Disposable Patterned Electroceutical Dressing (PED-10) Is Safe for Treatment of Open Clinical Chronic Wounds

被引:18
作者
Roy, Sashwati [1 ]
Prakash, Shaurya [2 ]
Mathew-Steiner, Shomita S. [1 ]
Ghatak, Piya Das [1 ]
Lochab, Varun [2 ]
Jones, Travis H. [2 ]
Sundaram, Prashanth Mohana [2 ]
Gordillo, Gayle M. [1 ]
Subramaniam, Vish V. [2 ]
Sen, Chandan K. [1 ]
机构
[1] Indiana Univ, Sch Med, Dept Surg, ICRME, 975 W Walnut St,Suite 450, Indianapolis, IN 46202 USA
[2] Ohio State Univ, Dept Mech & Aerosp Engn, Columbus, OH 43210 USA
关键词
electroceutical; wound care dressing; clinical; safety; chronic wound; ELECTRICAL-STIMULATION; PSEUDOMONAS-AERUGINOSA; STAPHYLOCOCCUS-AUREUS; BACTERIAL BIOFILMS; HIGH-VOLTAGE; SKIN WOUNDS; EFFICACY; ENHANCEMENT; FIELDS; INFLAMMATION;
D O I
10.1089/wound.2018.0915
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Objective: To evaluate if patterned electroceutical dressing (PED) is safe for human chronic wounds treatment as reported by wound care providers. Approach: This work reports a pilot feasibility study with the primary objective to determine physically observable effects of PED application on host tissue response from a safety evaluation point of view. For this pilot study, patients receiving a lower extremity amputation with at least one open wound on the part to be amputated were enrolled. Patients were identified through the Ohio State University Wexner Medical Center (OSUWMC) based on inclusion and exclusion criteria through prescreening through the Comprehensive Wound Center's (CWC) Limb Preservation Program and wound physicians and/or providers at OSUWMC. Wounds were treated with the PED before amputation surgery. Results: The intent of the study was to identify if PED was safe for clinical application based on visual observations of adverse or lack of adverse events on skin and wound tissue. The pilot testing performed on a small cohort (N = 8) of patients showed that with engineered voltage regulation of current flow to the open wound, the PED can be used with little to no visually observable adverse effects on chronic human skin wounds. Innovation: The PED was developed as a second-generation tunable electroceutical wound care dressing, which could potentially be used to treat wounds with deeper infections compared with current state of the art that treats wounds with treatment zone limited to the surface near topical application. Conclusion: Technology advances in design and fabrication of electroceutical dressings were leveraged to develop a tunable laboratory prototype that could be used as a disposable low-cost electroceutical wound care dressing on chronic wounds. Design revisions of PED-1 (1 k omega ballast resistor) circumvented previously observed adverse effects on the skin in the vicinity of an open wound. PED-10 (including a 10 k omega ballast resistor) was well tolerated in the small cohort of patients (N = 8) on whom it was tested, and the observations reported here warrant a larger study to determine the clinical impact on human wound healing and infection control.
引用
收藏
页码:149 / 159
页数:11
相关论文
共 75 条
[1]  
Abo-Neima S. E., 2016, INT J ENG SCI, V6, P13
[2]   THE HEALING OF SUPERFICIAL SKIN WOUNDS IS STIMULATED BY EXTERNAL ELECTRICAL-CURRENT [J].
ALVAREZ, OM ;
MERTZ, PM ;
SMERBECK, RV ;
EAGLSTEIN, WH .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1983, 81 (02) :144-148
[3]   Electrical stimulation disrupts biofilms in a human wound model and reveals the potential for monitoring treatment response with volatile biomarkers [J].
Ashrafi, Mohammed ;
Novak-Frazer, Lilyann ;
Morris, Julie ;
Baguneid, Mohamed ;
Rautemaa-Richardson, Riina ;
Bayat, Ardeshir .
WOUND REPAIR AND REGENERATION, 2019, 27 (01) :5-18
[4]   The efficacy of electrical stimulation in lower extremity cutaneous wound healing: A systematic review [J].
Ashrafi, Mohammed ;
Alonso-Rasgado, Teresa ;
Baguneid, Mohamed ;
Bayat, Ardeshir .
EXPERIMENTAL DERMATOLOGY, 2017, 26 (02) :171-178
[5]   The efficacy of electrical stimulation in experimentally induced cutaneous wounds in animals [J].
Ashrafi, Mohammed ;
Alonso-Rasgado, Teresa ;
Baguneid, Mohamed ;
Bayat, Ardeshir .
VETERINARY DERMATOLOGY, 2016, 27 (04) :235-E57
[6]   Silver-Zinc Redox-Coupled Electroceutical Wound Dressing Disrupts Bacterial Biofilm [J].
Banerjee, Jaideep ;
Das Ghatak, Piya ;
Roy, Sashwati ;
Khanna, Savita ;
Hemann, Craig ;
Deng, Binbin ;
Das, Amitava ;
Zweier, Jay L. ;
Wozniak, Daniel ;
Sen, Chandan K. .
PLOS ONE, 2015, 10 (03)
[7]   Improvement of Human Keratinocyte Migration by a Redox Active Bioelectric Dressing [J].
Banerjee, Jaideep ;
Das Ghatak, Piya ;
Roy, Sashwati ;
Khanna, Savita ;
Sequin, Emily K. ;
Bellman, Karen ;
Dickinson, Bryan C. ;
Suri, Prerna ;
Subramaniam, Vish V. ;
Chang, Christopher J. ;
Sen, Chandan K. .
PLOS ONE, 2014, 9 (03)
[8]   THE GLABROUS EPIDERMIS OF CAVIES CONTAINS A POWERFUL BATTERY [J].
BARKER, AT ;
JAFFE, LF ;
VANABLE, JW .
AMERICAN JOURNAL OF PHYSIOLOGY, 1982, 242 (03) :R358-R366
[9]   Electric Field Based Dressing Disrupts Mixed-Species Bacterial Biofilm Infection and Restores Functional Wound Healing [J].
Barki, Kasturi Ganesh ;
Das, Amitava ;
Dixith, Sriteja ;
Das Ghatak, Piya ;
Mathew-Steiner, Shomita ;
Schwab, Elizabeth ;
Khanna, Savita ;
Wozniak, Daniel J. ;
Roy, Sashwati ;
Sen, Chandan K. .
ANNALS OF SURGERY, 2019, 269 (04) :756-766
[10]   FURTHER STUDIES OF THE RATE OF HEALING OF HUMAN SKIN MEASURED BY THE ELECTRICAL WOUND POTENTIAL OF EXPERIMENTAL ABRASIONS [J].
BARNES, TC ;
KARASIC, J ;
AMOROSO, MD .
AMERICAN JOURNAL OF SURGERY, 1951, 82 (06) :720-726