Comparison of graphite calorimeter dosimetry system with Monte Carlo simulation at an industrial electron beam accelerator

被引:6
作者
Benny, P. G. [1 ]
Selvam, T. Palani [2 ]
Sarma, K. S. S. [1 ]
机构
[1] Bhabha Atom Res Ctr, Radiat Technol Dev Div, Bombay 400085, Maharashtra, India
[2] Bhabha Atom Res Ctr, Radiol Phys & Advisory Div, Bombay 400085, Maharashtra, India
关键词
Electron beam accelerator; Monte Carlo simulation; Graphite calorimeters; Dosimetry parameters;
D O I
10.1016/j.nima.2012.11.011
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In a previously published study, we reported the estimation of various dosimetry parameters such as average absorbed dose, energy fluence and practical range for an electron beam of 1.6 MeV using graphite calorimeters of different thicknesses. In this study, depth dose distribution in a graphite block was measured using cellulose triacetate (CTA) dosimeter strip for the scanned electron beam of kinetic energy 1.6 MeV and compared against the EGSnrc-based Monte Carlo calculations. Monte Carlo calculations agree to the measurements for depths beyond about 0.15 cm. Similarly absorbed dose measurements were carried out in graphite discs of different thicknesses and compared against the Monte Carlo calculations. The comparison is good for thicknesses of discs 0.4 cm and above. Although, the disagreement between measurements and calculations does not affect overall conclusion of the study, disagreement is attributed possibly due to presence of low energy spectrum of electrons in the measurements in addition to the prominent 1.6 MeV electrons. Both calculations and measurements demonstrate that energy absorption is linear up to 0.3 cm thick graphite. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:98 / 101
页数:4
相关论文
共 11 条
  • [1] [Anonymous], 51631 ISOASTM
  • [2] [Anonymous], 51649 ISOASTM
  • [3] [Anonymous], 35 ICRU
  • [4] Estimation of dosimetry parameters for an EB accelerator using graphite calorimeters of different thickness
    Benny, P. G.
    Khader, S. A.
    Sarma, K. S. S.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 680 : 108 - 111
  • [5] Derivation of electron and photon energy spectra from electron beam central axis depth dose curves
    Deng, J
    Jiang, SB
    Pawlicki, T
    Li, JS
    Ma, CM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05) : 1429 - 1449
  • [6] DEPTH-DOSE DISTRIBUTIONS IN A THIN-LAYER ABSORBER IRRADIATED BY 300-KEV ELECTRONS
    KIJIMA, T
    SUGA, H
    NAKASE, Y
    [J]. APPLIED RADIATION AND ISOTOPES, 1994, 45 (07) : 759 - 765
  • [7] McLaughlin William L, 1983, RADIAT PHYS CHEM, V21, P359
  • [8] Rogers D. W. O., 2006, PIRS701 NAT RES COUN
  • [9] ROGERS DWO, 2006, PIRS702 NAT RES COUN
  • [10] Sarma K.S.S., 2005, IANCAS B