A real-time in vivo dosimetric verification method for high-dose rate intracavitary brachytherapy of nasopharyngeal carcinoma

被引:24
作者
Qi, Zhen-Yu [1 ,2 ]
Deng, Xiao-Wu [1 ,2 ]
Cao, Xin-ping [1 ]
Huang, Shao-Min [1 ]
Lerch, Michael [2 ]
Rosenfeld, Anatoly [2 ]
机构
[1] Sun Yat Sen Univ, Ctr Canc, State Key Lab Oncol S China, Guangzhou 510060, Guangdong, Peoples R China
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
基金
中国国家自然科学基金;
关键词
MOSFET; in vivo dosimetry; real time dosimetry; intracaviatry brachytherapy; quality assurance; AVOID MISADMINISTRATIONS; SPATIAL-RESOLUTION; MOSFET DOSIMETRY; RADIOTHERAPY; CALIBRATION; TLD;
D O I
10.1118/1.4758067
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A real-time in vivo dosimetric verification method using metal-oxide-semiconductor field effect transistor (MOSFET) dosimeters has been developed for patient dosimetry in high-dose rate (HDR) intracavitary brachytherapy of nasopharyngeal carcinoma (NPC). Methods: The necessary calibration and correction factors for MOSFET measurements in (192)Iridium source were determined in a water phantom. With the detector placed inside a custom-made nasopharyngeal applicator, the actual dose delivered to the tumor was measured in vivo and compared to the calculated values using a commercial brachytherapy planning system. Results: Five MOSFETs were independently calibrated with the HDR source, yielding calibration factors of 0.48 +/- 0.007 cGy/mV. The maximum sensitivity variation was no more than 7% in the clinically relevant distance range of 1-5 cm from the source. A total of 70 in vivo measurements in 11 NPC patients demonstrated good agreement with the treatment planning. The mean differences between the planned and the actually delivered dose within a single treatment fraction were -0.1% +/- 3.8% and -0.1% +/- 3.7%, respectively, for right and left side assessments. The maximum dose deviation was less than 8.5%. Conclusions: In vivo measurement using the real-time MOSFET dosimetry system is possible to evaluate the actual dose to the tumor received by the patient during a treatment fraction and thus can offer another line of security to detect and prevent large errors. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4758067]
引用
收藏
页码:6757 / 6763
页数:7
相关论文
共 31 条
[1]   In-vivo rectal dose measurements with diodes to avoid misadministrations during intracavitary high dose rate brachytherapy for carcinoma of the cervix [J].
Alecu, R ;
Alecu, M .
MEDICAL PHYSICS, 1999, 26 (05) :768-770
[2]   A method to avoid misadministrations in high dose rate brachytherapy [J].
Alecu, R ;
Feldmeier, JJ ;
Court, WS ;
Alecu, M ;
Orton, CG .
MEDICAL PHYSICS, 1997, 24 (02) :259-261
[3]   In vivo thermoluminescence dosimetry dose verification of transperineal 192Ir high-dose-rate brachytherapy using CT-based planning for the treatment of prostate cancer [J].
Anagnostopoulos, G ;
Baltas, D ;
Geretschlaeger, A ;
Martin, T ;
Papagiannis, P ;
Tselis, N ;
Zamboglou, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04) :1183-1191
[4]   In vivo urethral dose measurements:: A method as verify high dose rate prostate treatments [J].
Brezovich, IA ;
Duan, J ;
Pareek, PN ;
Fiveash, J ;
Ezekiel, M .
MEDICAL PHYSICS, 2000, 27 (10) :2297-2301
[5]   Effects of temperature variation on MOSFET dosimetry [J].
Cheung, T ;
Butson, MJ ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13) :N191-N196
[6]   Low-dose measurement with a MOSFET in high-energy radiotherapy applications [J].
Cheung, T ;
Yu, PKN ;
Butson, MJ .
RADIATION MEASUREMENTS, 2005, 39 (01) :91-94
[7]   Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera [J].
Duan, J ;
Macey, DJ ;
Pareek, PN ;
Brezovich, IA .
MEDICAL PHYSICS, 2001, 28 (02) :167-173
[8]   Response of an implantable MOSFET dosimeter to 192Ir HDR radiation [J].
Fagerstrom, Jessica M. ;
Micka, John A. ;
DeWerd, Larry A. .
MEDICAL PHYSICS, 2008, 35 (12) :5729-5737
[9]   AUTOMATED DATA-COLLECTION AND ANALYSIS SYSTEM FOR MOSFET RADIATION DETECTORS [J].
GLADSTONE, DJ ;
CHIN, LM .
MEDICAL PHYSICS, 1991, 18 (03) :542-548
[10]   The use of TLD and Gafchromic film to assure submillimeter accuracy for image-guided radiosurgery [J].
Ho, Anthony K. ;
Gibbs, Iris C. ;
Chang, Steve D. ;
Main, Bill ;
Adler, John R. .
MEDICAL DOSIMETRY, 2008, 33 (01) :36-41