Evaluation of an immobilised macrocyclic material for on-line column preconcentration and separation of cadmium, copper and lead for electrothermal atomic absorption spectrometry

被引:45
作者
Hosten, E [1 ]
Welz, B [1 ]
机构
[1] Bodenseewerk Perkin Elmer & Co GmbH, Dept Appl Res, D-88662 Uberlingen, Germany
关键词
flow injection on-line preconcentration; macrocyclic materials; electrothermal atomic absorption spectrometry;
D O I
10.1016/S0003-2670(99)00052-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The determination of trace amounts of Cd, Cu and Pb in riverine, estuarine, sea water and in urine samples, using a flow injection on-line preconcentration procedure coupled with electrothermal atomic absorption spectrometry, was investigated, The sorption material used consists of macrocyclic ligand immobilised on a silica gel support. The sample at pH 1.6 is loaded onto a 50 mu l column for a specific time interval. The column is then washed with a dilute nitric acid solution (pH 3) and emptied by pumping air through the column. The column is eluted with 40 mu l of 2.5 mol l(-1) nitric acid solution and the eluate introduced into the electrothermal atomiser. Detection limits (3 sigma) of 0.8, 46 and 20 ng l(-1) were obtained for Cd, Cu and Pb, respectively. The accuracy of the procedure was tested with water and urine certified reference materials. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:55 / 65
页数:11
相关论文
共 17 条
[1]   THE INFLUENCE OF TEMPERATURE AND PH ON TRACE-METAL SPECIATION IN SEAWATER [J].
BYRNE, RH ;
KUMP, LR ;
CANTRELL, KJ .
MARINE CHEMISTRY, 1988, 25 (02) :163-181
[2]   LITERATURE SURVEY OF THE ONLINE PRECONCENTRATION IN FLOW-INJECTION ATOMIC SPECTROMETRIC ANALYSIS [J].
CARBONELL, V ;
SALVADOR, A ;
DELAGUARDIA, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 342 (07) :529-537
[3]   Determination of heavy metals by inductively coupled plasma mass spectrometry after on-line separation and preconcentration [J].
Dressler, VL ;
Pozebon, D ;
Curtius, AJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (11) :1527-1539
[4]   ONLINE ANALYTE PRECONCENTRATION WITH ATOMIC SPECTROMETRIC DETECTION [J].
EBDON, L ;
FISHER, AS ;
HILL, SJ ;
WORSFOLD, PJ .
JOURNAL OF AUTOMATIC CHEMISTRY, 1991, 13 (06) :281-286
[5]  
Fang Z., 1993, FLOW INJECTION SEPAR
[6]   Trends and potentials in flow injection on-line separation and preconcentration techniques for electrothermal atomic absorption spectrometry [J].
Fang, ZL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (10) :1371-1379
[7]  
FANG ZL, 1991, SPECTROCHIM ACTA REV, V14, P235
[8]  
Fang ZL, 1996, FRESEN J ANAL CHEM, V355, P576
[9]   Electrothermal atomic absorption spectrometric determination of cadmium and lead in blood using flow injection on-line sorption preconcentration in a knotted reactor [J].
Ivanova, E ;
Van Mol, W ;
Adams, F .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (6-8) :1041-1048
[10]   Determination of trace metals in sea-water by inductively coupled plasma mass spectrometry interfaced with an ion chromatographic separation system: effectiveness of nitrilotriacetate chelating resin as the column stationary phase for preconcentration and elimination of matrix effects [J].
Kumagai, H ;
Yamanaka, M ;
Sakai, T ;
Yokoyama, T ;
Suzuki, TM ;
Suzuki, T .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (06) :579-582