Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids

被引:79
作者
Fleming, Cherina [1 ]
Brouwer, Rob [1 ]
Lindemans, Jan [1 ]
de Jonge, Robert [1 ]
机构
[1] Univ Med Ctr Rotterdam, Dept Clin Chem, Erasmus MC, NL-3015 CE Rotterdam, Netherlands
关键词
body fluid; cerebrospinal fluid; counting chamber; hematology analyzer; leukocytes; Sysmex XN-1000; FLOW-CYTOMETRIC ANALYSIS; AUTOMATED HEMATOLOGY; PLEURAL FLUID; MODE; ANALYZERS; XE-5000;
D O I
10.1515/cclm-2011-0927
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: We evaluated the body fluid (BF) module on the new Sysmex XN-1000 for counting blood cells. Methods: One hundred and eighty-seven BF samples [73 cerebrospinal fluid (CSF), 48 continuous ambulatory peritoneal dialysis (CAPD), 46 ascites, and 20 pleural fluid] were used for method comparison between the XN-1000 and manual microscopy (Fuchs-Rosenthal chamber and stained cytospin slides) for counting red blood cells (RBCs) and white blood cells (WBCs) (differential). Results: Good agreement was found for counting WBCs (y=1.06x+0.09, n=67, R-2=0.96) and mononuclear cells (MNs) (y=1.04x-0.01, n=40, R-2=0.93) in CSF. However, the XN-1000 systematically counted more polymorphonuclear cells (PMNs) (y=1.48x+0.18, n=40, R-2=0.99) compared to manual microscopy. Excellent correlation for RBCs >1x109/L (y=0.99x+116.56, n=26, R-2=0.99) in CSF was found. For other fluids (CAPD, ascites and pleural fluid) excellent agreement was found for counting WBCs (y=1.06x+0.26, n=109, R-2=0.98), MNs (y=1.06x-0.41, n=93, R-2=0.96), PMNs (y=1.06x+0.81, n=93, R-2=0.98) and RBCs (y=1.04x+110.04, n=43, R-2=0.98). By using BF XN-check, the lower limit of quantitation (LLoQ) for WBC was defined at 5x10(6)/L. Linearity was excellent for both the WBCs (R-2=0.99) and RBCs (R-2=0.99) and carry-over never exceeded 0.05%. Conclusions: The BF module on the XN-1000 is a suitable tool for fast and accurate quantification of WBC (differential) and RBC counts in CSF and other BFs in a diagnostic setting.
引用
收藏
页码:1791 / 1798
页数:8
相关论文
共 35 条
  • [21] Evaluation of Mindray BC-6800 body fluid mode for automated cerebrospinal fluid cell counting
    Buoro, Sabrina
    Seghezzi, Michela
    Mecca, Tommaso
    Vavassori, Mauro
    Crippa, Alberto
    La Gioia, Antonio
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (11) : 1799 - 1810
  • [22] Preliminary evaluation of UF-5000 Body Fluid Mode for automated cerebrospinal fluid cell counting
    Seghezzi, Michela
    Manenti, Barbara
    Previtali, Giulia
    Alessio, Maria Grazia
    Dominoni, Paola
    Buoro, Sabrina
    CLINICA CHIMICA ACTA, 2017, 473 : 133 - 138
  • [23] A New Automated Technology for Cerebrospinal Fluid Cell Counts Comparison of Accuracy and Clinical Impact of GloCyte, Sysmex XN, and Manual Methods
    Sandhaus, Linda M.
    Dillman, Christine A.
    Hinkle, Warren P.
    MacKenzie, Julie M.
    Hong, George
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2017, 147 (05) : 507 - 514
  • [24] Use of the Cell-Dyn Sapphire Hematology Analyzer for Automated Counting of Blood Cells in Body Fluids
    De Smet, Dieter
    Van Moer, Guy
    Martens, Geert A.
    Nanos, Nikolaos
    Smet, Lutgarde
    Jochmans, Kristin
    De Waele, Marc
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2010, 133 (02) : 291 - 299
  • [25] Automated counting of white and red blood cells in the cerebrospinal fluid
    Aune, MW
    Sandberg, S
    CLINICAL AND LABORATORY HAEMATOLOGY, 2000, 22 (04): : 203 - 210
  • [26] Comparative Evaluation of Body Fluid Analysis by Sysmex XN Hematology Analyzers, CellaVision, Manual Microscopy and Multicolor Flow Cytometry
    Ivady, Gergely
    Barath, Sandor
    Szaraz-Szeles, Marianna
    Szabo, Eniko Kurti-G
    Kovacs, Kinga
    Petruska, Edina
    Hevessy, Zsuzsanna
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2022, 52 (02) : 314 - 322
  • [27] Extent of agreement between the body fluid model of Sysmex XN-20 and the manual microscopy method
    Huang, Wei-Hua
    Lu, Lin-Peng
    Wu, Kang
    Guo, Fang-Yu
    Guo, Jie
    Yu, Jing-Long
    Zhou, Dao-Yin
    Sun, Yi
    Deng, An-Mei
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2017, 31 (05)
  • [28] Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease
    Fleming, Cherina
    Russcher, Henk
    Lindemans, Jan
    de Jonge, Robert
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (11) : 1689 - 1706
  • [29] Detection of metastatic cells in body fluids by the automated Sysmex XN-9000 analyzer and flow-cytometric Infinicyt software
    Gerard, Delphine
    Henry, Sylvain
    Chatelain, Bernard
    Lesesve, Jean-Francois
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (04) : E146 - E149
  • [30] Quantification of the Cerebrospinal Fluid from a New Whole Body MRI Sequence
    Lebret, Alain
    Petit, Eric
    Durning, Bruno
    Hodel, Jerome
    Rahmouni, Alain
    Decq, Philippe
    MEDICAL IMAGING 2012: COMPUTER-AIDED DIAGNOSIS, 2012, 8315