Improved precision with Hologic Apex software

被引:21
作者
Fan, B. [1 ]
Lewiecki, E. M. [2 ]
Sherman, M. [1 ]
Lu, Y. [1 ]
Miller, P. D. [3 ]
Genant, H. K. [1 ,4 ]
Shepherd, J. A. [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94107 USA
[2] New Mexico Clin Res & Osteoporosis Ctr, Albuquerque, NM USA
[3] Colorado Ctr Bone Res, Lakewood, CO USA
[4] Synarc Inc, San Francisco, CA USA
关键词
bone mineral density (BMD); dual-energy X-ray absorptiometry (DXA); osteoporosis; software version;
D O I
10.1007/s00198-008-0594-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The precision of Hologic Apex v2.0 analysis software is significantly improved from Hologic Delphi v11.2 software and is comparable to GE Lunar Prodigy v7.5 software. Apex and Delphi precisions were, respectively, 1.0% vs. 1.2% (L1-L4 spine), 1.l % vs. 1.3% (total femur), 1.6% vs. 1.9% (femoral neck), and 0.7% vs. 0.9% (dual total femur). Introduction Precision of bone mineral density (BMD) measurements by dual-energy X-ray absorptiometry (DXA) is known to vary by manufacturer, model, and technologist. This study evaluated the precision of three analysis versions: Apex v2.0 and Delphi v11.2 (Hologic, Inc.), and Prodigy v7.5 (GE Healthcare, Inc.) independent of technologist skill. Methods Duplicate spine and dual hip scans on 90 women were acquired on both Delphi and Prodigy DXA systems at three clinics. BMD measures were converted to standardized BMD (sBMD) units. Precision errors were described as a root-mean-square (RMS) standard deviations and RMS percent coefficients of variation across the population. Results Apex and Delphi values were highly correlated (r ranged from 0.90 to 0.99). Excluding the right neck, the Apex precision error was found to be 20% to 25% lower than the Delphi (spine: 1.0% versus 1.2% (p<0.05), total hip: 1.1% versus 1.3% (p<0.05), right neck: 2.3% versus 2.6% (p> 0.1)). No statistically significant differences were found in the precision error of the Apex and Prodigy (p> 0.05) except for the right neck (2.3% versus 1.8% respectively, p=0.03). Conclusion The Apex software has significantly lower precision error compared to Delphi software with similar mean values, and similar precision to that of the Prodigy.
引用
收藏
页码:1597 / 1602
页数:6
相关论文
共 9 条
  • [1] Monitoring skeletal changes by radiological techniques
    Glüer, CC
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 1999, 14 (11) : 1952 - 1962
  • [2] ACCURATE ASSESSMENT OF PRECISION ERRORS - HOW TO MEASURE THE REPRODUCIBILITY OF BONE DENSITOMETRY TECHNIQUES
    GLUER, CC
    BLAKE, G
    LU, Y
    BLUNT, BA
    JERGAS, M
    GENANT, HK
    [J]. OSTEOPOROSIS INTERNATIONAL, 1995, 5 (04) : 262 - 270
  • [3] A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculation
    Hangartner, T. N.
    [J]. OSTEOPOROSIS INTERNATIONAL, 2007, 18 (04) : 513 - 523
  • [4] RATES OF CHANGE IN BONE-MINERAL DENSITY OF THE SPINE, HEEL, FEMORAL-NECK AND RADIUS IN HEALTHY POSTMENOPAUSAL WOMEN
    HARRIS, S
    DAWSONHUGHES, B
    [J]. BONE AND MINERAL, 1992, 17 (01): : 87 - 95
  • [5] Universal standardization of bone density measurements: A method with optimal properties for calibration among several instruments
    Hui, SL
    Gao, SJ
    Zhou, XH
    Johnston, CC
    Lu, Y
    Gluer, CC
    Grampp, S
    Genant, H
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (09) : 1463 - 1470
  • [6] Standardization of bone mineral density at femoral neck, trochanter and Ward's triangle
    Lu, Y
    Fuerst, T
    Hui, S
    Genant, HK
    [J]. OSTEOPOROSIS INTERNATIONAL, 2001, 12 (06) : 438 - 444
  • [7] THE CLINICAL-APPLICATION OF SERIAL BONE MASS MEASUREMENTS
    ROSS, PD
    DAVIS, JW
    WASNICH, RD
    VOGEL, JM
    [J]. BONE AND MINERAL, 1991, 12 (03): : 189 - 199
  • [8] Comparison of BMD precision for prodigy and delphi spine and femur scans
    Shepherd J.A.
    Fan B.
    Lu Y.
    Lewiecki E.M.
    Miller P.
    Genant H.K.
    [J]. Osteoporosis International, 2006, 17 (9) : 1303 - 1308
  • [9] Cross-calibration and minimum precision standards for dual-energy x-ray absorptiometry: The 2005 ISCD official positions
    Shepherd, JA
    Lu, Y
    Wilson, K
    Fuerst, T
    Genant, H
    Hangartner, TN
    Wilson, C
    Hans, D
    Leib, ES
    [J]. JOURNAL OF CLINICAL DENSITOMETRY, 2006, 9 (01) : 31 - 36