Small field measurements with a novel silicon position sensitive diode array

被引:13
|
作者
Manolopoulos, S. [1 ,2 ]
Wojnecki, C. [2 ]
Hugtenburg, R. [4 ]
Sidek, M. A. Jaafar [3 ]
Chalmers, G. [2 ]
Heyes, G. [2 ]
Green, S. [2 ]
机构
[1] Sci & Technol Facil Council, Rutherford Appleton Lab, Chilton OX11 0QX, England
[2] Univ Hosp Birmingham NHS Trust, Dept Med Phys, Birmingham B15 2TH, W Midlands, England
[3] Univ Birmingham, Dept Phys & Astron, Birmingham B15 2TT, W Midlands, England
[4] Swansea Univ, Sch Med, Swansea SA2 8PP, W Glam, Wales
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2009年 / 54卷 / 03期
基金
英国科学技术设施理事会;
关键词
MICROSTRIP DOSIMETER; DELTA-OSI; PROTOTYPE;
D O I
10.1088/0031-9155/54/3/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
DOSI, a novel dosimeter based on position sensitive detectors for particle physics experiments, was used for relative clinical dosimetry measurements in small radiotherapy fields. The device is capable of dynamic measurements in real time and provides sub-millimetre spatial resolution. The basic beam data for a stereotactic radiotherapy collimator system (BrainLAB) using 6 MV photons were measured and compared with the corresponding data acquired with a small diamond detector and a PinPoint ionization chamber. All measurements showed an excellent agreement between DOSI and the diamond detector. There was an increasing discrepancy between the relative output factors (ROF) measured with DOSI and those measured with the ionization chamber with decreasing field size, specifically for collimators with a diameter smaller than 15 mm. The percentage depth doses (PDD) were in agreement to better than 1% for all depths. The agreement on off-axis ratios (OAR) was better than 3% for all collimators, whereas the agreement on relative output factors (ROF) was at the 1% level. DOSI's fast read-out electronics made it possible for all measurements to be recorded within 45 min including time to change collimators. This should reduce the overall time for commissioning and QA measurements, an important factor especially for busy radiotherapy departments.
引用
收藏
页码:485 / 495
页数:11
相关论文
共 50 条
  • [21] AN EXAFS STUDY OF AQUEOUS CYANOCUPRATE(I) SOLUTIONS WITH A POSITION-SENSITIVE PHOTO-DIODE ARRAY
    SANO, M
    MARUO, T
    YAMATERA, H
    CHEMICAL PHYSICS LETTERS, 1983, 101 (02) : 211 - 214
  • [22] Novel Amorphous Silicon Detector Array for Linear Accelerator Electron Measurements
    Shaikh, M.
    Burmeister, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E248 - E248
  • [23] Small Field Dosimetry Comparison of a Sub-Mm Plastic Scintillator to Silicon Diode Detectors Measurements and Monte Carlo Calculations
    Northern, E.
    BaradaranGhahfarokhi, M.
    Cooney, A.
    Ding, G. X.
    Luo, G.
    Homann, K. L.
    MEDICAL PHYSICS, 2024, 51 (09) : 6623 - 6623
  • [24] Multicentre comparison for small field dosimetry using the new silicon diode RAZOR
    Talamonti, C.
    Falco, M. D.
    Tonghi, L. Barone
    Benecchi, G.
    Carbonini, C.
    Casale, M.
    Clemente, S.
    Consorti, R.
    Di Castro, E.
    Esposito, M.
    Fiandra, C.
    Gasperi, C.
    Iervolino, C.
    Luxardo, S.
    Marino, C.
    Mones, E.
    Oliviero, C.
    Pressello, M. C.
    Riccardi, S.
    Rosica, F.
    Spiazzi, L.
    Stasi, M.
    Strigari, L.
    Mancosu, P.
    Russo, S.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S924 - S925
  • [25] DIODE-ARRAY ON SILICON FOR CHEMICAL SENSING
    HUGHES, RC
    ZIPPERIAN, TE
    SCHUBERT, WK
    PLUT, TA
    RODRIGUEZ, JL
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1986, 133 (08) : C331 - C331
  • [26] Position-sensitive single-photon detection with a hybrid silicon pixel array tube
    CERN, Geneva, Switzerland
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997, 387 (1-2): : 131 - 133
  • [27] Amorphous Silicon Position Sensitive Detector Array for Fast 3-D Object Profiling
    Contreras, Javier
    Idzikowski, Marek
    Pereira, Sonia
    Filonovich, Sergej A.
    Fortunato, Elvira
    Martins, Rodrigo
    Ferreira, Isabel
    IEEE SENSORS JOURNAL, 2012, 12 (04) : 812 - 820
  • [28] A CAMERA TUBE WITH SILICON DIODE ARRAY TARGET
    CROWELL, MH
    BUCK, TM
    LABUDA, EF
    DALTON, JV
    WALSH, EJ
    BELL SYSTEM TECHNICAL JOURNAL, 1967, 46 (02): : 491 - +
  • [29] Position-sensitive single-photon detection with a hybrid silicon pixel array tube
    Gys, T
    DAmbrosio, C
    Heijne, E
    Leutz, H
    Piedigrossi, D
    Puertolas, D
    Rosso, E
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 387 (1-2): : 131 - 133
  • [30] Energy Dependence of the Response of a Silicon Diode Designed for Small-Field Dosimetry
    Aeffner, S.
    Yarahmadi, M.
    Sauer, O.
    MEDICAL PHYSICS, 2013, 40 (06)