A real-time in vivo dosimetric verification method for high-dose rate intracavitary brachytherapy of nasopharyngeal carcinoma

被引:24
|
作者
Qi, Zhen-Yu [1 ,2 ]
Deng, Xiao-Wu [1 ,2 ]
Cao, Xin-ping [1 ]
Huang, Shao-Min [1 ]
Lerch, Michael [2 ]
Rosenfeld, Anatoly [2 ]
机构
[1] Sun Yat Sen Univ, Ctr Canc, State Key Lab Oncol S China, Guangzhou 510060, Guangdong, Peoples R China
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
基金
中国国家自然科学基金;
关键词
MOSFET; in vivo dosimetry; real time dosimetry; intracaviatry brachytherapy; quality assurance; AVOID MISADMINISTRATIONS; SPATIAL-RESOLUTION; MOSFET DOSIMETRY; RADIOTHERAPY; CALIBRATION; TLD;
D O I
10.1118/1.4758067
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A real-time in vivo dosimetric verification method using metal-oxide-semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high-dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC). Methods: The necessary calibration and correction factors for MOSFET measurements in (192)Iridium source were determined in a water phantom. With the detector placed inside a custom-made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system. Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 +/- 0.007 cGy/mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1-5 cm from the source. A total of 70 in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were -0.1% +/- 3.8% and -0.1% +/- 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%. Conclusions: In vivo measurement using the real-time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4758067]
引用
收藏
页码:6757 / 6763
页数:7
相关论文
共 50 条
  • [1] In vivo real-time dosimetric verification in high dose rate prostate brachytherapy
    Seymour, Erin L.
    Downes, Simon J.
    Fogarty, Gerald B.
    Izard, Michael A.
    Metcalfe, Peter
    MEDICAL PHYSICS, 2011, 38 (08) : 4785 - 4794
  • [2] Real-time in vivo dosimetry in high dose rate prostate brachytherapy
    Mason, Josh
    Mamo, Arielle
    AI-Qaisieh, Bashar
    Henry, Ann M.
    Bownes, Peter
    RADIOTHERAPY AND ONCOLOGY, 2016, 120 (02) : 333 - 338
  • [3] Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera
    Duan, J
    Macey, DJ
    Pareek, PN
    Brezovich, IA
    MEDICAL PHYSICS, 2001, 28 (02) : 167 - 173
  • [4] In vivo dose verification method in catheter based high dose rate brachytherapy
    Jaselske, Evelina
    Adliene, Diana
    Rudzianskas, Viktoras
    Urbonavicius, Benas Gabrielis
    Inciura, Arturas
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 44 : 1 - 10
  • [5] Definitive radiation therapy for invasive carcinoma of the vagina: impact of high-dose rate intracavitary brachytherapy
    Nonaka, Tetsuo
    Nakayama, Yuko
    Mizoguchi, Nobutaka
    Onose, Ryo
    Kato, Hisamori
    Nakayama, Hiroki
    INTERNATIONAL JOURNAL OF CLINICAL ONCOLOGY, 2013, 18 (02) : 314 - 320
  • [6] Combined Interstitial and Intracavitary High-Dose Rate Brachytherapy of Cervical Cancer
    Itami, Jun
    Murakami, Naoya
    Watanabe, Miho
    Sekii, Shuhei
    Kasamatsu, Takahiro
    Kato, Shingo
    Hirowatari, Hisako
    Ikushima, Hitoshi
    Ando, Ken
    Ohno, Tatsuya
    Okamoto, Hiroyuki
    Okuma, Kae
    Igaki, Hiroshi
    FRONTIERS IN ONCOLOGY, 2022, 11
  • [7] Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
    Jamalludin, Zulaikha
    Jong, Wei Loong
    Malik, Rozita Abdul
    Rosenfeld, Anatoly
    Ung, Ngie Min
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 58 : 1 - 7
  • [8] Long-term results of high-dose rate intracavitary brachytherapy for squamous cell carcinoma of the uterine cervix
    Nakano, T
    Kato, S
    Ohno, T
    Tsujii, H
    Sato, S
    Fukuhisa, K
    Arai, T
    CANCER, 2005, 103 (01) : 92 - 101
  • [9] Treatment of Cervical Carcinoma with High-Dose Rate Intracavitary Brachytherapy: Two Years Follow-Up Study
    Das, Diptimay
    Chaudhuri, Snehamay
    Deb, Asit Ranjan
    Aich, Ranen Kanti
    Gangopadhyay, Subir
    Ray, Amitava
    ASIAN PACIFIC JOURNAL OF CANCER PREVENTION, 2011, 12 (03) : 807 - 810
  • [10] Evaluation of a real-time BeO ceramic fiber-coupled luminescence dosimetry system for dose verification of high dose rate brachytherapy
    Santos, Alexandre M. Caraca
    Mohammadi, Mohammad
    Afshar, Shahraam, V
    MEDICAL PHYSICS, 2015, 42 (11) : 6349 - 6356