High-efficiency SPECT MPI: Comparison of automated quantification, visual interpretation, and coronary angiography

被引:49
作者
Duvall, W. Lane [1 ]
Slomka, Piotr J. [2 ]
Gerlach, Jim R. [2 ]
Sweeny, Joseph M. [1 ]
Baber, Usman [1 ]
Croft, Lori B. [1 ]
Guma, Krista A. [1 ]
George, Titus [1 ]
Henzlova, Milena J. [1 ]
机构
[1] Mt Sinai Med Ctr, Mt Sinai Div Cardiol Mt Sinai Heart, New York, NY 10029 USA
[2] Cedars Sinai Med Ctr, Artificial Intelligence Med Program, Los Angeles, CA 90048 USA
基金
美国国家卫生研究院;
关键词
CZT SPECT MPI; high-efficiency SPECT MPI; automated quantification; coronary angiography; MYOCARDIAL-PERFUSION SPECT; SIMPLIFIED NORMAL LIMITS; ARTERY-DISEASE; HIGH-SPEED; CLINICAL VALIDATION; CAMERA; SUPINE;
D O I
10.1007/s12350-013-9735-x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Recently introduced high-efficiency (HE) SPECT cameras with solid-state CZT detectors have been shown to decrease imaging time and reduce radiation exposure to patients. An automated, computer-derived quantification of HE MPI has been shown to correlate well with coronary angiography on one HE SPECT camera system (D-SPECT), but has not been compared to visual interpretation on any of the HE SPECT platforms. Patients undergoing a clinically indicated Tc-99m sestamibi HE SPECT (GE Discovery 530c with supine and prone imaging) study over a 1-year period followed by a coronary angiogram within 2 months were included. Only patients with a history of CABG surgery were excluded. Both MPI studies and coronary angiograms were reinterpreted by blinded readers. One hundred and twenty two very low (risk of CAD < 5%) or low (risk of CAD < 10%) likelihood subjects with normal myocardial perfusion were used to create normal reference limits. Computer-derived quantification of the total perfusion deficit at stress and rest was obtained with QPS software. The visual and automated MPI quantification were compared to coronary angiography (a parts per thousand yen70% luminal stenosis) by receiver operating curve (ROC) analysis. Of the 3,111 patients who underwent HE SPECT over a 1-year period, 160 patients qualified for the correlation study (66% male, 52% with a history of CAD). The ROC area under the curve (AUC) was similar for both the automated and the visual interpretations using both supine only and combined supine and prone images (0.69-0.74). Using thresholds determined from sensitivity and specificity curves, the automated reads showed higher specificity (59%-67% vs 27%-60%) and lower sensitivity (71%-72% vs 79%-93%) than the visual reads. By including prone images sensitivity decreased slightly but specificity increased for both. By excluding patients with known CAD and cardiomyopathies, AUC and specificity increased for both techniques (0.72-0.82). The use of a difference score to evaluate ischemic burden resulted in lower sensitivities but higher specificities for both automated and visual quantification. There was good agreement between the visual interpretation and automated quantification in the entire cohort of 160 unselected consecutive patients (r = 0.70-0.81, P < .0001). Automated and visual quantification of high-efficiency SPECT MPI with the GE Discovery camera provides similar overall diagnostic accuracy when compared to coronary angiography. There was good correlation between the two methods of assessment. Combined supine and prone stress imaging provided the best diagnostic accuracy.
引用
收藏
页码:763 / 773
页数:11
相关论文
共 29 条
[1]   Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation [J].
Buechel, Ronny R. ;
Herzog, Bernhard A. ;
Husmann, Lars ;
Burger, Irene A. ;
Pazhenkottil, Aju P. ;
Treyer, Valerie ;
Valenta, Ines ;
von Schulthess, Patrick ;
Nkoulou, Rene ;
Wyss, Christophe A. ;
Kaufmann, Philipp A. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (04) :773-778
[2]   Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association [J].
Cerqueira, MD ;
Weissman, NJ ;
Dilsizian, V ;
Jacobs, AK ;
Kaul, S ;
Laskey, WK ;
Pennell, DJ ;
Rumberger, JA ;
Ryan, T ;
Verani, MS .
JOURNAL OF NUCLEAR CARDIOLOGY, 2002, 9 (02) :240-245
[3]   Reduced stress dose with rapid acquisition CZT SPECT MPI in a non-obese clinical population: Comparison to coronary angiography [J].
Duvall, W. Lane ;
Sweeny, Joseph M. ;
Croft, Lori B. ;
Ginsberg, Eric ;
Guma, Krista A. ;
Henzlova, Milena J. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2012, 19 (01) :19-27
[4]   Reduced isotope dose and imaging time with a high-efficiency CZT SPECT camera [J].
Duvall, W. Lane ;
Croft, Lori B. ;
Ginsberg, Eric S. ;
Einstein, Andrew J. ;
Guma, Krista A. ;
George, Titus ;
Henzlova, Milena J. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2011, 18 (05) :847-857
[5]   Comparison of high efficiency CZT SPECT MPI to coronary angiography [J].
Duvall, W. Lane ;
Sweeny, Joseph M. ;
Croft, Lori B. ;
Barghash, Maya H. ;
Kulkarni, Nitin K. ;
Guma, Krista A. ;
Henzlova, Milena J. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2011, 18 (04) :595-604
[6]   Reduced isotope dose with rapid SPECT MPI imaging: Initial experience with a CZT SPECT camera [J].
Duvall, W. Lane ;
Croft, Lori B. ;
Godiwala, Tapan ;
Ginsberg, Eric ;
George, Titus ;
Henzlova, Milena J. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2010, 17 (06) :1009-1014
[7]   Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras [J].
Esteves, Fabio P. ;
Raggi, Paolo ;
Folks, Russell D. ;
Keidar, Zohar ;
Askew, J. Wells ;
Rispler, Shmuel ;
O'Connor, Michael K. ;
Verdes, Liudmilla ;
Garcia, Ernest V. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2009, 16 (06) :927-934
[8]   Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: first validation versus invasive coronary angiography [J].
Fiechter, Michael ;
Ghadri, Jelena R. ;
Kuest, Silke M. ;
Pazhenkottil, Aju P. ;
Wolfrum, Mathias ;
Nkoulou, Rene N. ;
Goetti, Robert ;
Gaemperli, Oliver ;
Kaufmann, Philipp A. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 (11) :2025-2030
[9]   Cardiac Dedicated Ultrafast SPECT Cameras: New Designs and Clinical Implications [J].
Garcia, Ernest V. ;
Faber, Tracy L. ;
Esteves, Fabio P. .
JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (02) :210-217
[10]   High diagnostic accuracy of low-dose gated-SPECT with solid-state ultrafast detectors: preliminary clinical results [J].
Gimelli, Alessia ;
Bottai, Matteo ;
Genovesi, Dario ;
Giorgetti, Assuero ;
Di Martino, Fabio ;
Marzullo, Paolo .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 (01) :83-90