MR imaging of lung parenchyma at 0.2 T: evaluation of imaging techniques, comparative study with chest radiography and interobserver analysis

被引:14
作者
Abolmaali, ND
Schmitt, J
Krauss, S
Bretz, F
Deimling, M
Jacobi, V
Vogl, TJ
机构
[1] Goethe Univ Frankfurt, Inst Diagnost & Intervent Radiol, D-60590 Frankfurt, Germany
[2] Leibniz Univ Hannover, Dept Bioinformat, D-30419 Hannover, Germany
[3] Siemens Med Solut, D-91052 Erlangen, Germany
关键词
magnetic resonance; low-field-strength imaging; comparative studies; lung; MR; radiography; images; analysis;
D O I
10.1007/s00330-003-2215-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to evaluate low-field MR imaging of the lung parenchyma in comparison with postero-anterior (PA) and lateral chest radiographs (CR). One hundred one prospectively randomized patients who had received routine CR were additionally examined with magnetic resonance imaging (MRI) at 0.2 T. Utilized sequences were: constructive interference in steady state (CISS), true fast imaging in steady state precession (True-FISP) and T1-weighted spin-echo (T1SE). Consensus reading of two observers was performed for CR. Three other observers analyzed hardcopies of the MRI examinations for each sequence independently. The individual results for the comparisons between the sequences and CR were calculated using kappa coefficients with their corresponding confidence intervals. Additionally, an interobserver analysis was performed.The proportions of agreement for the three sequences compared with CR were high, with 0.93 for CISS, 0.89 for True-FISP and 0.91 for T1SE. The kappa coefficients and the corresponding confidence intervals were 0.81 [0.68; 0.95] for CISS, 0.72 [0.57; 0.88] for True-FISP and 0.78 [0.65; 0.92] for T1SE. Concerning CISS, differences between MRI and CR were mainly related to advantages resulting from cross-sectional imaging. The smallest 95% lower confidence bound of the three kappa measures for comparing the MR readers with each other was 0.97, indicating a high interobserver agreement. Low-field MRI of the lung parenchyma using the CISS sequence is well comparable with chest radiography and demonstrates slight advantages resulting from the cross-sectional imaging technique.
引用
收藏
页码:703 / 708
页数:6
相关论文
共 22 条
  • [1] Gadomer-17-enhanced 3D navigator-echo MR angiography of the pulmonary arteries in pigs
    Abolmaali, ND
    Hietschold, V
    Appold, S
    Ebert, W
    Vogl, TJ
    [J]. EUROPEAN RADIOLOGY, 2002, 12 (03) : 692 - 697
  • [2] Abolmaali ND, 2001, RADIOLOGY, V221, P630
  • [3] MULTISLICE, BREATHHOLD IMAGING OF THE LUNG WITH SUBMILLISECOND ECHO TIMES
    ALSOP, DC
    HATABU, H
    BONNET, M
    LISTERUD, J
    GEFTER, W
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) : 678 - 682
  • [4] LUNG PARENCHYMA - PROJECTION RECONSTRUCTION MR-IMAGING
    BERGIN, CJ
    PAULY, JM
    MACOVSKI, A
    [J]. RADIOLOGY, 1991, 179 (03) : 777 - 781
  • [5] Lung morphology: Fast MR imaging assessment with a volumetric interpolated breath-hold technique: Initial experience with patients
    Biederer, J
    Both, M
    Graessner, J
    Liess, C
    Jakob, P
    Reuter, M
    Heller, M
    [J]. RADIOLOGY, 2003, 226 (01) : 242 - 249
  • [6] COHEN AM, 1984, RADIOL CLIN N AM, V22, P829
  • [7] Lung air spaces:: MR imaging evaluation with hyperpolarized 3He gas
    de Lange, EE
    Mugler, JP
    Brookeman, JR
    Knight-Scott, J
    Truwit, JD
    Teates, CD
    Daniel, TM
    Bogorad, PL
    Cates, GD
    [J]. RADIOLOGY, 1999, 210 (03) : 851 - 857
  • [8] DEIMLING M, 2000, P INT SOC MAG RESON, V8, P2202
  • [9] HAUSCHKE D, 2000, STAT SIGNIFICANCE, P493
  • [10] Prospective feasibility study of chest X-ray vs. thoracic MRI in breath-hold technique at an open low-field scanner
    Heussel, CP
    Sandner, A
    Voigtländer, T
    Heike, M
    Deimling, M
    Kuth, R
    Rupprecht, T
    Schreiber, WG
    Kauczor, HU
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (07): : 854 - 861