Enzymatic Biosensors with Integrated Thin Film a-Si:H Photodiodes

被引:2
作者
Pereira, A. T. [1 ,2 ]
Chu, V. [1 ]
Prazeres, D. M. F. [2 ,3 ]
Conde, J. P. [1 ,3 ]
机构
[1] INESC Microsistemas & Nanotecnol, Rua Alves Redol 9, Lisbon, Portugal
[2] Inst Super Tecn, Inst Biotechnol & Bioengn, Ctr Engn Biol Quim, Lisbon, Portugal
[3] Inst Super Tecn, Dept Chem & Biol Engn, Lisbon, Portugal
来源
AMORPHOUS AND POLYCRYSTALLINE THIN FILM SILICON SCIENCE AND TECHNOLOGY - 2009, VOL 1153 | 2009年 / 1153卷
关键词
AMORPHOUS-SILICON PHOTODIODE;
D O I
10.1557/PROC-1153-A08-04
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A microfabricated amorphous silicon photodiode is used to detect chemiluminescent and colorimetric horseradish peroxidase (HRP) enzymatic reactions. Detections limits of 1 nM and 1 pM of HRP are obtained for chemiluminescent and colorimetric measurements, respectively, with the reactions carried out in solution volume of 50 mu L in polystyrene microwells. Surface-adsorbed HRP can be detected with a limit of 1 fmol.cm(-2) by both detection methods. Immunoassays were performed using HRP-labeled antibodies and the detection of specific antibody-antigen molecular recognition is demonstrated both in the plastic well and inside a microfluidic channel. The application of the a-Si:H/HRP system is extended by coupling HRP with oxidase enzyme systems for glucose detection and a sensitivity of 0.1 mmol/L was achieved.
引用
收藏
页数:6
相关论文
共 5 条
  • [1] Azevedo AM, 2003, BIOTECHNOL ANN REV, V9, P199, DOI 10.1016/S1387-2656(03)09003-3
  • [2] Colorimetric detection of molecular recognition reactions with an enzyme biolabel using a thin-film amorphous silicon photodiode on a glass substrate
    Gouvea, A.
    Pereira, A. T.
    Pimentel, A. C.
    Prazeres, D. M. F.
    Chu, V.
    Conde, J. P.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2008, 135 (01) : 102 - 107
  • [3] Pereira A. T., IEEE SENSOR IN PRESS
  • [4] Detection of chemiluminescence using an amorphous silicon photodiode
    Pimentel, Alexandra C.
    Pereira, A. T.
    Chu, V.
    Prazeres, D. M. F.
    Conde, J. P.
    [J]. IEEE SENSORS JOURNAL, 2007, 7 (3-4) : 415 - 416
  • [5] Street R. A., 2005, HYDROGENATED AMORPHO