Background: Recording surface electromyogram (EMG) signals from sphincter muscles has become of increasing interest due to potential applications in diagnosis or investigation of the mechanisms of incontinence. Recently developed probes allow high-resolution detection of EMG signal from the external anal sphincter. One of the main issues in the interpretation of experimental signals is the detection volume of the recording electrodes. Methods: An analytical model of generation of EMG from the anal sphincter was applied to simulate single fiber action potentials varying the fiber length, conductivity and thickness of the mucosa, shape of the electrodes, and detection system (monopolar, bipolar, and double differential). Results: The new probe is more selective than classical probes with big electrodes and large interelectrode distance. The decay of the recorded surface EMG action potentials with distance of the fiber was faster for increasing depth than axial distance (this is due to the geometry of the volume conductor, as indicated by comparison with simulations with a planar volume conductor). Point electrodes led to smaller detection volumes than rectangular electrodes, and the double differential system was the most selective recording configuration. The relative decrease of the detected potential generated by more distant fibers was lower for increasing thickness and conductivity of the mucosa. End-of-fiber components are enhanced by the circular geometry of sphincter muscles with respect to the case of a planar volume conductor. Conclusions: The detection volume of EMG recording systems from sphincter muscles extends more in the axial than in the depth direction and is affected by many parameters that cannot be estimated in practical situations, thus introducing a rather large variability in the muscle portion investigated among subjects. More selective information can be obtained using the recently developed multi-channel probe with respect to classical probes. Selectivity is increased with small electrodes and double differential recordings. (C) 2007 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:896 / 904
页数:9
相关论文
共 17 条
  • [1] SURFACE MYOELECTRIC SIGNAL CROSS-TALK AMONG MUSCLES OF THE LEG
    DE LUCA, CJ
    MERLETTI, R
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1988, 69 (06): : 568 - 575
  • [2] Simulation analysis of the ability of different types of multi-electrodes to increase selectivity of detection and to reduce cross-talk
    Dimitrov, GV
    Disselhorst-Klug, C
    Dimitrova, NA
    Schulte, E
    Rau, G
    [J]. JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2003, 13 (02) : 125 - 138
  • [3] Neither high-pass filtering nor mathematical differentiation of the EMG signals can considerably reduce cross-talk
    Dimitrova, NA
    Dimitrov, GV
    Nikitin, OA
    [J]. JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2002, 12 (04) : 235 - 246
  • [4] Improvement of spatial resolution in surface-EMG: A theoretical and experimental comparison of different spatial filters
    DisselhorstKlug, C
    Silny, J
    Rau, G
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (07) : 567 - 574
  • [5] Anal sphincter morphology in patients with full-thickness rectal prolapse
    Dvorkin, LS
    Chan, CLH
    Knowles, CH
    Williams, NS
    Lunniss, PJ
    Scott, SM
    [J]. DISEASES OF THE COLON & RECTUM, 2004, 47 (02) : 198 - 203
  • [6] The external anal sphincter and the role of surface electromyography
    Enck, P
    Hinninghofen, H
    Merletti, R
    Azpiroz, F
    [J]. NEUROGASTROENTEROLOGY AND MOTILITY, 2005, 17 : 60 - 67
  • [7] A surface EMG generation model with multilayer cylindrical description of the volume conductor
    Farina, D
    Mesin, L
    Martina, S
    Merletti, R
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2004, 51 (03) : 415 - 426
  • [8] Comparison of spatial filter selectivity in surface myoelectric signal detection: influence of the volume conductor model
    Farina, D
    Mesin, L
    Martina, S
    Merletti, R
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2004, 42 (01) : 114 - 120
  • [9] Influence of anatomical, physical, and detection-system parameters on surface EMG
    Farina, D
    Cescon, C
    Merletti, R
    [J]. BIOLOGICAL CYBERNETICS, 2002, 86 (06) : 445 - 456
  • [10] A novel approach for precise simulation of the EMG signal detected by surface electrodes
    Farina, D
    Merletti, R
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (06) : 637 - 646