A real-time in vivo dosimetric verification method for high-dose rate intracavitary brachytherapy of nasopharyngeal carcinoma

被引:24
|
作者
Qi, Zhen-Yu [1 ,2 ]
Deng, Xiao-Wu [1 ,2 ]
Cao, Xin-ping [1 ]
Huang, Shao-Min [1 ]
Lerch, Michael [2 ]
Rosenfeld, Anatoly [2 ]
机构
[1] Sun Yat Sen Univ, Ctr Canc, State Key Lab Oncol S China, Guangzhou 510060, Guangdong, Peoples R China
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
基金
中国国家自然科学基金;
关键词
MOSFET; in vivo dosimetry; real time dosimetry; intracaviatry brachytherapy; quality assurance; AVOID MISADMINISTRATIONS; SPATIAL-RESOLUTION; MOSFET DOSIMETRY; RADIOTHERAPY; CALIBRATION; TLD;
D O I
10.1118/1.4758067
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A real-time in vivo dosimetric verification method using metal-oxide-semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high-dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC). Methods: The necessary calibration and correction factors for MOSFET measurements in (192)Iridium source were determined in a water phantom. With the detector placed inside a custom-made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system. Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 +/- 0.007 cGy/mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1-5 cm from the source. A total of 70 in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were -0.1% +/- 3.8% and -0.1% +/- 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%. Conclusions: In vivo measurement using the real-time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4758067]
引用
收藏
页码:6757 / 6763
页数:7
相关论文
共 50 条
  • [21] The usefulness of an independent patient-specific treatment planning verification method using a benchmark plan in high-dose-rate intracavitary brachytherapy for carcinoma of the uterine cervix
    Takahashi, Yutaka
    Koizumi, Masahiko
    Sumida, Iori
    Isohashi, Fumiaki
    Ogata, Toshiyuki
    Akino, Yuichi
    Yoshioka, Yasuo
    Maruoka, Shintaro
    Inoue, Shinichi
    Konishi, Koji
    Ogawa, Kazuhiko
    JOURNAL OF RADIATION RESEARCH, 2012, 53 (06) : 936 - 944
  • [22] A phantom for verification of dwell position and time of a high dose rate brachytherapy source
    Madebo, M.
    Pillainayagam, J.
    Kron, T.
    Franich, R.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (03) : 335 - 339
  • [23] High-dose-rate remote afterloading intracavitary brachytherapy for the treatment of extrahepatic biliary duct carcinoma
    Lu, JJ
    Bains, YS
    Abdel-Wahab, M
    Brandon, AH
    Wolfson, AH
    Raub, WA
    Wilkinson, CM
    Markoe, AM
    CANCER JOURNAL, 2002, 8 (01) : 74 - 78
  • [24] Comparison of high-dose-rate intracavitary brachytherapy dosimetry with and without anesthesia in patients with cervical carcinoma
    Sharma, Daya N.
    Chaudhari, Pritee
    Sharma, Seema
    Gupta, Leena
    Jagadesan, Pandjatcharam
    Rath, Goura K.
    Julka, Pramod K.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 60 - 66
  • [25] Adjuvant high-dose-rate brachytherapy after external beam radiotherapy in nasopharyngeal carcinoma
    Özyar, E
    Yildz, F
    Akyol, FH
    Atahan, IL
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (01): : 101 - 108
  • [26] Treatment of facial cutaneous carcinoma with high-dose rate contact brachytherapy with customized molds
    Maronas, Maria
    Luis Guinot, Jose
    Arribas, Leoncio
    Carrascosa, Maria
    Isabel Tortajada, Maria
    Carmona, Ruth
    Estornell, Marian
    Muelas, Rodrigo
    BRACHYTHERAPY, 2011, 10 (03) : 221 - 227
  • [27] Impact of different dose prescription schedules on EQD2 in high-dose-rate intracavitary brachytherapy of carcinoma cervix
    Kumar, Mohan
    Thangaraj, Revathy
    Alva, Ram Charith
    Koushik, Kirthi
    Penni, Arul
    Achar, Janaki M. G.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2019, 11 (02) : 189 - 193
  • [28] A phantom study of an in vivo dosimetry system using plastic scintillation detectors for real-time verification of 192Ir HDR brachytherapy
    Therriault-Proulx, Francois
    Briere, Tina M.
    Mourtada, Firas
    Aubin, Sylviane
    Beddar, Sam
    Beaulieu, Luc
    MEDICAL PHYSICS, 2011, 38 (05) : 2542 - 2551
  • [29] Dosimetric impact of cylinder size in high-dose rate vaginal cuff brachytherapy (VCBT) for primary endometrial cancer
    Zhang, Hualin
    Gopalakrishnan, Mahesh
    Lee, Plato
    Kang, Zhuang
    Sathiaseelan, Vythialingam
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05): : 262 - 272
  • [30] Technical Note: Scintillation markers for real-time visual source tracking during skin high dose rate brachytherapy