Avalon© Bicaval Dual-Lumen Cannula for Venovenous Extracorporeal Membrane Oxygenation: Survey of Cannula Use in France

被引:62
作者
Chimot, Loic [1 ,2 ]
Marque, Sophie [1 ]
Gros, Antoine [1 ]
Gacouin, Arnaud [1 ]
Lavoue, Sylvain [1 ]
Camus, Christophe [1 ]
Le Tulzo, Yves [1 ]
机构
[1] Univ Rennes 1, Hop Pontchaillou, Rennes, France
[2] Ctr Hosp Perigueux, Intens Care Unit, F-29019 Perigueux, France
关键词
ECMO; epidemiology; ARDS; echocardiography; procedure; database; RESPIRATORY-DISTRESS-SYNDROME; ACUTE LUNG INJURY; INTENSIVE-CARE UNITS; FAILURE; EPIDEMIOLOGY; MORTALITY; INFLUENZA;
D O I
10.1097/MAT.0b013e31827db6f3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We conducted an epidemiologic survey in France on the use of bicaval dual-lumen cannulas for extracorporeal membrane oxygenation (ECMO). Every service that used the Avalon (c) cannula was contacted. Practitioners answered questions concerning its practical usage and complications that were attributable to its usage. We report data for 52 instances of cannula usage. The primary indication was acute respiratory distress syndrome (ARDS) in 77% of cases. Of all of the patients who required cannulas, 46% died. The maximum flow was 2,175 +/- 556 ml/minute for 20-Fr.-diameter cannulas, 3,207 +/- 653 ml/minute for 23 Fr., 3,963 +/- 729 ml/minute for 27 Fr., and 5,490 +/- 984 ml/minute for 31 Fr. Surgeons placed the cannulas in 52% of cases, intensivists placed the cannulas in 23% of cases, and multidisciplinary teams placed the cannulas in 25% of cases. The mean insertion time was 26 +/- 13 minutes, and insertion was performed under transesophageal electrocardiography (TEE) (67%), transthoracic echocardiography (TTE) (25%), fluoroscopy (4%), or no guidance (4%). The main complication was migration into the right ventricle. Problems with hemolysis were described in 21% of cases. No case of cannula thrombosis was found. No case of infection was reported. Bleeding was noted in 17% of cases. The mean time of use was 8 +/- 7 days. Modifications to the supportive care system were required in 15% of cases. Monitoring was performed by chest x-rays (90%), TTE (42%), and TEE (46%). Five extubations occurred during the support period. Nine patients were mobilized. The use of this cannula yielded satisfactory results. We suggest placing these cannulas using TTE or TEE and recommend the use of large-caliber cannulas in hypoxemic patients. ASAIO Journal 2013;59:157-161.
引用
收藏
页码:157 / 161
页数:5
相关论文
共 16 条
[1]   Initial Experience With Single Cannulation for Venovenous Extracorporeal Oxygenation in Adults [J].
Bermudez, Christian A. ;
Rocha, Rodolfo V. ;
Sappington, Penny L. ;
Toyoda, Yoshiya ;
Murray, Holt N. ;
Boujoukos, Arthur J. .
ANNALS OF THORACIC SURGERY, 2010, 90 (03) :991-995
[2]   Incidence and mortality of acute lung injury and the acute respiratory distress syndrome in three Australian states [J].
Bersten, AD ;
Edibam, C ;
Hunt, T ;
Moran, J .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2002, 165 (04) :443-448
[3]   Epidemiology and outcome of acute lung injury in European intensive care units - Results from the ALIVE study [J].
Brun-Buisson, C ;
Minelli, C ;
Bertolini, G ;
Brazzi, L ;
Pimentel, J ;
Lewandowski, K ;
Bion, J ;
Rornand, JA ;
Villar, J ;
Thorsteinsson, A ;
Damas, P ;
Armaganidis, A ;
Lemaire, FO .
INTENSIVE CARE MEDICINE, 2004, 30 (01) :51-61
[4]  
Cornejo R, 2011, MINERVA ANESTESIOL, V77, P510
[5]   Extracorporeal Membrane Oxygenation for 2009 Influenza A(H1N1) Acute Respiratory Distress Syndrome [J].
Davies, Andrew ;
Jones, Daryl ;
Bailey, Michael ;
Beca, John ;
Bellomo, Rinaldo ;
Blackwell, Nikki ;
Forrest, Paul ;
Gattas, David ;
Granger, Emily ;
Herkes, Robert ;
Jackson, Andrew ;
McGuinness, Shay ;
Nair, Priya ;
Pellegrino, Vincent ;
Pettilae, Ville ;
Plunkett, Brian ;
Pye, Roger ;
Torzillo, Paul ;
Webb, Steve ;
Wilson, Michael ;
Ziegenfuss, Marc .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2009, 302 (17) :1888-1895
[6]   Ambulatory extracorporeal membrane oxygenation: A new approach for bridge-to-lung transplantation [J].
Garcia, Jose P. ;
Iacono, Aldo ;
Kon, Zachary N. ;
Griffith, Bartley P. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 139 (06) :E137-E139
[7]   Functional Disability 5 Years after Acute Respiratory Distress Syndrome [J].
Herridge, Margaret S. ;
Tansey, Catherine M. ;
Matte, Andrea ;
Tomlinson, George ;
Diaz-Granados, Natalia ;
Cooper, Andrew ;
Guest, Cameron B. ;
Mazer, C. David ;
Mehta, Sangeeta ;
Stewart, Thomas E. ;
Kudlow, Paul ;
Cook, Deborah ;
Slutsky, Arthur S. ;
Cheung, Angela M. .
NEW ENGLAND JOURNAL OF MEDICINE, 2011, 364 (14) :1293-1304
[8]   Use of Bicaval Dual-Lumen Catheter for Adult Venovenous Extracorporeal Membrane Oxygenation [J].
Javidfar, Jeffrey ;
Brodie, Daniel ;
Wang, Dongfang ;
Ibrahimiye, Ali N. ;
Yang, Jonathan ;
Zwischenberger, Joseph B. ;
Sonett, Joshua ;
Bacchetta, Matthew .
ANNALS OF THORACIC SURGERY, 2011, 91 (06) :1763-1769
[9]   Review of ECMO (Extra Corporeal Membrane Oxygenation) Support in Critically Ill Adult Patients [J].
Marasco, Silvana F. ;
Lukas, George ;
McDonald, Michael ;
McMillan, James ;
Ihle, Benno .
HEART LUNG AND CIRCULATION, 2008, 17 :S41-S47
[10]   Efficacy and economic assessment of conventional ventilatory support versus extracorporeal membrane oxygenation for severe adult respiratory failure (CESAR): a multicentre randomised controlled trial [J].
Peek, Giles J. ;
Mugford, Miranda ;
Tiruvoipati, Ravindranath ;
Wilson, Andrew ;
Allen, Elizabeth ;
Thalanany, Mariamma M. ;
Hibbert, Clare L. ;
Truesdale, Ann ;
Clemens, Felicity ;
Cooper, Nicola ;
Firmin, Richard K. ;
Elbourne, Diana .