fast, automatic channel calibration.
引用
收藏
页码:2546 / 2550
页数:5
相关论文
共 19 条
  • [1] Performance evaluation of the LabPET™ APD-based digital PET scanner
    Bergeron, Melanie
    Cadorette, Jules
    Beaudoin, Jean-Francois
    Rousseau, Jacques A.
    Dumoulin, Marc
    Lepage, Martin D.
    Robert, Ghislain
    Selivanov, Vitali
    Tetrault, Marc-Andre
    Viscogliosi, Nicolas
    Dumouchel, Tyler
    Thorn, Stephanie
    DaSilva, Jean
    deKemp, Robert A.
    Norenberg, Jeffrey P.
    Fontaine, Rejean
    Lecomte, Roger
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4185 - +
  • [2] A microvolumetric β blood counter for pharmacokinetic PET studies in small animals
    Convert, L.
    Morin-Brassard, G.
    Cadorette, J.
    Rouleau, D.
    Croteau, E.
    Archambault, M.
    Fontaine, R.
    Lecomte, R.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (01) : 173 - 180
  • [3] Real time digital signal processing implementation for an APD-based PET scanner with phoswich detectors
    Fontaine, R.
    Tetrault, M. -A.
    Belanger, F.
    Viscogliosi, N.
    Himmich, R.
    Michaud, J. -B.
    Robert, S.
    Leroux, J. -D.
    Semmaoui, H.
    Berard, P.
    Cadorette, J.
    Pepin, C. M.
    Lecomte, R.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) : 784 - 788
  • [4] FONTAINE R, 2005, P IEEE NSS MIC C REC, V5, P2785
  • [5] Digital signal processing applied to crystal identification in Positron Emission Tomography dedicated to small animals
    Fontaine, Rejean
    Viscogliosi, Nicolas
    Semmaoui, Hicham
    Belanger, Francois
    Lemieux, Francois
    Tetrault, Marc-Andre
    Michaud, Jean-Baptiste
    Berard, Philippe
    Cadorette, Jules
    Pepin, Catherine M.
    Lecomte, Roger
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) : 385 - 388
  • [6] Next generation of real time data acquisition, calibration and control system for the RatCAP scanner
    Junnarkar, Sachin S.
    Fried, Jack
    Southekal, Sudeepti
    Pratte, Jean-Francois
    O'Connor, Paul
    Radeka, Veljko
    Vaska, Paul
    Purschke, Martin
    Tornasi, Dardo
    Woody, Craig
    Fontaine, Rejean
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (01) : 220 - 224
  • [7] Simplified FPGA-based data acquisition system for PET
    Laymon, CM
    Miyaoka, RS
    Park, BK
    Lewellen, TK
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) : 1483 - 1486
  • [8] LECOMTE R, 2006, J NUCL MED S, V47, pP194
  • [9] LEROUX JD, IEEE T NUCL IN PRESS
  • [10] A sampling ADC data acquisition system for positron emission tomography
    Mann, A
    Grube, B
    Konorov, I
    Paul, S
    Schmitt, L
    McElroy, DP
    Ziegler, SI
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 297 - 303