A fast 3D Look-Locker method for volumetric T1 mapping

被引:109
作者
Henderson, E [1 ]
McKinnon, G [1 ]
Lee, TY [1 ]
Rutt, BK [1 ]
机构
[1] John P Robarts Res Inst, Imaging Res Labs, London, ON N6A 5K8, Canada
关键词
longitudinal relaxation time; T-1; 3D; Look-Locker;
D O I
10.1016/S0730-725X(99)00025-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We introduce a fast technique, based on the principles of the 2D Look-Locker T-1 measurement scheme, to rapidly acquire the data for accurate maps of T-1 in three dimensions. The acquisition time has been shortened considerably by segmenting the acquisition of the k(y) phase encode lines. Using this technique, the data for a 256 x 128 x 32 volumetric T-1 measurement can be acquired in 7.6 min. T-1 measurements made in phantoms with T(1)s between 200 and 1200 ms had an accuracy of 4% and a reproducibility of 3.5%. Measurements of T-1 made in normal brain using the fast 3D sequence corresponded well with inversion-recovery fast spin-echo measurements. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:1163 / 1171
页数:9
相关论文
共 40 条
  • [1] SPIN-LATTICE RELAXATION-TIME MEASUREMENT BY MEANS OF A TURBOFLASH TECHNIQUE
    BLUML, S
    SCHAD, LR
    STEPANOW, B
    LORENZ, WJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (03) : 289 - 295
  • [2] FAST AND PRECISE T1 IMAGING USING A TOMROP SEQUENCE
    BRIX, G
    SCHAD, LR
    DEIMLING, M
    LORENZ, WJ
    [J]. MAGNETIC RESONANCE IMAGING, 1990, 8 (04) : 351 - 356
  • [3] Measurement of spin-lattice relaxation times with FLASH for dynamic MRI of the breast
    Brookes, JA
    Redpath, TW
    Gilbert, FJ
    Needham, G
    Murray, AD
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (819) : 206 - 214
  • [4] A COMPARISON OF ONE-SHOT AND RECOVERY METHODS IN T1-IMAGING
    CRAWLEY, AP
    HENKELMAN, RM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (01) : 23 - 34
  • [5] QUANTIFICATION OF T1 VALUES BY SNAPSHOT-FLASH NMR IMAGING
    DEICHMANN, R
    HAASE, A
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1992, 96 (03): : 608 - 612
  • [6] Pharmacokinetic analysis of Gd-DTPA enhancement in dynamic three-dimensional MRI of breast lesions
    denBoer, JA
    Hoenderop, RKKM
    Smink, J
    Dornseiffen, G
    Koch, PWAA
    Mulder, JH
    Slump, CH
    Volker, EDP
    deVos, RAI
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (04): : 702 - 715
  • [7] RAPID CALCULATION OF T1 USING VARIABLE FLIP ANGLE GRADIENT REFOCUSED IMAGING
    FRAM, EK
    HERFKENS, RJ
    JOHNSON, GA
    GLOVER, GH
    KARIS, JP
    SHIMAKAWA, A
    PERKINS, TG
    PELC, NJ
    [J]. MAGNETIC RESONANCE IMAGING, 1987, 5 (03) : 201 - 208
  • [8] T1 MAPPING FROM SPIN-ECHO AND STIMULATED ECHOES
    FRANCONI, F
    SEGUIN, F
    SONIER, CB
    LEPAPE, A
    AKOKA, S
    [J]. MEDICAL PHYSICS, 1995, 22 (11) : 1763 - 1769
  • [9] Optimization of the ultrafast Look-Locker echo-planar imaging T1 mapping sequence
    Freeman, AJ
    Gowland, PA
    Mansfield, P
    [J]. MAGNETIC RESONANCE IMAGING, 1998, 16 (07) : 765 - 772
  • [10] MEASUREMENT OF RADIATION-DOSE DISTRIBUTIONS BY NUCLEAR MAGNETIC-RESONANCE (NMR) IMAGING
    GORE, JC