Development of a novel cardiopulmonary resuscitation measurement tool using real-time feedback from wearable wireless instrumentation

被引:7
作者
Ward, Sarah R. [1 ]
Scott, Bronwyn C. [1 ]
Rubin, David M. [1 ]
Pantanowitz, Adam [1 ]
机构
[1] Univ Witwatersrand, Sch Elect & Informat Engn, Biomed Engn Res Grp, Private Bag 3, ZA-2050 Johannesburg, South Africa
关键词
Cardiopulmonary resuscitation (CPR); Quality; Dual-quaternions; Electromyogram (EMG); Inertial Measurement Unit (IMU); MYO; Machine learning; IMPROVES CPR QUALITY; AUDIOVISUAL FEEDBACK; PERFORMANCE; COMPRESSION;
D O I
10.1016/j.resuscitation.2019.02.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Aim: The design and implementation of a wearable training device to improve cardiopulmonary resuscitation (CPR) is presented. Methods: The MYO contains both Electromyography (EMG) and Inertial Measurement Unit (IMU) sensors which are used to detect effective CPR, and the four common incorrect hand and arm positions viz. relaxed fingers; hands too low on the sternum; patient too close; or patient too far. The device determines the rate and depth of compressions calculated using a Fourier transform and dual-quaternions respectively. In addition, common positional mistakes are determined using classification algorithms (six machine learning algorithms are considered and tested). Feedback via Graphical User Interface (GUI) and audio is integrated. Results: The system is tested by performing CPR on a mannequin and comparing real-time results to theoretical values. Tests show that although the classification algorithm performed well in testing (98%), in real time, it had low accuracy for certain categories (60%) which are attributable to the MYO calibration, sampling rate and misclassification of similar hand positions. Combining these similar incorrect positions into more general categories significantly improves accuracy, and produces the same improved outcome of improved CPR. The rate and depth measures have a general accuracy of 97%. Conclusion: The system allows for portable, real-time feedback for use in training and in the field, and shows promise toward classifying and improving the administration of CPR.
引用
收藏
页码:183 / 189
页数:7
相关论文
共 32 条
  • [1] Compression depth estimation for CPR quality assessment using DSP on accelerometer signals
    Aase, SO
    Myklebust, H
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (03) : 263 - 268
  • [2] Abe S, 2010, ADV PATTERN RECOGNIT, P1, DOI 10.1007/978-1-84996-098-4
  • [3] CPR quality improvement during in-hospital cardiac arrest using a real-time audiovisual feedback system
    Abella, Benjamin S.
    Edelson, Dana P.
    Kim, Salem
    Retzer, Elizabeth
    Myklebust, Helge
    Barry, Anne M.
    O'Hearn, Nicholas
    Hoek, Terry L. Vanden
    Becker, Lance B.
    [J]. RESUSCITATION, 2007, 73 (01) : 54 - 61
  • [4] American Heart Association Circulation, 2010, CIRCULATION, P5640
  • [5] Anderson R, 2018, RESUSCITATION
  • [6] [Anonymous], COS511 THEORETICAL M
  • [7] Real-time visual feedback during training improves laypersons' CPR quality: a randomized controlled manikin study
    Baldi, Enrico
    Cornara, Stefano
    Contri, Enrico
    Epis, Francesco
    Fina, Dario
    Zelaschi, Beatrice
    Dossena, Cinzia
    Fichtner, Ferdinando
    Tonani, Michela
    DiMaggio, Marzia
    Zambaiti, Elisa
    Somaschini, Alberto
    [J]. CANADIAN JOURNAL OF EMERGENCY MEDICINE, 2017, 19 (06) : 480 - 487
  • [8] Viewing an ultra-brief chest compression only video improves some measures of bystander CPR performance and responsiveness at a mass gathering event
    Beskind, Daniel L.
    Stolz, Uwe
    Thiede, Rebecca
    Hoyer, Riley
    Robertson, Whitney
    Brown, Jeffrey
    Ludgate, Melissa
    Tiutan, Timothy
    Shane, Romy
    McMorrow, Deven
    Pleasants, Michael
    Kern, Karl B.
    Panchal, Ashish R.
    [J]. RESUSCITATION, 2017, 118 : 96 - 100
  • [9] PROLONGED PAUSES IN CARDIOPULMONARY RESUSCITATION ARE ASSOCIATED WITH POOR SURVIVAL DURING IN-HOSPITAL CARDIAC ARREST
    Birchak, John
    Abdul-Kafi, Owais
    Pham, Trinh
    Viner, Maya
    Nehmer, Maria
    Rao, Birju
    Ardati, Amer
    Fraidenburg, Dustin
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 71 (11) : 459 - 459
  • [10] EVALUATION OF CARDIOPULMONARY RESUSCITATION (CPR) TECHNIQUES
    BOSSAERT, L
    VANHOEYWEGHEN, R
    [J]. RESUSCITATION, 1989, 17 : S99 - S109