Can Microcomputed Tomography Measure Retrieved Polyethylene Wear? Comparing Fixed-bearing and Rotating-platform Knees

被引:24
作者
Engh, Charles A., Jr. [1 ,2 ]
Zimmerman, Rebecca L. [1 ]
Hopper, Robert H., Jr. [1 ]
Engh, Gerard A. [1 ,2 ]
机构
[1] Anderson Orthopaed Res Inst, Alexandria, VA 22307 USA
[2] Mt Vernon Hosp, Inova Ctr Joint Replacement, Alexandria, VA USA
关键词
TIBIAL INSERTS; VOLUMETRIC WEAR; MOBILE-BEARING; FOLLOW-UP; ARTHROPLASTY; THICKNESS; QUANTIFICATION; PROSTHESIS;
D O I
10.1007/s11999-012-2513-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Wear of total knee polyethylene has been quantified gravimetrically with thickness measurements and evaluation of surface wear modes. However, these techniques do not localize volumetric wear. We used micro-CT scans of retrieved total knee liners and unworn, new liners to determine the volume and location of wear. We retrieved 12 fixed and 12 rotating-platform bearings after a mean 52 months of use. Inserts were weighed and thickness was measured. Micro-CT scans of retrieved and matched new liners were superimposed to compare the location and magnitude of wear. The average total wear was 254 +/- A 248 mm(3). The average wear rate was 58 +/- A 41 mm(3)/year. Wear was 69% of penetration, demonstrating the contribution of deformation to knee wear. Rotating-platform wear rate was 43 +/- A 25 mm(3)/year and the fixed-bearing rate was 74 +/- A 49 mm(3)/year. Five percent of the rotating-platform wear rate came from the backside compared with 14% of the fixed-bearing wear rate. Micro-CT can determine the volume and location of wear of retrieved tibial liners. Because the magnitude of the manufacturing tolerances was approximately half the magnitude of the total wear on average, accounting for the potential influence of tolerances is important to accurately measure volumetric wear if the unworn (preimplantation) geometry of the insert is unknown. Without accounting for tolerances, this technique may not be applicable for retrievals with a short followup that have low wear. However, application of micro-CT could be of value in determining the exact location of wear in knee simulator studies in which the same insert is measured repeatedly and manufacturing tolerances are not a concern.
引用
收藏
页码:86 / 93
页数:8
相关论文
共 22 条
[1]   Clinical wear measurement on low contact stress rotating platform knee bearings [J].
Atwood, Sara A. ;
Currier, John H. ;
Mayor, Michael B. ;
Collier, John P. ;
Van Citters, Douglas W. ;
Kennedy, Francis E. .
JOURNAL OF ARTHROPLASTY, 2008, 23 (03) :431-440
[2]  
Benjamin J, 2001, CLIN ORTHOP RELAT R, P131
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Wear in retrieved condylar knee arthroplasties - A comparison of wear in different designs of 280 retrieved condylar knee prostheses [J].
Blunn, GW ;
Joshi, AB ;
Minns, RJ ;
Lidgren, L ;
Lilley, P ;
Ryd, L ;
Engelbrecht, E ;
Walker, PS .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :281-290
[5]   Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners [J].
Bowden, AE ;
Kurtz, SM ;
Edidin, AA .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2005, 75B (01) :205-209
[6]   Radiographic assessment of the thickness lost from polyethylene tibial inserts that had been sterilized differently [J].
Collier, Matthew B. ;
Engh, C. Anderson, Jr. ;
Hatten, Kyle M. ;
Ginn, Stuart D. ;
Sheils, Todd M. ;
Engh, Gerard A. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (07) :1543-1552
[7]   Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty [J].
Collier, Matthew B. ;
Engh, C. Anderson, Jr. ;
McAuley, James P. ;
Engh, Gerard A. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (06) :1306-1314
[8]   Clinical assessment of tibial polyethylene thickness [J].
Collier, MB ;
Jewett, BA ;
Engh, CA .
JOURNAL OF ARTHROPLASTY, 2003, 18 (07) :860-866
[9]   Backside wear of polyethylene tibial inserts: Mechanism and magnitude of material loss [J].
Conditt, MA ;
Thompson, MT ;
Usrey, MM ;
Ismaily, SK ;
Noble, PC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (02) :326-331
[10]   Analysis of Wear in Retrieved Mobile and Fixed Bearing Knee Inserts [J].
Engh, Gerard A. ;
Zimmerman, Rebecca L. ;
Parks, Nancy L. ;
Engh, C. Anderson .
JOURNAL OF ARTHROPLASTY, 2009, 24 (06) :28-32